Comparison of an amorphous silicon/cesium iodide flat-panel digital chest radiography system with screen/film and computed radiography systems--a contrast-detail phantom study.

X J Rong1, C C Shaw, X Liu, M R Lemacks, S K Thompson.   

Abstract

Flat-panel (FP) based digital radiography systems have recently been introduced as a new and improved digital radiography technology; it is important to evaluate and compare this new technology with currently widely used conventional screen/film (SF) and computed radiography (CR) techniques. In this study, the low-contrast performance of an amorphous silicon/cesium iodide (aSi/Csl)-based flat-panel digital chest radiography system is compared to those of a screen/film and a computed radiography system by measuring their contrast-detail curves. Also studied were the effects of image enhancement in printing the digital images and dependence on kVp and incident exposure. It was found that the FP system demonstrated significantly better low-contrast performance than the SF or CR systems. It was estimated that a dose savings of 70%-90% could be achieved to match the low-contrast performance of the FP images to that of the SF images. This dose saving was also found to increase with the object size. No significant difference was observed in low-contrast performances between the SF and CR systems. The use of clinical enhancement protocols for printing digital images was found to be essential and result in better low-contrast performance. No significant effects were observed for different kVps. From the results of this contrast-detail phantom study, the aSi/CsI-based flat-panel digital chest system should perform better under clinical situations for detection of low-contrast objects such as lung nodules. However, proper processing prior to printing would be essential to realizing this better performance.

Entities:  

Mesh:

Substances:

Year:  2001        PMID: 11764040     DOI: 10.1118/1.1408620

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  19 in total

1.  A clinical evaluation of the image quality computer program, CoCIQ.

Authors:  E Norrman; M Gårdestig; J Persliden; H Geijer
Journal:  J Digit Imaging       Date:  2005-06       Impact factor: 4.056

2.  Direct detector radiography versus dual reading computed radiography: feasibility of dose reduction in chest radiography.

Authors:  Michael Gruber; Martin Uffmann; Michael Weber; Mathias Prokop; Csilla Balassy; Cornelia Schaefer-Prokop
Journal:  Eur Radiol       Date:  2006-01-11       Impact factor: 5.315

3.  Structural noise from automatic exposure control device and its relationship to X-ray tube voltage used for calibration of a flat-panel detector system.

Authors:  Masayoshi Mizuta; Hiroyuki Akazawa; Toshifumi Kasai; Shigeru Sanada; Shuji Abe; Shigeki Mitou
Journal:  Radiol Phys Technol       Date:  2011-08-23

4.  An investigation into the validity of utilising the CDRAD 2.0 phantom for optimisation studies in digital radiography.

Authors:  Sadeq Al-Murshedi; Peter Hogg; Andrew England
Journal:  Br J Radiol       Date:  2018-07-05       Impact factor: 3.039

5.  Basic imaging properties of an indirect flat-panel detector system employing irradiation side sampling (ISS) technology for chest radiography: comparison with a computed radiographic system.

Authors:  Nobukazu Tanaka; Yuki Yano; Hidetake Yabuuchi; Tsutomu Akasaka; Masayuki Sasaki; Masafumi Ohki; Junji Morishita
Journal:  Radiol Phys Technol       Date:  2012-11-10

6.  Optimisation of radiological protocols for chest imaging using computed radiography and flat-panel X-ray detectors.

Authors:  G Compagnone; M Casadio Baleni; E Di Nicola; M Valentino; M Benati; L F Calzolaio; N Oberhofer; E Fabbri; S Domenichelli; L Barozzi
Journal:  Radiol Med       Date:  2012-10-22       Impact factor: 3.469

7.  A feasibility study on reduction of the entrance-surface dose to neonates by use of a new digital mobile X-ray system.

Authors:  Satoru Utsunomiya; Hajime Monzen; Mami Akimoto; Nobutaka Mukumoto; Yoshitomo Ishihara; Takehiko Shiinoki; Mitsuhiro Nakamura; Yuki Miyabe; Sayaka Sato; Satoru Matsuo; Masahiro Hiraoka
Journal:  Radiol Phys Technol       Date:  2012-10-20

8.  Investigation of optimum anti-scatter grid selection for digital radiography: physical imaging properties and detectability of low-contrast signals.

Authors:  Nobukazu Tanaka; Kentaro Naka; Aya Saito; Junji Morishita; Fukai Toyofuku; Masafumi Ohki; Yoshiharu Higashida
Journal:  Radiol Phys Technol       Date:  2012-08-08

9.  Low dose high energy x-ray in-line phase sensitive imaging prototype: Investigation of optimal geometric conditions and design parameters.

Authors:  Muhammad U Ghani; Aimin Yan; Molly D Wong; Yuhua Li; Liqiang Ren; Xizeng Wu; Hong Liu
Journal:  J Xray Sci Technol       Date:  2015       Impact factor: 1.535

10.  Dose and detectability improvements with high energy phase sensitive x-ray imaging in comparison to low energy conventional imaging.

Authors:  Molly Donovan Wong; Aimin Yan; Muhammad Ghani; Yuhua Li; Laurie Fajardo; Xizeng Wu; Hong Liu
Journal:  Phys Med Biol       Date:  2014-04-15       Impact factor: 3.609

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.