A feasibility study on reduction of the entrance-surface dose to neonates by use of a new digital mobile X-ray system.

Satoru Utsunomiya1, Hajime Monzen, Mami Akimoto, Nobutaka Mukumoto, Yoshitomo Ishihara, Takehiko Shiinoki, Mitsuhiro Nakamura, Yuki Miyabe, Sayaka Sato, Satoru Matsuo, Masahiro Hiraoka.   

Abstract

We investigated the neonatal entrance-surface dose (ESD) and doses of scattered radiation emitted by a digital mobile X-ray system. The system is equipped with a novel flat-panel detector and is used in the neonatal intensive care unit. In the present study, the following three experiments were performed on frequently used X-ray-imaging condition: (1) the digital characteristics of the FPD were evaluated; (2) the ESD to a water-equivalent phantom was measured with a patient skin dosimeter (PSD); and (3) the scattered radiation around the incubator was measured with an ionization chamber survey meter. The digital characteristic curve showed that the system had excellent linearity and that the contrast characteristics were not affected by the tube voltage in the range of 50-110 kV. The ESD was 51-52 μGy with an 8-cm-thick phantom and 33-34 μGy with a 4.5-cm phantom, for one exposure. The doses measured around the incubator were 0.1-0.6 μSv or below measurable limits. Use of the new device demonstrates the potential of reducing the ESD to the patient and operator.

Entities:  

Mesh:

Year:  2012        PMID: 23086629     DOI: 10.1007/s12194-012-0182-1

Source DB:  PubMed          Journal:  Radiol Phys Technol        ISSN: 1865-0333


  13 in total

1.  Imaging the neonate in the incubator: an investigation of the technical, radiological and nursing issues.

Authors:  S J Mutch; S D P Wentworth
Journal:  Br J Radiol       Date:  2007-09-17       Impact factor: 3.039

2.  [Comparison of basic imaging properties for digital radiographic systems based on a direct conversion type of flat panel detector].

Authors:  Shinji Sakai; Reiji Katayama; Junji Morishita; Taro Sakaguchi; Seiji Ohkubo; Takashi Maeda; Naofumi Hayabuchi
Journal:  Nihon Hoshasen Gijutsu Gakkai Zasshi       Date:  2008-01-20

3.  [Input-output characteristics].

Authors:  Kenji Kishimoto
Journal:  Nihon Hoshasen Gijutsu Gakkai Zasshi       Date:  2009-07-20

4.  Radiological protection and safety in medicine. A report of the International Commission on Radiological Protection.

Authors: 
Journal:  Ann ICRP       Date:  1996

5.  Correlation of contrast-detail analysis and clinical image quality assessment in chest radiography with a human cadaver study.

Authors:  An De Crop; Klaus Bacher; Tom Van Hoof; Peter V Smeets; Barbara S Smet; Merel Vergauwen; Urszula Kiendys; Philippe Duyck; Koenraad Verstraete; Katharina D'Herde; Hubert Thierens
Journal:  Radiology       Date:  2011-11-04       Impact factor: 11.105

6.  Radiation exposure during chest X-ray examinations in a premature intensive care unit: phantom studies.

Authors:  T Duetting; B Foerste; T Knoch; K Darge; J Troeger
Journal:  Pediatr Radiol       Date:  1999-03

7.  Dose optimization in pediatric cardiac x-ray imaging.

Authors:  Amber J Gislason; Andrew G Davies; Arnold R Cowen
Journal:  Med Phys       Date:  2010-10       Impact factor: 4.071

8.  Comparison of an amorphous silicon/cesium iodide flat-panel digital chest radiography system with screen/film and computed radiography systems--a contrast-detail phantom study.

Authors:  X J Rong; C C Shaw; X Liu; M R Lemacks; S K Thompson
Journal:  Med Phys       Date:  2001-11       Impact factor: 4.071

Review 9.  Solid-state, flat-panel, digital radiography detectors and their physical imaging characteristics.

Authors:  A R Cowen; S M Kengyelics; A G Davies
Journal:  Clin Radiol       Date:  2008-01-31       Impact factor: 2.350

10.  Patient dose in neonatal units.

Authors:  K Smans; L Struelens; M Smet; H Bosmans; F Vanhavere
Journal:  Radiat Prot Dosimetry       Date:  2008-08-30       Impact factor: 0.972

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.