MR imaging in the presence of vascular stents: A systematic assessment of artifacts for various stent orientations, sequence types, and field strengths.

T Klemm1, S Duda, J Machann, K Seekamp-Rahn, L Schnieder, C D Claussen, F Schick.   

Abstract

A systematic evaluation of the potential quality of magnetic resonance images recorded in the presence of metallic stents was performed on a low-field open imager operating at 0.2 T and on a high-field closed unit operating at 1.0 T. Eight different stent types were examined by two-dimensional gradient-echo sequences with echo times of 4 and 10 msec and by a fast spin-echo technique. In addition, a three-dimensional gradient-echo sequence was applied with an echo time of 2.4 msec. A set of sequence and slice parameters was used on both scanners. Thus, artifacts due to susceptibility effects depending on the magnetic field strength could be distinguished from radiofrequency shielding effects in the lumen of the stents (independent of the field strength). Nine different orthogonal orientations of the stent axis and the image (in terms of slice, read, and phase-encoding direction) were tested, and the artifacts (extension of signal void and visibility of the lumen) were compared. The optimal strategy for visualization of vascular and perivascular regions outside the stents was fast spin-echo imaging with the stent axis and read direction parallel to the static field. Susceptibility-induced signal void in gradient-echo images was minimal using the three-dimensional approach. Increased transmitter amplitudes above usual values provided clearly improved insight in the lumen using gradient-echo sequences. Copyright 2000 Wiley-Liss, Inc.

Mesh:

Substances:

Year:  2000        PMID: 11042644     DOI: 10.1002/1522-2586(200010)12:4<606::aid-jmri14>3.0.co;2-j

Source DB:  PubMed          Journal:  J Magn Reson Imaging        ISSN: 1053-1807            Impact factor:   4.813


  23 in total

1.  Stent-assisted coil placement in a wide-necked persistent trigeminal artery aneurysm with jailing of the trigeminal artery: a case report.

Authors:  Mohammed I Mohammed; Johnny S Sandhu; Ajay K Wakhloo
Journal:  AJNR Am J Neuroradiol       Date:  2002-03       Impact factor: 3.825

2.  Comparing different MR angiography strategies of carotid stents in a vascular flow model: toward stent-specific recommendations in MR follow-up.

Authors:  Andreas M J Frölich; Sara M Pilgram-Pastor; Marios N Psychogios; Alexander Mohr; Michael Knauth
Journal:  Neuroradiology       Date:  2010-08-19       Impact factor: 2.804

3.  Real-time magnetic resonance imaging to guide pediatric endovascular procedures.

Authors:  A N Raval; R J Lederman
Journal:  Pediatr Cardiol       Date:  2005 May-Jun       Impact factor: 1.655

4.  Carotid artery stents: in vitro comparison of different stent designs and sizes using CT angiography and contrast-enhanced MR angiography at 1.5T and 3T.

Authors:  M Lettau; A Sauer; S Heiland; S Rohde; M Bendszus; S Hähnel
Journal:  AJNR Am J Neuroradiol       Date:  2009-09-12       Impact factor: 3.825

5.  Contrast-Enhanced and Time-of-Flight MRA at 3T Compared with DSA for the Follow-Up of Intracranial Aneurysms Treated with the WEB Device.

Authors:  C Timsit; S Soize; A Benaissa; C Portefaix; J-Y Gauvrit; L Pierot
Journal:  AJNR Am J Neuroradiol       Date:  2016-04-21       Impact factor: 3.825

6.  Angiographic CT: in vitro comparison of different carotid artery stents-does stent orientation matter?

Authors:  Michael Lettau; Martin Bendszus; Stefan Hähnel
Journal:  Neuroradiology       Date:  2013-02-15       Impact factor: 2.804

7.  Carotid stent delivery in an XMR suite: immediate assessment of the physiologic impact of extracranial revascularization.

Authors:  Alastair J Martin; David A Saloner; Timothy P L Roberts; Heidi Roberts; Oliver M Weber; William Dillon; Sean Cullen; Van Halbach; Christopher F Dowd; Randall T Higashida
Journal:  AJNR Am J Neuroradiol       Date:  2005-03       Impact factor: 3.825

8.  Off-resonance magnetic resonance angiography improves visualization of in-stent lumen in peripheral nitinol stents compared to conventional T1-weighted acquisitions: an in vitro comparison study.

Authors:  Gitsios Gitsioudis; Philipp Fortner; Matthias Stuber; Anna Missiou; Florian Andre; Oliver J Müller; Hugo A Katus; Grigorios Korosoglou
Journal:  Int J Cardiovasc Imaging       Date:  2016-08-17       Impact factor: 2.357

9.  In vitro comparison of different carotid artery stents: a pixel-by-pixel analysis using CT angiography and contrast-enhanced MR angiography at 1.5 and 3 T.

Authors:  Michael Lettau; Annett Sauer; Sabine Heiland; Stefan Rohde; Julia Reinhardt; Martin Bendszus; Stefan Hähnel
Journal:  Neuroradiology       Date:  2009-11-19       Impact factor: 2.804

10.  Mid-term 3T MRA follow-up of intracranial aneurysms treated with the Woven EndoBridge.

Authors:  Sbt van Rooij; J P Peluso; M Sluzewski; H G Kortman; I Boukrab; W J van Rooij
Journal:  Interv Neuroradiol       Date:  2018-07-12       Impact factor: 1.610

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.