Measuring anterior chamber depth with Orbscan Topography System.

G U Auffarth1, M R Tetz, Y Biazid, H E Völcker.   

Abstract

PURPOSE: To assess the reliability of the Orbscan Topography System in measuring anterior chamber depth (ACD).
SETTING: Department of Ophthalmology, Ruprecht-Karls-University of Heidelberg, Heidelberg, Germany.
METHODS: The Orbscan Topography System is a 3-D, scanning slit beam system for analyzing corneal surfaces as well as structures of the anterior segment, e.g., iris, lens. Reflected surface data points are measured in the x, y, and z axes, creating color-coded true surface topography maps. We measured ACD prior to cataract surgery in 56 eyes of 37 patients (mean age 70.4 years + 13.2 [SD]) using the Orbscan system and, as a control, the Grieshaber Biometric System (ultrasound) and the Jäger ACD slitlamp measurement system.
RESULTS: Mean ACD values measured with the Orbscan system were 3.23 +/- 0.55 mm; ultrasound measurements were 3.27 +/- 0.54 mm and Jäger measurements, 3.10 +/- 0.55 mm. The average difference between Orbscan and ultrasound values was 0.04 +/- 0.15 mm. The correlation coefficient between Orbscan and ultrasound was 0.96 (P < .00001) and between Orbscan and Jäger, 0.899 (P < .0001).
CONCLUSION: The ACD measurements by the Orbscan system were equivalent to common reference measurements.

Entities:  

Mesh:

Year:  1997        PMID: 9423907     DOI: 10.1016/s0886-3350(97)80114-9

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  5 in total

1.  Evaluation of corneal thickness and topography in normal eyes using the Orbscan corneal topography system.

Authors:  Z Liu; A J Huang; S C Pflugfelder
Journal:  Br J Ophthalmol       Date:  1999-07       Impact factor: 4.638

2.  [Intermittent ciliary block as a result of haemodialysis].

Authors:  H-G Schäfer; K Kaczmarek; R Gockeln; M W Meyer; R Winter; C Erb
Journal:  Ophthalmologe       Date:  2006-03       Impact factor: 1.059

3.  [Scheimpflug and topography systems in ophthalmologic diagnostics].

Authors:  G U Auffarth; A F M Borkenstein; A Ehmer; A Mannsfeld; T M Rabsilber; M P Holzer
Journal:  Ophthalmologe       Date:  2008-09       Impact factor: 1.059

4.  Comparison of techniques for measuring anterior chamber depth: Orbscan imaging, Smith's technique, and van Herick's method.

Authors:  Frank Eperjesi; Claire Holden
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2010-09-10       Impact factor: 3.117

5.  Anterior chamber depth measured by two methods in myopic and hyperopic phakic IOL implant.

Authors:  M Vetrugno; N Cardascia; L Cardia
Journal:  Br J Ophthalmol       Date:  2000-10       Impact factor: 4.638

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.