Literature DB >> 10819281

Cemented rotating-platform total knee replacement. A nine to twelve-year follow-up study.

J J Callaghan1, M W Squire, D D Goetz, P M Sullivan, R C Johnston.   

Abstract

BACKGROUND: Although the LCS (low contact stress) rotating-platform mobile-bearing knee replacement has been used extensively, there have been few intermediate or long-term clinical and radiographic follow-up studies evaluating the device. The purpose of this study was to report the nine to twelve-year results of a consecutive series of patients who had a primary total knee replacement performed with this device.
METHODS: Between November 1985 and November 1988, the senior author (R. C. J.) performed 119 consecutive total knee arthroplasties in eighty-six patients with LCS rotating-platform femoral and tibial components and a Townley all-polyethylene dome patellar component. All components were fixed with cement. The average age of the patients at the time of the operation was seventy years (range, thirty-seven to eighty-eight years). Fifty-two patients (seventy-six knees) were female, and thirty-four patients (forty-three knees) were male. The patients were evaluated with clinical knee ratings and radiographic analysis nine to twelve years following the knee replacement.
RESULTS: At the time of the nine to twelve-year follow-up, sixty-four patients (eighty-six knees) were alive, eighteen patients (twenty-eight knees) had died, and four patients (five knees) had been lost to follow-up. Of the 114 knees in the eighty-two patients for whom the final outcome was known, none required a reoperation and none had a dislocation of the mobile-bearing prosthesis. For the forty-five patients (sixty-six knees) who returned for final clinical and radiographic follow-up examinations at nine to twelve years, the average clinical and functional Knee Society ratings were 30 points (range, 2 to 70 points) and 44 points (range, 0 to 80 points) preoperatively and 90 points (range, 63 to 102 points) and 75 points (range, 30 to 100 points) at the final follow-up evaluation. The average Hospital for Special Surgery knee rating was 57 points (range, 28 to 80 points) preoperatively and 84 points (range, 59 to 97 points) at the final follow-up evaluation. The average active range of knee flexion was from 0 degrees (range, 0 to 10 degrees) to 102 degrees (range, 15 to 120 degrees) at the final follow-up evaluation. Seven of the sixty-six knees were painful anteriorly. There was no periprosthetic osteolysis and no evidence of loosening on follow-up radiographs.
CONCLUSIONS: After nine to twelve years of follow-up, the cemented LCS rotating-platform knee replacement was found to be performing well, with durable clinical and radiographic results.

Entities:  

Mesh:

Substances:

Year:  2000        PMID: 10819281

Source DB:  PubMed          Journal:  J Bone Joint Surg Am        ISSN: 0021-9355            Impact factor:   5.284


  28 in total

1.  [Total knee arthroplasty--navigation as the standard].

Authors:  L Kinzl; F Gebhard; P Keppler
Journal:  Chirurg       Date:  2004-10       Impact factor: 0.955

2.  Dynamic in-vivo tibio-femoral and bearing motions in mobile bearing knee arthroplasty.

Authors:  S Fantozzi; A Leardini; S A Banks; M Marcacci; S Giannini; F Catani
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2003-07-09       Impact factor: 4.342

3.  Sex-related outcome differences after implantation of low-contact-stress mobile-bearing total knee arthroplasty.

Authors:  Norbert Kastner; Gerald Gruber; Birgit A Aigner; Jörg Friesenbichler; Michael Pechmann; Florentine Fürst; Patrick Vavken; Andreas Leithner; Patrick Sadoghi
Journal:  Int Orthop       Date:  2012-01-25       Impact factor: 3.075

4.  International multi-centre survivorship analysis of mobile bearing total knee arthroplasty.

Authors:  James B Stiehl; Karel J Hamelynck; Paul E Voorhorst
Journal:  Int Orthop       Date:  2006-03-18       Impact factor: 3.075

5.  The Self-Aligning knee prosthesis: clinical and radiological outcome and survival analysis at 5-year follow up.

Authors:  M J F Diks; P G Anderson; J C A D Janssen; G van Stralen; A B Wymenga
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2007-11-14       Impact factor: 4.342

6.  Two-year follow-up on joint stability and muscular function comparing rotating versus fixed bearing TKR.

Authors:  C Luring; H Bathis; F Oczipka; C Trepte; H Lufen; L Perlick; J Grifka
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2005-12-14       Impact factor: 4.342

7.  Comparison among total knee arthroplasties with a mobile bearing: menisci versus rotating platform versus AP glide platform.

Authors:  Giuseppe Solarino; Andrea Luca; Luigi Marzo; Lorenzo Scialpi; Giovanni B Solarino
Journal:  Chir Organi Mov       Date:  2008-06-27

8.  Mid-term results with a highly congruous mobile-bearing knee prosthesis.

Authors:  Roger G Lemaire
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2009-08-22       Impact factor: 4.342

9.  Gender-specific outcome after implantation of low-contact-stress mobile-bearing total knee arthroplasty with a minimum follow-up of ten years.

Authors:  Norbert Kastner; Birgit A Aigner; Tobias Meikl; Jörg Friesenbichler; Matthias Wolf; Mathias Glehr; Gerald Gruber; Andreas Leithner; Patrick Sadoghi
Journal:  Int Orthop       Date:  2014-07-16       Impact factor: 3.075

10.  Long-term outcome of low contact stress total knee arthroplasty with different mobile bearing designs.

Authors:  Giuseppe Solarino; Antonio Spinarelli; Massimiliano Carrozzo; Andrea Piazzolla; Giovanni Vicenti; Biagio Moretti
Journal:  Joints       Date:  2014-08-01
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.