Stent longitudinal flexibility: a comparison of 13 stent designs before and after balloon expansion.

J A Ormiston1, S R Dixon, M W Webster, P N Ruygrok, J T Stewart, I Minchington, T West.   

Abstract

Longitudinal flexibility is an important property of coronary stents, facilitating delivery and allowing the expanded stent to conform to vessel contour. Subjective descriptions of flexibility abound, but there are few independent quantitative data to aid stent selection. A three-point bend test was employed to measure stiffness, the reciprocal of flexibility, for 13 stent designs in the unexpanded (bare) state, then after expansion with a 3.5-mm balloon. For eight of the designs, stiffness of the proprietary stent/balloon delivery system was also measured. In the unexpanded state, there was a wide spread of stiffness, which ranged from 0.5+/-0.2 to 91.5+/-10.0 g force/mm, depending on design. Stiffness was least for the coil (Wiktor and Crossflex) and hybrid (AVE GFX and Bard XT) designs. The MultiLink was the most flexible and the Crown the stiffest of the slotted tube designs. All stents became stiffer upon expansion. For most manufacturer-mounted stents, the delivery balloon was the main determinant of stent/balloon delivery system stiffness. Manufacturer-mounted stent profile ranged from 1.15+/-0.11 mm for the Jostent to 1.53 +/- 0.05 mm for the MultiLink system. Independent quantitative assessment of characteristics such as flexibility and profile should aid rational comparison of stent designs.

Entities:  

Mesh:

Year:  2000        PMID: 10816296     DOI: 10.1002/(sici)1522-726x(200005)50:1<120::aid-ccd26>3.0.co;2-t

Source DB:  PubMed          Journal:  Catheter Cardiovasc Interv        ISSN: 1522-1946            Impact factor:   2.692


  17 in total

1.  First human trial of KW39 slotted-tube stents: for percutaneous coronary intervention.

Authors:  Minoru Tanaka; Nelson Hirokazu Tsuno; Kazuaki Mitsudo; Kazushige Kadota; Ryozo Tatami; Masayuki Kato; Kenichi Kato; Akihiko Nogami; Osamu Ishikawa; Koki Takahashi
Journal:  Tex Heart Inst J       Date:  2011

2.  Mechanical behavior of fully expanded commercially available endovascular coronary stents.

Authors:  Josip Tambaca; Suncica Canic; Mate Kosor; R David Fish; David Paniagua
Journal:  Tex Heart Inst J       Date:  2011

3.  Three-dimensional analysis of in vivo coronary stent--coronary artery interactions.

Authors:  Robert Liao; Nathan E Green; S Y James Chen; John C Messenger; Adam R Hansgen; Bertron M Groves; John D Carroll
Journal:  Int J Cardiovasc Imaging       Date:  2004-08       Impact factor: 2.357

Review 4.  Coronary artery stents: identification and evaluation.

Authors:  J Butany; K Carmichael; S W Leong; M J Collins
Journal:  J Clin Pathol       Date:  2005-08       Impact factor: 3.411

5.  Mechanical properties and biocompatibility of plasma-nitrided laser-cut 316L cardiovascular stents.

Authors:  Erdem Arslan; Mustafa C Iğdil; Hilal Yazici; Candan Tamerler; Hakan Bermek; Levent Trabzon
Journal:  J Mater Sci Mater Med       Date:  2007-10-30       Impact factor: 3.896

6.  [New developments in drug-eluting stents].

Authors:  M Kollum; C Bode
Journal:  Herz       Date:  2011-05       Impact factor: 1.443

7.  Asymmetric vascular stent: feasibility study of a new low-porosity patch-containing stent.

Authors:  Ciprian N Ionita; Ann M Paciorek; Kenneth R Hoffmann; Daniel R Bednarek; Junichi Yamamoto; John Kolega; Elad I Levy; L Nelson Hopkins; Stephen Rudin; J Mocco
Journal:  Stroke       Date:  2008-04-24       Impact factor: 7.914

8.  The resolute™ integrity zotarolimus-eluting stent in coronary artery disease: a review.

Authors:  Subhash Banerjee
Journal:  Cardiol Ther       Date:  2013-01-04

Review 9.  Structural Design of Vascular Stents: A Review.

Authors:  Chen Pan; Yafeng Han; Jiping Lu
Journal:  Micromachines (Basel)       Date:  2021-06-29       Impact factor: 2.891

10.  Partially polyurethane-covered stent for cerebral aneurysm treatment.

Authors:  Hussain S Rangwala; Ciprian N Ionita; Stephen Rudin; Robert E Baier
Journal:  J Biomed Mater Res B Appl Biomater       Date:  2009-05       Impact factor: 3.405

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.