Literature DB >> 10733185

Comparison of electrode position in the human cochlea using various perimodiolar electrode arrays.

M Tykocinski1, L T Cohen, B C Pyman, T Roland, C Treaba, J Palamara, M C Dahm, R K Shepherd, J Xu, R S Cowan, N L Cohen, G M Clark.   

Abstract

OBJECTIVE: This study was conducted to evaluate the insertion properties and intracochlear trajectories of three perimodiolar electrode array designs and to compare these designs with the standard Cochlear/Melbourne array.
BACKGROUND: Advantages to be expected of a perimodiolar electrode array include both a reduction in stimulus thresholds and an increase in dynamic range, resulting in a more localized stimulation pattern of the spiral ganglion cells, reduced power consumption, and, therefore, longer speech processor battery life.
METHODS: The test arrays were implanted into human temporal bones. Image analysis was performed on a radiograph taken after the insertion. The cochleas were then histologically processed with the electrode array in situ, and the resulting sections were subsequently assessed for position of the electrode array as well as insertion-related intracochlear damage.
RESULTS: All perimodiolar electrode arrays were inserted deeper and showed trajectories that were generally closer to the modiolus compared with the standard electrode array. However, although the precurved array designs did not show significant insertion trauma, the method of insertion needed improvement. After insertion of the straight electrode array with positioner, signs of severe insertion trauma in the majority of implanted cochleas were found.
CONCLUSIONS: Although it was possible to position the electrode arrays close to the modiolus, none of the three perimodiolar designs investigated fulfilled satisfactorily all three criteria of being easy, safe, and atraumatic to implant.

Entities:  

Mesh:

Year:  2000        PMID: 10733185     DOI: 10.1016/s0196-0709(00)80010-1

Source DB:  PubMed          Journal:  Am J Otol        ISSN: 0192-9763


  24 in total

1.  [Cochlear implantation with preservation of residual deep frequency hearing].

Authors:  W Gstöttner; S-M Pok; S Peters; J Kiefer; O Adunka
Journal:  HNO       Date:  2005-09       Impact factor: 1.284

2.  Multisection CT as a valuable tool in the postoperative assessment of cochlear implant patients.

Authors:  Berit M Verbist; Johan H M Frijns; Jakob Geleijns; Mark A van Buchem
Journal:  AJNR Am J Neuroradiol       Date:  2005-02       Impact factor: 3.825

Review 3.  The development of the Nucleus Freedom Cochlear implant system.

Authors:  James F Patrick; Peter A Busby; Peter J Gibson
Journal:  Trends Amplif       Date:  2006-12

4.  Frequency map for the human cochlear spiral ganglion: implications for cochlear implants.

Authors:  Olga Stakhovskaya; Divya Sridhar; Ben H Bonham; Patricia A Leake
Journal:  J Assoc Res Otolaryngol       Date:  2007-02-21

5.  Cochlear implant electrode configuration effects on activation threshold and tonotopic selectivity.

Authors:  Russell L Snyder; John C Middlebrooks; Ben H Bonham
Journal:  Hear Res       Date:  2007-10-11       Impact factor: 3.208

6.  Intracochlear assessment of electrode position after cochlear implant surgery by means of multislice computer tomography.

Authors:  Gijs K A van Wermeskerken; Mathias Prokop; Adriaan F van Olphen; Frans W J Albers
Journal:  Eur Arch Otorhinolaryngol       Date:  2007-07-14       Impact factor: 2.503

7.  Considerations for design of future cochlear implant electrode arrays: electrode array stiffness, size, and depth of insertion.

Authors:  Stephen J Rebscher; Alexander Hetherington; Ben Bonham; Peter Wardrop; David Whinney; Patricia A Leake
Journal:  J Rehabil Res Dev       Date:  2008

8.  Cochlear Histopathology as Observed in Two Patients With a Cochlear Implant Electrode With Positioner.

Authors:  Takefumi Kamakura; Joseph B Nadol
Journal:  Otol Neurotol       Date:  2016-07       Impact factor: 2.311

9.  Matched Cohort Comparison Indicates Superiority of Precurved Electrode Arrays.

Authors:  Jourdan T Holder; Robert J Yawn; Ashley M Nassiri; Robert T Dwyer; Alejandro Rivas; Robert F Labadie; René H Gifford
Journal:  Otol Neurotol       Date:  2019-10       Impact factor: 2.311

Review 10.  Cochlear implants: system design, integration, and evaluation.

Authors:  Fan-Gang Zeng; Stephen Rebscher; William Harrison; Xiaoan Sun; Haihong Feng
Journal:  IEEE Rev Biomed Eng       Date:  2008-11-05
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.