The effect of the antiscatter grid on full-field digital mammography phantom images.

D P Chakraborty1.   

Abstract

Computer Analysis of Mammography Phantom Images (CAMPI) is a method for making quantitative measurements of image quality. This article reports on a recent application of this method to a prototype full-field digital mammography (FFDM) machine. Images of a modified ACR phantom were acquired on the General Electric Diagnostic Molybdenum Rhodium (GE-DMR) FFDM machine at a number of x-ray techniques, both with and without the scatter reduction grid. The techniques were chosen so that one had sets of grid and non-grid images with matched doses (200 mrads) and matched gray-scale values (1500). A third set was acquired at constant 26 kVp and varying mAs for both grid conditions. Analyses of the images yielded signal-to-noise-ratio (SNR), contrast and noise corresponding to each target object, and a non-uniformity measure. The results showed that under conditions of equal gray-scale value the grid images were markedly superior, albeit at higher doses than the non-grid images. Under constant dose conditions, the non-grid images were slightly superior in SNR (7%) but markedly less uniform (60%). Overall, the grid images had substantially greater contrast and superior image uniformity. These conclusions applied to the whole kVp range studied for the Mo-Mo target filter combination and 4 cm of breast equivalent material of average composition. These results suggest that use of the non-grid technique in digital mammography with the GE-DMR-FFDM unit, is presently not warranted. With improved uniformity correction procedure, this conclusion would change and one should be able to realize a 14% reduction in patient dose at the same SNR by using a non-grid technique.

Entities:  

Mesh:

Substances:

Year:  1999        PMID: 10036663      PMCID: PMC3452430          DOI: 10.1007/bf03168622

Source DB:  PubMed          Journal:  J Digit Imaging        ISSN: 0897-1889            Impact factor:   4.056


  16 in total

1.  Spectral dependence of glandular tissue dose in screen-film mammography.

Authors:  X Wu; G T Barnes; D M Tucker
Journal:  Radiology       Date:  1991-04       Impact factor: 11.105

2.  Scatter-glare corrections in quantitative dual-energy fluoroscopy.

Authors:  S Y Molloi; C A Mistretta
Journal:  Med Phys       Date:  1988 May-Jun       Impact factor: 4.071

3.  X-ray scatter removal by deconvolution.

Authors:  J A Seibert; J M Boone
Journal:  Med Phys       Date:  1988 Jul-Aug       Impact factor: 4.071

4.  Analysis of the significance of scattered radiation in reduced dose mammography, including magnification effects, scatter suppression, and focal spot and detector blurring.

Authors:  E P Muntz
Journal:  Med Phys       Date:  1979 Mar-Apr       Impact factor: 4.071

5.  An approach to specifying a minimum dose system for mammography using multiparameter optimization techniques.

Authors:  E P Muntz; H Jafroudi; R Jennings; H Bernstein
Journal:  Med Phys       Date:  1985 Jan-Feb       Impact factor: 4.071

6.  Contrast and dose with Mo-Mo, Mo-Rh, and Rh-Rh target-filter combinations in mammography.

Authors:  E L Gingold; X Wu; G T Barnes
Journal:  Radiology       Date:  1995-06       Impact factor: 11.105

7.  A model for optimization of spectral shape in digital mammography.

Authors:  R Fahrig; M J Yaffe
Journal:  Med Phys       Date:  1994-09       Impact factor: 4.071

8.  Optimization of electrostatic imaging systems for minimum patient dose or minimum exposure in mammography.

Authors:  E P Muntz; M Welkowsky; E Kaegi; L Morsell; E Wilkinson; G Jacobson
Journal:  Radiology       Date:  1978-05       Impact factor: 11.105

9.  Image information content and patient exposure.

Authors:  J W Motz; M Danos
Journal:  Med Phys       Date:  1978 Jan-Feb       Impact factor: 4.071

10.  Effect of reduced scatter on radiographic information content and patient exposure: a quantitative demonstration.

Authors:  R F Wagner; G T Barnes; B S Askins
Journal:  Med Phys       Date:  1980 Jan-Feb       Impact factor: 4.071

View more
  3 in total

1.  Full breast digital mammography with an amorphous silicon-based flat panel detector: physical characteristics of a clinical prototype.

Authors:  S Vedantham; A Karellas; S Suryanarayanan; D Albagli; S Han; E J Tkaczyk; C E Landberg; B Opsahl-Ong; P R Granfors; I Levis; C J D'Orsi; R E Hendrick
Journal:  Med Phys       Date:  2000-03       Impact factor: 4.071

2.  An alternate method for using a visual discrimination model (VDM) to optimize soft-copy display image quality.

Authors:  Dev P Chakraborty
Journal:  J Soc Inf Disp       Date:  2006-10       Impact factor: 2.140

Review 3.  Breast cancer imaging: a perspective for the next decade.

Authors:  Andrew Karellas; Srinivasan Vedantham
Journal:  Med Phys       Date:  2008-11       Impact factor: 4.071

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.