Effect of reduced scatter on radiographic information content and patient exposure: a quantitative demonstration.

R F Wagner, G T Barnes, B S Askins.   

Abstract

If a normal screen-film exposure is made using a conventional grid, a reduced exposure may be used with a superior scatter removal device to record the same image information. This theoretical conclusion is not readily demonstrated because of the limited visibility on the underexposed film. In this study radiographs underexposed by a factor of three with a scanning multiple slit assembly (SMSA) were enhanced autoradiographically and found visually to contain more information than properly exposed radiographs obtained under similar conditions using a conventional Bucky grid. This observation was confirmed by contrast scale and noise measurements. Furthermore, the entrance skin exposure of the radiograph using the SMSA was 45% of that required with the grid technique at the same beam quality.

Entities:  

Mesh:

Year:  1980        PMID: 7366537     DOI: 10.1118/1.594773

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  6 in total

1.  The effect of the antiscatter grid on full-field digital mammography phantom images.

Authors:  D P Chakraborty
Journal:  J Digit Imaging       Date:  1999-02       Impact factor: 4.056

2.  On image quality metrics and the usefulness of grids in digital mammography.

Authors:  Han Chen; Mats Danielsson; Cheng Xu; Björn Cederström
Journal:  J Med Imaging (Bellingham)       Date:  2015-02-17

3.  An automatic and accurate x-ray tube focal spot/grid alignment system for mobile radiography: system description and alignment accuracy.

Authors:  David M Gauntt; Gary T Barnes
Journal:  Med Phys       Date:  2010-12       Impact factor: 4.071

Review 4.  Physics and statistics of medical imaging.

Authors:  D G Brown; R F Wagner
Journal:  J Digit Imaging       Date:  1989-11       Impact factor: 4.056

5.  Unified snr analysis of medical imaging systems.

Authors:  Robert F Wagner; David G Brown
Journal:  Phys Med Biol       Date:  1985-06       Impact factor: 3.609

6.  Efficiency of the Human Observer Compared to an Ideal Observer Based on a Generalized NEQ Which Incorporates Scatter and Geometric Unsharpness: Evaluation with a 2AFC Experiment.

Authors:  Iacovos S Kyprianou; Arundhuti Ganguly; Stephen Rudin; Daniel R Bednarek; Brandon D Gallas; Kyle J Myers
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2005
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.