Evaluation in volunteers of the VIA V-ABG automated bedside blood gas, chemistry, and hematocrit monitor.

P L Bailey1, S W McJames, M L Cluff, D T Wells, J A Orr, D R Westenskow, S E Kern.   

Abstract

OBJECTIVE: To evaluate the VIA V-ABG (VIA Medical Corp.) point-of-care blood gas and chemistry monitor in healthy human volunteers, with particular emphasis on the measurement of blood gases.
METHODS: Experimental conditions were varied by intermittently subjecting volunteers to either isocapnic hypercapnia (end-tidal (ET), PETCO2 = 50+/-2 mmHg, ETPO2 = 130+/-5 mmHg) or isocapnic hypoxia (PETCO2 = 42+/-2, PETO2 + 45+/-2 mmHg) in addition to room air breathing. Measurements by the VIA V-ABG device were compared with paired samples and measurements performed by two ABL Radiometers (505 and 500). Analysis of results includes bias and precision plots and comparison of results with minimal performance criteria as established by CLIA.
RESULTS: Nineteen volunteers yielded 222 matched samples. The range of values were 7.32-7.61 for pH, 20.9-51.6 mmHg for PCO2, 27.9-184.5 mmHg for PO2, 134-141 mmol/l for Na, 3.1-4.1 mmol/l for K, and 30.0-50.4% for hematocrit. Bias and precision (+/-2 sd) for pH was 0.01 and 0.04, for PCO2 was 0.4 and 4.8, for PO2 was 1.0 and 17.0, for Na was -0.3 and 5.2, for K was 0.1 and 0.2, and for Hct was 2.0 and 5.4.
CONCLUSIONS: Over the range of blood gas values assessed, blood gas measurements by the VIA V-ABG device were clinically acceptable and met minimal performance criteria utilizing current Medicare CLIA proficiency standards. Performance criteria were also met by the VIA V-ABG device for Na, K, and Hct measurements but the range of values was too narrow to allow characterization of clinical acceptability. The VIA V-ABG device appears to perform well compared with the results which have been published for other point-of-care devices. Comparison between different studies investigating point-of-care devices is difficult due to several factors (range of values measured, comparison device, population studied, etc.). Some of these instruments, including the VIA V-ABG device, may serve quite well as point-of-care devices to perform certain tests at the bedside. Whether or not any of these devices can substitute for traditional laboratory blood gas and chemistry measurements remains an issue that is not adequately studied.

Entities:  

Mesh:

Year:  1998        PMID: 9951759     DOI: 10.1023/a:1009991527491

Source DB:  PubMed          Journal:  J Clin Monit Comput        ISSN: 1387-1307            Impact factor:   2.502


  6 in total

1.  Microprocessor-based near real-time bedside blood chemistry monitor.

Authors:  D K Wong; W S Jordan
Journal:  Int J Clin Monit Comput       Date:  1992

2.  1990 Medicare/CLIA final rules for proficiency testing: minimum intralaboratory performance characteristics (CV and bias) needed to pass.

Authors:  S S Ehrmeyer; R H Laessig; J E Leinweber; J J Oryall
Journal:  Clin Chem       Date:  1990-10       Impact factor: 8.327

3.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

Review 4.  In-line measurement of electrolytes, glucose, and blood gases.

Authors:  M L Smith; N T Smith; E S Nesseler
Journal:  Int Anesthesiol Clin       Date:  1993

5.  Accuracy and precision of a new, portable, handheld blood gas analyzer, the IRMA.

Authors:  J A Wahr; W Lau; K K Tremper; L Hallock; K Smith
Journal:  J Clin Monit       Date:  1996-07

6.  Hand-held blood gas analyzer is accurate in the critical care setting.

Authors:  G P Zaloga; P R Roberts; K Black; J T Santamauro; E Klase; M Suleiman
Journal:  Crit Care Med       Date:  1996-06       Impact factor: 7.598

  6 in total
  3 in total

Review 1.  Clinical need for continuous glucose monitoring in the hospital.

Authors:  Jeffrey I Joseph; Brian Hipszer; Boris Mraovic; Inna Chervoneva; Mark Joseph; Zvi Grunwald
Journal:  J Diabetes Sci Technol       Date:  2009-11-01

Review 2.  Vascular Glucose Sensor Symposium: Continuous Glucose Monitoring Systems (CGMS) for Hospitalized and Ambulatory Patients at Risk for Hyperglycemia, Hypoglycemia, and Glycemic Variability.

Authors:  Jeffrey I Joseph; Marc C Torjman; Paul J Strasma
Journal:  J Diabetes Sci Technol       Date:  2015-06-15

3.  Evaluation of the VIA Blood Chemistry Monitor for Glucose in Healthy and Diabetic Volunteers.

Authors:  Arjunan Ganesh; Brian Hipszer; Navdeep Loomba; Barbara Simon; Marc C Torjman; Jeffrey Joseph
Journal:  J Diabetes Sci Technol       Date:  2008-03
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.