Accuracy and precision of a new, portable, handheld blood gas analyzer, the IRMA.

J A Wahr1, W Lau, K K Tremper, L Hallock, K Smith.   

Abstract

OBJECTIVE: The accuracy and precision of the new IRMA (Immediate Response Mobile Analysis System, Diametrics, Inc., St. Paul, MN) handheld blood gas analyzer was compared with that of two benchtop blood gas analyzers. The IRMA consists of a notebook-sized machine and disposable cartridges, each containing a pH, a CO2 and an O2 electrode, and provides bedside (point-of-care) blood gas analysis.
METHODS: A total of 172 samples (arterial and mixed venous) were obtained from 25 informed, consenting patients undergoing cardiopulmonary bypass. The pH, PCO2 and PO2 of each sample was determined on four blood gas analyzers: NOVA Statlabs Profile 5 (NOVA Biomedical, Waltham, MA), the ABL-50 (Radiometer, West Lake, OH), and two IRMA machines. Linear regression and bias +/- precision were determined, comparing each of the analyzers with the NOVA.
RESULTS: All three machines showed a similar, high degree of correlation with the NOVA for pH, PCO2, and PO2. The bias and precision of the IRMA machines compared with the NOVA was similar to that of the ABL compared with the NOVA for pH (NOVA:ABL -0.005 +/- 0.011; NOVA:IRMA 1 = 0.0026 +/- 0.025; NOVA:IRMA 2 = 0.0021 +/- 0.025), for PCO2 (NOVA:ABL = -1.4 +/- 1.3 mmHg; NOVA: IRMA 1 = -1.3 +/- 1.9 mmHg; NOVA: IRMA 2 = -1.2 +/- 2.1 mmHg) and PO2 (NOVA:ABL = 3.6 +/- 21.1 mmHg; NOVA:IRMA 1 = 3.4 +/- 19.9 mmHg; NOVA:IRMA 2 = 6.3 +/- 20.9 mmHg). The bias found for pH, PCO2, and PO2 was not affected by extremes of temperature (range 25.5-40 degrees C) or hematocrit (range 11-44%) for any machine.
CONCLUSIONS: The new technology incorporated in the IRMA blood gas analyzer provides results with an accuracy that is similar to that of benchtop analyzers, but with all of the advantages of point-of-care analysis.

Entities:  

Mesh:

Substances:

Year:  1996        PMID: 8863112     DOI: 10.1007/bf02221753

Source DB:  PubMed          Journal:  J Clin Monit        ISSN: 0748-1977


  6 in total

1.  Safety aspects of outstationed laboratory equipment.

Authors:  K Cartwright; H Turner
Journal:  J Hosp Infect       Date:  1992-04       Impact factor: 3.926

2.  Sending blood gas specimens through pressurized transport tube systems exaggerates the error in oxygen tension measurements created by the presence of air bubbles.

Authors:  M H McKane; P A Southorn; P J Santrach; M F Burritt; D J Plevak
Journal:  Anesth Analg       Date:  1995-07       Impact factor: 5.108

3.  The effects of air bubbles and time delay on blood gas analysis.

Authors:  S Ishikawa; A Fornier; C Borst; M S Segal
Journal:  Ann Allergy       Date:  1974-08

4.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

5.  Arterial blood gas analysis: potential errors due to the addition of heparin.

Authors:  R D Hamilton; R J Crockett; J H Alpers
Journal:  Anaesth Intensive Care       Date:  1978-08       Impact factor: 1.669

6.  Blood gas analysis: effect of air bubbles in syringe and delay in estimation.

Authors:  C K Biswas; J M Ramos; B Agroyannis; D N Kerr
Journal:  Br Med J (Clin Res Ed)       Date:  1982-03-27
  6 in total
  5 in total

1.  Transcutaneous PTCCO2 measurement in combination with arterial blood gas analysis provides superior accuracy and reliability in ICU patients.

Authors:  Oliver Spelten; Fritz Fiedler; Robert Schier; Wolfgang A Wetsch; Jochen Hinkelbein
Journal:  J Clin Monit Comput       Date:  2015-12-01       Impact factor: 2.502

2.  Continuous neonatal blood gas monitoring using a multiparameter intra-arterial sensor.

Authors:  C Morgan; S J Newell; D A Ducker; J Hodgkinson; D K White; C J Morley; J M Church
Journal:  Arch Dis Child Fetal Neonatal Ed       Date:  1999-03       Impact factor: 5.747

3.  Evaluation in volunteers of the VIA V-ABG automated bedside blood gas, chemistry, and hematocrit monitor.

Authors:  P L Bailey; S W McJames; M L Cluff; D T Wells; J A Orr; D R Westenskow; S E Kern
Journal:  J Clin Monit Comput       Date:  1998-07       Impact factor: 2.502

4.  Influence of an acetate- and a lactate-based balanced infusion solution on acid base physiology and hemodynamics: an observational pilot study.

Authors:  Klaus F Hofmann-Kiefer; Daniel Chappell; Tobias Kammerer; Matthias Jacob; Michaela Paptistella; Peter Conzen; Markus Rehm
Journal:  Eur J Med Res       Date:  2012-07-06       Impact factor: 2.175

5.  Blood gas measurements using the Bayer Rapid Point 405: are we basing our decisions on accurate data?

Authors:  Frederic Sarrazin; Michael J Tessler; Kenneth Kardash; Elizabeth McNamara; Christina Holcroft
Journal:  J Clin Monit Comput       Date:  2007-06-27       Impact factor: 1.977

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.