Literature DB >> 9870015

Interrelation of tissue temperature versus flow velocity in two different kinds of temperature controlled catheter radiofrequency energy applications.

S Grumbrecht1, J Neuzner, H F Pitschner.   

Abstract

UNLABELLED: The influence of blood flow cooling down the energy delivering electrode during temperature controlled radiofrequency energy application is an important factor for ablation success. In this experimental in-vitro study, using tempered saline as blood equivalent, we observed a highly significant increase in tissue temperature, lesion depth and required energy amount with increasing flow velocity. Second, we found significant deeper lesions with use of pulsed radiofrequency energy application compared to continuous application. We conclude that, even with lower electrode temperatures, success can be achieved dependent on the local blood flow velocity, and deeper lesions can be created with the use of pulsed radiofrequency energy application.
BACKGROUND: Success in temperature-controlled radiofrequency (RF) catheter ablation of arrhythmogenic areas in human hearts depend largely (among others) on the size of the electrode, developed pressure of electrode against tissue, as well as on the localization of the thermistor sensor within the electrode. In addition, the blood flow velocity at various sites of ablation is an important factor for the calculation of heat transport from the electrode, which obviously has not been given much consideration of in the past. The aim of the present in-vitro study, therefore, was to evaluate this important factor's influence on the temperature developed at the electrode and within the myocardial tissue. METHODS AND
RESULTS: All experiments were carried out in a bath containing NaCl solution at 37 degrees C. Four different flow velocities were applied (0, 110, 180, 320 ml/cm2 *min). During and after temperature-controlled unipolar radiofrequency energy delivery (60 degrees C, 40 sec) the electrode temperature, the tissue temperature 5 mm in depth, and the total energy delivered were measured, as well as the actual depth of the lesion. The amount of energy applied to the electrode was regulated by the thermosensor in the electrode to obtain a maximum temperature of 60 degrees C. Two different kinds of radiofrequency energy delivery have been used: (1) continuous radiofrequency energy delivery as usual regarding clinical use, (2) pulsed radiofrequency energy delivery with a duty cycle length of 10 ms and a pause of at least the same duration during two consecutive duty cycles. At pulsed radiofrequency energy application, the energy for each duty cycle was held constant during delivery. The amount of pulses delivered to the electrode was regulated by the electrode's thermosensor. With both modes of radiofrequency energy delivery a uniform observation could be made. The more the flow velocity applied accelerated, the more the tissue temperature rose (R = 0.85; p < 0.00000001), and the lesion depth increased in spite of electrode temperature being held constant. The amount of the total energy delivered rose in proportion to the cooling down of the electrode dependent on the flow velocity (R = 0.69, p < 0.0000004). Steady-state temperatures had not been accomplished after 40 sec time. When energy was delivered at the pulsed mode, intramyocardial temperatures proved higher compared to the continuous mode with significant differences (p < 0.05) at comparable flow velocities applied between 180 and 320 ml/cm2*min and at same electrode temperatures. This resulted in significantly (p < 0.05) larger lesion depths in pulsed radiofrequency energy delivery. We suppose that this significant difference can be explained by a higher amount of total energy delivered at comparable electrode temperature in the pulsed mode as compared to the continuous mode.

Entities:  

Mesh:

Year:  1998        PMID: 9870015     DOI: 10.1023/a:1009720003138

Source DB:  PubMed          Journal:  J Interv Card Electrophysiol        ISSN: 1383-875X            Impact factor:   1.900


  14 in total

1.  Subendocardial and intramural temperature response during radiofrequency catheter ablation in chronic myocardial infarction and normal myocardium.

Authors:  H Kottkamp; G Hindricks; E Horst; T Baal; C Fechtrup; G Breithardt; M Borggrefe
Journal:  Circulation       Date:  1997-04-15       Impact factor: 29.690

2.  Temperature monitoring during radiofrequency catheter ablation of accessory pathways.

Authors:  J J Langberg; H Calkins; R el-Atassi; M Borganelli; A Leon; S J Kalbfleisch; F Morady
Journal:  Circulation       Date:  1992-11       Impact factor: 29.690

3.  Radiofrequency catheter ablation: the effect of electrode size on lesion volume in vivo.

Authors:  J J Langberg; M A Lee; M C Chin; M Rosenqvist
Journal:  Pacing Clin Electrophysiol       Date:  1990-10       Impact factor: 1.976

4.  Assessment of effects of a radiofrequency energy field and thermistor location in an electrode catheter on the accuracy of temperature measurement.

Authors:  L T Blouin; F I Marcus; L Lampe
Journal:  Pacing Clin Electrophysiol       Date:  1991-05       Impact factor: 1.976

5.  Temperature and impedance monitoring during slow pathway ablation in patients with AV nodal reentrant tachycardia.

Authors:  S A Strickberger; A Zivin; E G Daoud; F Bogun; M Harvey; R Goyal; M Niebauer; K C Man; F Morady
Journal:  J Cardiovasc Electrophysiol       Date:  1996-04

Review 6.  Temperature monitoring during radiofrequency ablation.

Authors:  J L Dinerman; R D Berger; H Calkins
Journal:  J Cardiovasc Electrophysiol       Date:  1996-02

7.  Tissue heating during radiofrequency catheter ablation: a thermodynamic model and observations in isolated perfused and superfused canine right ventricular free wall.

Authors:  D E Haines; D D Watson
Journal:  Pacing Clin Electrophysiol       Date:  1989-06       Impact factor: 1.976

Review 8.  Coagulation of ventricular myocardium using radiofrequency alternating current: bio-physical aspects and experimental findings.

Authors:  W Haverkamp; G Hindricks; H Gulker; U Rissel; W Pfennings; M Borggrefe; G Breithardt
Journal:  Pacing Clin Electrophysiol       Date:  1989-01       Impact factor: 1.976

9.  Effects of heating with radiofrequency power on myocardial impulse conduction: is radiofrequency ablation exclusively thermally mediated?

Authors:  T A Simmers; J M de Bakker; F H Wittkampf; R N Hauer
Journal:  J Cardiovasc Electrophysiol       Date:  1996-03

10.  Temperature monitoring during radiofrequency catheter ablation procedures using closed loop control. Atakr Multicenter Investigators Group.

Authors:  H Calkins; E Prystowsky; M Carlson; L S Klein; J P Saul; P Gillette
Journal:  Circulation       Date:  1994-09       Impact factor: 29.690

View more
  4 in total

1.  Electrode impedance: an indicator of electrode-tissue contact and lesion dimensions during linear ablation.

Authors:  X Zheng; G P Walcott; J A Hall; D L Rollins; W M Smith; G N Kay; R E Ideker
Journal:  J Interv Card Electrophysiol       Date:  2000-12       Impact factor: 1.900

2.  Homogeneity and diameter of linear lesions induced with multipolar ablation catheters: in vitro and in vivo comparison of pulsed versus continuous radiofrequency energy delivery.

Authors:  A Erdogan; S Grumbrecht; J Carlsson; H Roederich; B Schulte; J Sperzel; A Berkowitsch; J Neuzner; H F Pitschner
Journal:  J Interv Card Electrophysiol       Date:  2000-12       Impact factor: 1.900

3.  The effect of ablation sequence and duration on lesion shape using rapidly pulsed radiofrequency energy through multiple electrodes.

Authors:  I D McRury; S Diamond; G Falwell; A Schlichting; C Wilson
Journal:  J Interv Card Electrophysiol       Date:  2000-04       Impact factor: 1.900

4.  Comparison between pulsed and continuous radiofrequency delivery.

Authors:  Ali Erdogan; Eiko Walleck; Sebastian Rueckleben; Thomas Neumann; Harald H Tillmanns; Bernd Waldecker; Hans Hoelschermann; Martin Heidt
Journal:  J Interv Card Electrophysiol       Date:  2006-12-13       Impact factor: 1.900

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.