The effect of ablation sequence and duration on lesion shape using rapidly pulsed radiofrequency energy through multiple electrodes.

I D McRury1, S Diamond, G Falwell, A Schlichting, C Wilson.   

Abstract

Sequences of energy application to multiple electrodes and a study of ablation duration with distal tip and multi-electrode ablations were explored with a radiofrequency controller that distributes energy from a generator to up to 4 electrodes with various duty cycles. In vitro ablations were performed on bovine left ventricle in circulating blood and lesions in goats were performed to verify the in vitro results. All of the ablation sequences with simultaneous electrode activation of contiguous electrodes resulted in deeper lesions than those created in sequence. There was also no scalloping of the lesion if contiguous electrodes were activated simultaneously. During all distal tip ablations, lesion volume and depth was greater after 3 minutes of energy delivery than after 1 minute, but did not increase from 3 minutes to 5 minutes. There was a significant increase in multi-electrode ablation lesion depth with each additional minute in the ablation cycle. The in vivo ablations verified these results at 120 and 300 second ablations. Pulsed energy distal tip ablations resulted in deeper lesions than continuous only if power amplitudes over 50 W were employed. In conclusion, contiguous electrodes in simultaneous use create lesions that resemble one large lesion rather than two lesions positioned next to each other. Multi-electrode ablation lesions continue to grow at ablation durations of up to 5 minutes compared to distal tip lesions which reach steady-state between 1 and 3 minutes. Pulsed energy delivery to distal tips may result in deeper lesions than conventional if high powers are employed.

Entities:  

Mesh:

Year:  2000        PMID: 10729852     DOI: 10.1023/a:1009898504174

Source DB:  PubMed          Journal:  J Interv Card Electrophysiol        ISSN: 1383-875X            Impact factor:   1.900


  22 in total

Review 1.  The biophysics of radiofrequency catheter ablation in the heart: the importance of temperature monitoring.

Authors:  D E Haines
Journal:  Pacing Clin Electrophysiol       Date:  1993-03       Impact factor: 1.976

2.  Successful irrigated-tip catheter ablation of atrial flutter resistant to conventional radiofrequency ablation.

Authors:  P Jaïs; M Haïssaguerre; D C Shah; A Takahashi; M Hocini; T Lavergne; S Lafitte; A Le Mouroux; B Fischer; J Clémenty
Journal:  Circulation       Date:  1998-09-01       Impact factor: 29.690

3.  Simultaneous multipolar radiofrequency ablation in the monopolar mode increases lesion size.

Authors:  S Mackey; L Thornton; D S He; F I Marcus; L F Lampe
Journal:  Pacing Clin Electrophysiol       Date:  1996-07       Impact factor: 1.976

4.  Radiofrequency catheter ablation of common atrial flutter in 200 patients.

Authors:  B Fischer; P Jaïs; D Shah; S Chouairi; M Haïssaguerre; S Garrigues; F Poquet; L Gencel; J Clémenty; F I Marcus
Journal:  J Cardiovasc Electrophysiol       Date:  1996-12

5.  Increasing catheter ablation lesion size by simultaneous application of radiofrequency current to two adjacent sites.

Authors:  R J Chang; W G Stevenson; L A Saxon; J Parker
Journal:  Am Heart J       Date:  1993-05       Impact factor: 4.749

6.  High energy radiofrequency catheter ablation for common atrial flutter targeting the isthmus between the inferior vena cava and tricuspid valve annulus using a super long tip electrode.

Authors:  Y Iesaka; A Takahashi; M Goya; T Yamane; T Tokunaga; H Amemiya; H Fujiwara; J Nitta; A Nogami; K Aonuma; M Hiroe; F Marumo; M Hiraoka
Journal:  Pacing Clin Electrophysiol       Date:  1998-02       Impact factor: 1.976

7.  Electrode radius predicts lesion radius during radiofrequency energy heating. Validation of a proposed thermodynamic model.

Authors:  D E Haines; D D Watson; A F Verow
Journal:  Circ Res       Date:  1990-07       Impact factor: 17.367

8.  Successful catheter ablation of atrial fibrillation.

Authors:  M Haïssaguerre; L Gencel; B Fischer; P Le Métayer; F Poquet; F I Marcus; J Clémenty
Journal:  J Cardiovasc Electrophysiol       Date:  1994-12

9.  Radiofrequency ablation of the inferior vena cava-tricuspid valve isthmus in common atrial flutter.

Authors:  F G Cosio; M López-Gil; A Goicolea; F Arribas; J L Barroso
Journal:  Am J Cardiol       Date:  1993-03-15       Impact factor: 2.778

10.  Sinus node-atrioventricular node isolation: long-term results with the "corridor" operation for atrial fibrillation.

Authors:  J W Leitch; G Klein; R Yee; G Guiraudon
Journal:  J Am Coll Cardiol       Date:  1991-03-15       Impact factor: 24.094

View more
  1 in total

1.  Remote magnetic versus manual catheters: evaluation of ablation effect in atrial fibrillation by myocardial marker levels.

Authors:  Eivind Solheim; Morten Kristian Off; Per Ivar Hoff; Alessandro De Bortoli; Peter Schuster; Ole-Jørgen Ohm; Jian Chen
Journal:  J Interv Card Electrophysiol       Date:  2011-04-08       Impact factor: 1.900

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.