Comparison of decentration and tilt between one piece and three piece polymethyl methacrylate intraocular lenses.

K Hayashi1, H Hayashi, F Nakao, F Hayashi.   

Abstract

BACKGROUND: The extent of the decentration and tilt was prospectively compared between one piece polymethyl methacrylate (PMMA) and three piece PMMA intraocular lenses (IOLs) which were implanted in the capsular bag after performing continuous curvilinear capsulorhexis.
METHODS: 91 patients underwent a one piece PMMA IOL implantation in one eye as well as the implantation of the three piece PMMA IOL with polyvinylidene fluoride loops in the opposite eye. The length of the lens decentration and the angle of the tilt were quantitated using the anterior eye segment analysis system (EAS-1000) at 1 week as well as 1, 3, and 6 months postoperatively.
RESULTS: The mean length of the decentration in the one piece IOL was smaller than that in the three piece IOL at 1 week (p = 0.0092), 1 month (p = 0.0044), 3 months (p = 0.0069), and 6 months (p = 0.0010) postoperatively. However, no significant difference was found in the degree of the tilt between the two types of IOLs throughout the observation periods.
CONCLUSION: These results clarified that the one piece PMMA IOL with rigid PMMA haptics implanted in the capsular bag provides a better centration than the three piece PMMA IOL with flexible haptics, whereas the tilt was the same between the two types of IOLs.

Entities:  

Mesh:

Substances:

Year:  1998        PMID: 9640193      PMCID: PMC1722563          DOI: 10.1136/bjo.82.4.419

Source DB:  PubMed          Journal:  Br J Ophthalmol        ISSN: 0007-1161            Impact factor:   4.638


  19 in total

1.  Prospective experimental study of factors related to posterior chamber intraocular lens decentration.

Authors:  U F Legler; E I Assia; V E Castaneda; J P Hoggatt; D J Apple
Journal:  J Cataract Refract Surg       Date:  1992-09       Impact factor: 3.351

2.  Intraobserver and interobserver reproducibility of the Nidek EAS-1000 Anterior Eye Segment Analysis System.

Authors:  K A Baez; S Orengo; S Gandham; G L Spaeth
Journal:  Ophthalmic Surg       Date:  1992-06

3.  The correlation between incision size and corneal shape changes in sutureless cataract surgery.

Authors:  K Hayashi; H Hayashi; F Nakao; F Hayashi
Journal:  Ophthalmology       Date:  1995-04       Impact factor: 12.079

4.  Experimental evaluation of posterior capsule opacification and intraocular lens decentration: comparison of intraocular lenses of 12.5 mm and 14.0 mm diameter.

Authors:  S Ohmi; K Uenoyama
Journal:  J Cataract Refract Surg       Date:  1993-05       Impact factor: 3.351

5.  Comparison of shape recovery ratios of single-piece poly(methyl methacrylate) intraocular lens haptics.

Authors:  W Kimura; T Kimura; T Sawada; T Kikuchi; H Nagai; Y Yamada
Journal:  J Cataract Refract Surg       Date:  1993-09       Impact factor: 3.351

6.  Intraocular lens encapsulation by shrinkage of the capsulorhexis opening.

Authors:  O Nishi; K Nishi
Journal:  J Cataract Refract Surg       Date:  1993-07       Impact factor: 3.351

7.  Practice styles and preferences of ASCRS members--1994 survey.

Authors:  D V Leaming
Journal:  J Cataract Refract Surg       Date:  1995-07       Impact factor: 3.351

8.  Current trends in cataract and refractive surgery in Japan--1994 survey.

Authors:  T Oshika; K Masuda; F Hayashi; Y Majima; D V Leaming
Journal:  Jpn J Ophthalmol       Date:  1995       Impact factor: 2.447

9.  Divide and conquer nucleofractis phacoemulsification: development and variations.

Authors:  H V Gimbel
Journal:  J Cataract Refract Surg       Date:  1991-05       Impact factor: 3.351

10.  Progressive constriction of the anterior capsular opening following intact capsulorhexis.

Authors:  S O Hansen; A S Crandall; R J Olson
Journal:  J Cataract Refract Surg       Date:  1993-01       Impact factor: 3.351

View more
  5 in total

1.  [Measurement method for the determination of rotation and decentration of intraocular lenses].

Authors:  K A Becker; G U Auffarth; H E Völcker
Journal:  Ophthalmologe       Date:  2004-06       Impact factor: 1.059

2.  [Clinical results of the aspheric intraocular lens FY-60AD (Hoya) with particular respect to decentration and tilt].

Authors:  U Mester; S Heinen; H Kaymak
Journal:  Ophthalmologe       Date:  2010-09       Impact factor: 1.059

3.  Predictive Value of Excellent Uncorrected Visual Acuity Post-Operative Day One After Cataract Surgery.

Authors:  Jonathan W Young; Nathan W Law; Daniel C Tu
Journal:  Clin Ophthalmol       Date:  2020-09-23

4.  A prospective, randomised comparison of single and three piece acrylic foldable intraocular lenses.

Authors:  R Nejima; K Miyata; M Honbou; T Tokunaga; T Tanabe; M Sato; T Oshika
Journal:  Br J Ophthalmol       Date:  2004-06       Impact factor: 4.638

Review 5.  Factors Affecting Posterior Capsule Opacification in the Development of Intraocular Lens Materials.

Authors:  Grace Cooksley; Joseph Lacey; Marcus K Dymond; Susan Sandeman
Journal:  Pharmaceutics       Date:  2021-06-10       Impact factor: 6.321

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.