Literature DB >> 15197576

[Measurement method for the determination of rotation and decentration of intraocular lenses].

K A Becker1, G U Auffarth, H E Völcker.   

Abstract

PURPOSE: In the past 50 years the IOL has been improved. Nowadays it is possible to exactly calculate the necessary IOL power and correct toric refractive errors. In this study we developed methods for measuring rotation and decentration using the example of a hydrophilic acrylic IOL.
METHODS: Following cataract surgery, the first retroilluminated photograph was taken. The position of the IOL was determined with the image analysis program Adobe Photoshop.
RESULTS: Measurement of a rotational movement or of a decentration of an IOL can be done accurately and rapidly. Evaluation of the stability of rotation of the hydrophilic acrylic IOL measured here showed an average rotation of the IOL of 5.3+/-1.4 degrees after 6 months compared to the position directly after implantation.
CONCLUSION: The measurement methods presented here are easy to use and provide reliable results. Examiners must have basic knowledge of the computer programs used. Also, these methods depend on good quality of the retroilluminated photographs.

Entities:  

Mesh:

Year:  2004        PMID: 15197576     DOI: 10.1007/s00347-003-0951-7

Source DB:  PubMed          Journal:  Ophthalmologe        ISSN: 0941-293X            Impact factor:   1.059


  7 in total

1.  Effect of haptic angulation on posterior capsule opacification in modern foldable lenses with a square, truncated optic edge.

Authors:  Josef M Schmidbauer; Marcela Escobar-Gomez; David J Apple; Qun Peng; Stella N Arthur; Luis G Vargas
Journal:  J Cataract Refract Surg       Date:  2002-07       Impact factor: 3.351

2.  Further observations on intraocular acrylic lenses in cataract surgery.

Authors:  H RIDLEY
Journal:  Trans Am Acad Ophthalmol Otolaryngol       Date:  1953 Jan-Feb

3.  Intra-ocular acrylic lenses after cataract extraction.

Authors:  H RIDLEY
Journal:  Lancet       Date:  1952-01-19       Impact factor: 79.321

4.  Centration of intraocular lenses with circular haptics.

Authors:  V De Groot; P Jonckheere; M J Tassignon
Journal:  J Cataract Refract Surg       Date:  1997-10       Impact factor: 3.351

5.  Centration and fixation of silicone intraocular lenses: clinicopathological findings in human autopsy eyes.

Authors:  G U Auffarth; C McCabe; M Wilcox; J C Sims; T A Wesendahl; D J Apple
Journal:  J Cataract Refract Surg       Date:  1996       Impact factor: 3.351

6.  [Centering the posterior chamber lens in autopsy eyes with and without pseudoexfoliation syndrome].

Authors:  G U Auffarth; K Tsao; T A Wesendahl; D J Apple
Journal:  Ophthalmologe       Date:  1995-10       Impact factor: 1.059

7.  Comparison of decentration and tilt between one piece and three piece polymethyl methacrylate intraocular lenses.

Authors:  K Hayashi; H Hayashi; F Nakao; F Hayashi
Journal:  Br J Ophthalmol       Date:  1998-04       Impact factor: 4.638

  7 in total
  6 in total

1.  [Classification of biomedical research reports as a reference for evidence-based medicine in ophthalmology. A survey considering as example the journal Der Ophthalmologe].

Authors:  H P N Scholl; M Fleckenstein; T U Krohne; F G Holz
Journal:  Ophthalmologe       Date:  2005-12       Impact factor: 1.059

2.  [Accuracy of lens power calculation and centration of an aspheric intraocular lens].

Authors:  K A Becker; M P Holzer; A J Reuland; G U Auffarth
Journal:  Ophthalmologe       Date:  2006-10       Impact factor: 1.059

Review 3.  [Toric IOLs after cataract surgery and refractive lens exchange].

Authors:  G U Auffarth; T M Rabsilber
Journal:  Ophthalmologe       Date:  2007-12       Impact factor: 1.059

4.  Evaluation of Decentration, Tilt and Angular Orientation of Toric Intraocular Lens.

Authors:  Martin Fus; Sarka Pitrova
Journal:  Clin Ophthalmol       Date:  2021-12-19

5.  Rotational stability of the AcrySof SA60TT toric intraocular lenses: a cohort study.

Authors:  Bruno Zuberbuhler; Theo Signer; Richard Gale; Eduard Haefliger
Journal:  BMC Ophthalmol       Date:  2008-05-06       Impact factor: 2.209

6.  Efficacy of Toric intraocular lens implantation in eyes with high myopia: A prospective, case-controlled observational study.

Authors:  Tao Guo; Peng Gao; Li Fang; Li Guo; Yuchen Fan; Changgen Liu
Journal:  Exp Ther Med       Date:  2018-04-25       Impact factor: 2.447

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.