Abstract

AIMS: To compare the postoperative performance of single and three piece acrylic foldable intraocular lenses (IOLs).
METHODS: 20 patients underwent bilateral cataract surgery with a single piece SA30AL IOL in one eye and a three piece MA30BA IOL in the other eye. The eyes were randomly assigned to either a single or three piece lens. The amount of IOL decentration and tilt, area of anterior capsule opening, and degree of posterior capsule opacification were measured using the Scheimpflug anterior segment analysis system (Nidek EAS-1000). Visual acuity and contrast sensitivity were examined. Measurements were performed by masked examiners before and 1 day, 1 week, 1, 3, 6, and 18 months after surgery.
RESULTS: There were no significant differences between the two groups (p>0.05, paired t test) in the amount of IOL decentration, IOL tilt, area of anterior capsule opening, degree of posterior capsule opacification, best corrected visual acuity, and contrast sensitivity throughout the 18 month follow up period.
CONCLUSION: The single and three piece acrylic foldable IOLs are equally stable in the eye after surgery.

Entities:  

Mesh:

Substances:

Year:  2004        PMID: 15148204      PMCID: PMC1772204          DOI: 10.1136/bjo.2003.037663

Source DB:  PubMed          Journal:  Br J Ophthalmol        ISSN: 0007-1161            Impact factor:   4.638


  35 in total

1.  Preventing lens epithelial cell migration using intraocular lenses with sharp rectangular edges.

Authors:  O Nishi; K Nishi; K Wickström
Journal:  J Cataract Refract Surg       Date:  2000-10       Impact factor: 3.351

2.  Anterior capsule contraction and intraocular lens decentration and tilt after hydrogel lens implantation.

Authors:  K Hayashi; H Hayashi; F Nakao; F Hayashi
Journal:  Br J Ophthalmol       Date:  2001-11       Impact factor: 4.638

3.  Preliminary results of cataract extraction with implantation of a single-piece AcrySof intraocular lens.

Authors:  Aldo Caporossi; Fabrizio Casprini; Gian Marco Tosi; Stefano Baiocchi
Journal:  J Cataract Refract Surg       Date:  2002-04       Impact factor: 3.351

4.  Effect of round-edged acrylic intraocular lenses on preventing posterior capsule opacification.

Authors:  O Nishi; K Nishi; J Akura; T Nagata
Journal:  J Cataract Refract Surg       Date:  2001-04       Impact factor: 3.351

5.  Comparison of AcrySof MA30BA and Sensar AR40 acrylic intraocular lenses.

Authors:  Fabrizio Casprini; Gian Marco Tosi; Pier Paolo Quercioli; Aldo Caporossi
Journal:  J Cataract Refract Surg       Date:  2002-07       Impact factor: 3.351

6.  Posterior capsule opacification with AcrySof and poly(methyl methacrylate) intraocular lenses. Comparative study with a 3-year follow-up.

Authors:  K Sundelin; Y Friberg-Riad; A Ostberg; J Sjöstrand
Journal:  J Cataract Refract Surg       Date:  2001-10       Impact factor: 3.351

7.  Anterior capsular contraction after cataract surgery in eyes of diabetic patients.

Authors:  S Kato; T Oshika; J Numaga; Y Hayashi; M Oshiro; T Yuguchi; T Kaiya
Journal:  Br J Ophthalmol       Date:  2001-01       Impact factor: 4.638

8.  Current trends in cataract and refractive surgery in Japan: 1999 survey.

Authors:  T Oshika; S Amano; M Araie; Y Majima; D V Leaming
Journal:  Jpn J Ophthalmol       Date:  2001 Jul-Aug       Impact factor: 2.447

9.  Relationship of AcrySof acrylic and PhacoFlex silicone intraocular lenses to visual acuity and posterior capsule opacification.

Authors:  Michael T Halpern; Dave Covert; Carmelina Battista; Arthur J Weinstein; Ralph D Levinson; Lihan Yan
Journal:  J Cataract Refract Surg       Date:  2002-04       Impact factor: 3.351

10.  Clinical performance of Alcon SA30AL and SA60AT single-piece acrylic intraocular lenses.

Authors:  James A Davison
Journal:  J Cataract Refract Surg       Date:  2002-07       Impact factor: 3.351

View more
  13 in total

1.  [Influence of intraocular lens material and design on the development of posterior capsule opacification].

Authors:  O Nishi
Journal:  Ophthalmologe       Date:  2005-06       Impact factor: 1.059

2.  Risk factors for posterior capsule opacification.

Authors:  S D McLeod
Journal:  Br J Ophthalmol       Date:  2005-11       Impact factor: 4.638

3.  Nd:YAG capsulotomy rates after use of the AcrySof acrylic three piece and one piece intraocular lenses.

Authors:  S I Mian; K Fahim; A Marcovitch; H Gada; D C Musch; A Sugar
Journal:  Br J Ophthalmol       Date:  2005-11       Impact factor: 4.638

4.  Prospective, non-randomised, long term clinical evaluation of a foldable hydrophilic single piece intraocular lens: results of the Centerflex FDA study.

Authors:  K A Becker; M Martin; T M Rabsilber; B B Entz; A J Reuland; G U Auffarth
Journal:  Br J Ophthalmol       Date:  2006-05-10       Impact factor: 4.638

5.  Influence of tilt and decentration of scleral-sutured intraocular lens on ocular higher-order wavefront aberration.

Authors:  T Oshika; G Sugita; K Miyata; T Tokunaga; T Samejima; C Okamoto; Y Ishii
Journal:  Br J Ophthalmol       Date:  2006-08-16       Impact factor: 4.638

6.  [Refractive predictability and stability of three-piece versus single-piece intraocular lenses in patients with high axial myopia].

Authors:  A Frings; I Steinbrugger; E Wissiak; B Vidic; Y El-Shabrawi; N Ardjomand
Journal:  Ophthalmologe       Date:  2012-06       Impact factor: 1.059

7.  Influence of three-piece and single-piece designs of two sharp-edge optic hydrophobic acrylic intraocular lenses on the prevention of posterior capsule opacification: a prospective, randomised, long-term clinical trial.

Authors:  Reda Zemaitiene; Vytautas Jasinskas; Gerd U Auffarth
Journal:  Br J Ophthalmol       Date:  2006-11-23       Impact factor: 4.638

8.  Comparisons of the in-the-bag stabilities of single-piece and three-piece intraocular lenses for age-related cataract patients: a randomized controlled trial.

Authors:  Xiaojian Zhong; Erping Long; Wan Chen; Wu Xiang; Zhaochuan Liu; Hui Chen; Jingjing Chen; Zhuoling Lin; Haotian Lin; Weirong Chen
Journal:  BMC Ophthalmol       Date:  2016-07-08       Impact factor: 2.209

9.  Comparison of Nd: YAG capsulotomy rate between 1-piece and 3-piece acrylic intraocular lenses: A STROBE-compliant article.

Authors:  Sun Young Kim; Young Kwon Chung; Hye Young Shin; Mee Yon Lee; Young-Chun Lee; Su-Young Kim
Journal:  Medicine (Baltimore)       Date:  2017-07       Impact factor: 1.889

10.  IOL tilt and decentration estimation from 3 dimensional reconstruction of OCT image.

Authors:  Xiaogang Wang; Jing Dong; Xiaoliang Wang; Qiang Wu
Journal:  PLoS One       Date:  2013-03-15       Impact factor: 3.240

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.