Evaluation of a pulse oximeter sensor tester.

Shuba Dugani1, Iljaz Hodzovic, Seema Sindhakar, Aida Nadra, Clare Dunstan, Anthony Richard Wilkes, John Mecklenburgh.   

Abstract

OBJECTIVE: The Lightman is intended to test the optical and electrical properties of a pulse oximeter probe including the wavelength of the light emitting diode by means of a micro spectrometer. The aim of this study was to evaluate the ability of the Lightman to detect faulty pulse oximeter finger probes by testing the accuracy of the wavelength of the light emitting diode in isolation from the monitor.
METHODS: The pulse oximeter measurements of arterial oxygen saturation from the "accurate" and "inaccurate" probes, as identified by the Lightman, were compared with arterial saturation determined by a co-oximeter. Data was analysed from 63 sets of measurements. In addition, we conducted a national survey to determine the testing procedures used by the Biomedical Engineering departments to evaluate the accuracy of pulse oximeter devices.
RESULTS: The bias [95% limits of agreement] for accurate, over-reading and under-reading probes were 0.17% [3.6 to -3.3], 1.44% [5.4 to -2.5] and -1.6% [2.6 to -5.8] respectively. The response rate to the national survey was 75% (142/189); a pulse oximeter tester was used by 93/142 (65%) trusts.
CONCLUSIONS: Our findings suggest that the Lightman can detect faulty probes and predict reasonably accurately the direction of the probe's error. The Lightman may be considered as a useful tool to assess the accuracy of pulse oximeters. The national survey highlighted a wide variation in the testing procedure utilised to evaluate the accuracy of pulse oximeters. Introduction of guidelines regarding the testing procedure would promote a uniform practice.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21590358     DOI: 10.1007/s10877-011-9283-3

Source DB:  PubMed          Journal:  J Clin Monit Comput        ISSN: 1387-1307            Impact factor:   2.502


  19 in total

1.  BTS guideline for emergency oxygen use in adult patients.

Authors:  B R O'Driscoll; L S Howard; A G Davison
Journal:  Thorax       Date:  2008-10       Impact factor: 9.139

Review 2.  Pulse oximetry.

Authors:  J A Wahr; K K Tremper; M Diab
Journal:  Respir Care Clin N Am       Date:  1995-09

3.  The effect of fingerprinting ink on pulse oximetry.

Authors:  M F Battito
Journal:  Anesth Analg       Date:  1989-08       Impact factor: 5.108

4.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

5.  The penumbra effect: vasomotion-dependent pulse oximeter artifact due to probe malposition.

Authors:  J F Kelleher; R H Ruff
Journal:  Anesthesiology       Date:  1989-11       Impact factor: 7.892

6.  Accuracy of pulse oximetry in the intensive care unit.

Authors:  A Van de Louw; C Cracco; C Cerf; A Harf; P Duvaldestin; F Lemaire; L Brochard
Journal:  Intensive Care Med       Date:  2001-10       Impact factor: 17.440

Review 7.  Pulse oximetry. Gold standard or false sense of security?

Authors:  R F Grace
Journal:  Med J Aust       Date:  1994-05-16       Impact factor: 7.738

8.  Infrared heat lamps interfere with pulse oximeters.

Authors:  T D Brooks; D A Paulus; W E Winkle
Journal:  Anesthesiology       Date:  1984-11       Impact factor: 7.892

9.  Accuracy of pulse oximetry in children with cyanotic congenital heart disease.

Authors:  H J Schmitt; W H Schuetz; P A Proeschel; C Jaklin
Journal:  J Cardiothorac Vasc Anesth       Date:  1993-02       Impact factor: 2.628

10.  Accuracy of two pulse oximeters at low arterial hemoglobin-oxygen saturation.

Authors:  B G Carter; J B Carlin; J Tibballs; H Mead; M Hochmann; A Osborne
Journal:  Crit Care Med       Date:  1998-06       Impact factor: 7.598

View more
  3 in total

1.  Comments on 'evaluation of a pulse oximeter sensor tester'.

Authors:  G R Mathews; V M Hickson
Journal:  J Clin Monit Comput       Date:  2011-08-28       Impact factor: 2.502

2.  Comments on 'evaluation of a pulse oximeter sensor tester': authors' response.

Authors:  Iljaz Hodzovic; Antony Robert Wilkes
Journal:  J Clin Monit Comput       Date:  2011-08-27       Impact factor: 2.502

3.  Design and construct an optical device to determine relative blood volume in patients undergoing hemodialysis.

Authors:  Banafshe Dormanesh; Shahnaz Tofangchiha; Vahid Abouei; Hani Sharifian
Journal:  Iran Red Crescent Med J       Date:  2014-04-05       Impact factor: 0.611

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.