On the dynamic range of different X-ray photon detectors in intra-oral radiography. A comparison of image quality in film, charge-coupled device and storage phosphor systems.

E Borg1, H G Gröndahl.   

Abstract

OBJECTIVE: To compare film for intra-oral radiography with two charge-coupled device (CCD) and one storage phosphor system for digital imaging in respect of subjective image quality, detectability of small mass differences and appearance of burn-out effects and blooming phenomena at various exposure times.
METHODS: Dried mandibles with teeth from different areas were radiographed at exposures covering a relative range from 1 to 100. Image quality was subjectively evaluated after image processing, when applicable, using a visual grading scale from 0 to 10. The number of visible holes in an aluminium block was used to measure the detectability of small mass differences. Burn-out effects and blooming were evaluated by measuring widths of roots and of aluminium and plastic cylinders.
RESULTS: Radiographs with the storage phosphor system achieved image quality scores similar to those of film but over a larger exposure range, while CCD images were rated lower and over a smaller range. All holes in the aluminium block were only detected with the storage phosphor system. While the widths of roots were strongly affected by sensor saturation in CCD images and by burn-out in film images, smaller effects were seen with the storage phosphor system. Similar results were obtained with aluminium and plastic cylinders.
CONCLUSIONS: Higher image quality was achieved over a much wider exposure range with the storage phosphor system than with either film or the CCD systems.

Entities:  

Mesh:

Year:  1996        PMID: 9446978     DOI: 10.1259/dmfr.25.2.9446978

Source DB:  PubMed          Journal:  Dentomaxillofac Radiol        ISSN: 0250-832X            Impact factor:   2.419


  12 in total

1.  Prospective study on the reproducibility of cephalometric landmarks on conventional and digital lateral headfilms.

Authors:  K Hagemann; D Vollmer; T Niegel; U Ehmer; I Reuter
Journal:  J Orofac Orthop       Date:  2000       Impact factor: 1.938

2.  Diagnostic accuracy of proximal enamel subsurface demineralization and its relationship with calcium loss and lesion depth.

Authors:  E Onem; B G Baksi; B H Sen; O Sögüt; A Mert
Journal:  Dentomaxillofac Radiol       Date:  2011-11-10       Impact factor: 2.419

3.  A comprehensive in vitro study of image accuracy and quality for periodontal diagnosis. Part 1: the influence of X-ray generator on periodontal measurements using conventional and digital receptors.

Authors:  Bart Vandenberghe; Livia Corpas; Hilde Bosmans; Jie Yang; Reinhilde Jacobs
Journal:  Clin Oral Investig       Date:  2010-05-05       Impact factor: 3.573

4.  Relationship between radiation dose reduction and image quality change in photostimulable phosphor luminescence X-ray imaging systems.

Authors:  T Sakurai; R Kawamata; Y Kozai; Y Kaku; K Nakamura; M Saito; H Wakao; I Kashima
Journal:  Dentomaxillofac Radiol       Date:  2010-05       Impact factor: 2.419

5.  Effect of alternative photostimulable phosphor plates erasing times on subjective digital image quality.

Authors:  D P Melo; A Dos Anjos Pontual; S M de Almeida; P S F Campos; M C Alves; G M Tosoni
Journal:  Dentomaxillofac Radiol       Date:  2010-01       Impact factor: 2.419

6.  The usage of digital radiography and cone beam computed tomography among Turkish dentists.

Authors:  Semanur Dölekoğlu; E Fişekçioğlu; M İlgüy; D İlgüy
Journal:  Dentomaxillofac Radiol       Date:  2011-09       Impact factor: 2.419

7.  Comparison of the delay in processing time and protective plastic cases in two phosphor plate systems.

Authors:  Ali Murat Aktan; Mehmet Ertuğrul Ciftçi; Faruk Akgünlü
Journal:  ScientificWorldJournal       Date:  2012-11-07

8.  Diagnostic accuracy of digital and conventional radiography in the detection of non-cavitated approximal dental caries.

Authors:  Farida Abesi; Alireza Mirshekar; Ehsan Moudi; Maryam Seyedmajidi; Sina Haghanifar; Nima Haghighat; Ali Bijani
Journal:  Iran J Radiol       Date:  2012-03-25       Impact factor: 0.212

9.  The influence of tube potential on periodontal bone level measurements and subjective image quality using a digital photostimulable storage phosphor sensor.

Authors:  Bart Vandenberghe; Reinhilde Jacobs
Journal:  J Oral Maxillofac Res       Date:  2010-04-01

10.  Digital radiographic evaluation of optical density of porcelains used for fabrication of inlay/onlay prosthesis.

Authors:  Júlio Cezar de Melo Castilho; Wilton Mitsunari Takeshita; Luis Roque de Araújo dos Santos; Luiz César de Moraes; Edmundo Medici Filho; Mari Eli Leonelli de Moraes
Journal:  J Appl Oral Sci       Date:  2006-01       Impact factor: 2.698

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.