Literature DB >> 9399454

Accuracy of using computed tomography to identify pedicle screw placement in cadaveric human lumbar spine.

J U Yoo1, A Ghanayem, C Petersilge, J Lewin.   

Abstract

STUDY
DESIGN: Utility of using computed tomography to predict pedicle screw misplacement.
OBJECTIVE: This study defines the sensitivity and specificity of predicting pedicle screw placement by experienced clinicians using a CT scan image. SUMMARY OF BACKGROUND DATA: In clinical and research settings, the method most commonly used to evaluate pedicle screws placement has been computed tomography. However, no current literature describes the accuracy of this method of evaluating screw placement.
METHOD: Cobalt-chrome and titanium alloy pedicle screws of identical size were placed in six cadaveric human lumbar spine. Wide laminectomy was performed to allow complete visualization of the pedicles. Three consecutive lumbar levels were instrumented in each spine, giving 36 pedicle screw placements to identify. The instrumented spines were imaged, and four orthopaedic spine surgeons and a musculoskeletal radiologist were asked to read the images to identify the accuracy of screw placement within the pedicles.
RESULTS: The sensitivity rate of identifying a misplaced screw was 67 +/- 6% for cobalt-chrome screws compared with 86 +/- 5% for titanium screws (P < 0.005). The specificity rates of radiographic diagnosis of misplaced pedicle screws were 66 +/- 10% for cobalt-chrome screws and 88 +/- 8% for titanium screws (P < 0.005). Similarly, a statistically significant difference was found in the sensitivity rates of identifying screws placed correctly in the pedicle: 70 +/- 10% for cobalt-chrome screws versus 89 +/- 8% for titanium screws (P < 0.005). Overall accuracy rates were 68 +/- 7% for cobalt chrome screws versus 87 +/- 3% for titanium screws (P < 0.002).
CONCLUSION: Reliance on the computed tomography scan data alone in determining accuracy of pedicle screws can lead to inaccuracies in both clinical and research conditions.

Entities:  

Mesh:

Substances:

Year:  1997        PMID: 9399454     DOI: 10.1097/00007632-199711150-00016

Source DB:  PubMed          Journal:  Spine (Phila Pa 1976)        ISSN: 0362-2436            Impact factor:   3.468


  8 in total

1.  [Accuracy of CT-based navitation of pedicle screws in the thoracic spine compared with conventional technique].

Authors:  K J Schnake; B König; U Berth; R J Schroeder; F Kandziora; U Stöckle; M Raschke; N P Haas
Journal:  Unfallchirurg       Date:  2004-02       Impact factor: 1.000

2.  Pedicle Screw Placement in the Thoracolumbar Spine Using a Novel, Simple, Safe, and Effective Guide-Pin : A Computerized Tomography Analysis.

Authors:  Seung-Jae Hyun; Yongjung J Kim; Seung-Chul Rhim; Gene Cheh; Samuel K Cho
Journal:  J Korean Neurosurg Soc       Date:  2015-07-31

3.  Assessment of CAOS as a training model in spinal surgery: a randomised study.

Authors:  P J Richards; I C Kurta; V Jasani; C H Wynn Jones; A Rahmatalla; G Mackenzie; J Dove
Journal:  Eur Spine J       Date:  2006-05-09       Impact factor: 3.134

4.  Accuracy of a new intraoperative cone beam CT imaging technique (Artis zeego II) compared to postoperative CT scan for assessment of pedicle screws placement and breaches detection.

Authors:  Virginie Cordemans; Ludovic Kaminski; Xavier Banse; Bernard G Francq; Olivier Cartiaux
Journal:  Eur Spine J       Date:  2017-05-20       Impact factor: 3.134

5.  Effect of brightness in the evaluation of lumbar pedicular screws position: clinical study.

Authors:  L M Romero-Muñoz; M Alfonso; C Villas; J L Zubieta
Journal:  Musculoskelet Surg       Date:  2013-04-05

6.  Known-component 3D-2D registration for quality assurance of spine surgery pedicle screw placement.

Authors:  A Uneri; T De Silva; J W Stayman; G Kleinszig; S Vogt; A J Khanna; Z L Gokaslan; J-P Wolinsky; J H Siewerdsen
Journal:  Phys Med Biol       Date:  2015-09-30       Impact factor: 3.609

7.  Inter- and intra-observer reliability of measurement of pedicle screw breach assessed by postoperative CT scans.

Authors:  William F Lavelle; Ashish Ranade; Amer F Samdani; John P Gaughan; Linda P D'Andrea; Randal R Betz
Journal:  Int J Spine Surg       Date:  2014-12-01

8.  Lumbar pedicle screw placement: Using only AP plane imaging.

Authors:  Anil Sethi; Adrienne Lee; Rahul Vaidya
Journal:  Indian J Orthop       Date:  2012-07       Impact factor: 1.251