Performance of the automated COBAS AMPLICOR system for the detection of hepatitis C virus RNA.

H H Kessler1, E A Dragon, K Pierer, B I Santner, Y Liao, D Stünzner, E Stelzl, E Marth.   

Abstract

BACKGROUND: The COBAS AMPLICOR (CA) instrument for the amplification and detection steps of the AMPLICOR molecular diagnostic assays has recently been introduced. The system contains a single thermal cycler with two independently controlled heating/cooling blocks, a pipettor, a magnetic particle washer, a photometer and an incubator.
OBJECTIVE: The performance of the CA instrument was evaluated in a routine diagnostic laboratory for the detection of hepatitis C virus (HCV) RNA. The new system was compared with the corresponding microwell plate assay (AMPLICOR HCV Test). STUDY
DESIGN: Routine clinical sera (350) from hemodialysis patients and patients with chronic HCV infection and interferon therapy were studied. If discrepant results were obtained, both assays were repeated (specimen preparation, amplification and detection); in addition, the HCV copy number was determined with the AMPLICOR HCV MONITOR Test.
RESULTS: There was a correlation between the CA HCV Test and the AMPLICOR HCV Test in 341 of 350 specimens (97%). After resolution of 9 discrepant results, the CA HCV Test gave a sensitivity of 97.8% and a specificity of 99.4%. The most common reason for discrepant results was a low HCV RNA copy number.
CONCLUSION: The CA system was found to be a labor-saving, fast and reliable instrument for the amplification and detection steps of a RT-PCR molecular assay for detection of HCV RNA.

Entities:  

Mesh:

Substances:

Year:  1997        PMID: 9126682     DOI: 10.1016/s0928-0197(96)00263-2

Source DB:  PubMed          Journal:  Clin Diagn Virol        ISSN: 0928-0197


  5 in total

1.  Effects of storage and type of blood collection tubes on hepatitis C virus level in whole blood samples.

Authors:  H H Kessler; E Stelzl; R B Raggam; J Haas; F Kirchmeir; K Hegenbarth; E Daghofer; B I Santner; E Marth; R E Stauber
Journal:  J Clin Microbiol       Date:  2001-05       Impact factor: 5.948

2.  Evaluation of AMPLILINK software for the COBAS AMPLICOR system.

Authors:  H H Kessler; D Jungkind; E Stelzl; S Direnzo; S K Vellimedu; K Pierer; B Santner; E Marth
Journal:  J Clin Microbiol       Date:  1999-02       Impact factor: 5.948

3.  Fully automated detection of hepatitis C virus RNA in serum and whole-blood samples.

Authors:  Harald H Kessler; Alexandra M K Clarici; Evelyn Stelzl; Gerhard Mühlbauer; Elisabeth Daghofer; Brigitte I Santner; Egon Marth; Rudolf E Stauber
Journal:  Clin Diagn Lab Immunol       Date:  2002-11

4.  Semiautomated quantification of hepatitis B virus DNA in a routine diagnostic laboratory.

Authors:  H H Kessler; E Stelzl; E Daghofer; B I Santner; E Marth; H Lackner; R E Stauber
Journal:  Clin Diagn Lab Immunol       Date:  2000-09

5.  Evaluation of an automated sample preparation protocol for quantitative detection of hepatitis C virus RNA.

Authors:  Evelyn Stelzl; Andrea Kormann-Klement; Josef Haas; Elisabeth Daghofer; Brigitte I Santner; Egon Marth; Harald H Kessler
Journal:  J Clin Microbiol       Date:  2002-04       Impact factor: 5.948

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.