| Literature DB >> 9081259 |
J M Wittenberg1, D P Mukherjee, B R Smith, R N Kruse.
Abstract
This biomechanical study investigated the effectiveness of fixation devices of simulated angle fractures in sheep mandibles. The fractures were stabilized by a Leibinger 8-hole three-dimensional (3-D) plate, a Synthes eight-hole mesh plate, and a Synthes six-hole reconstruction plate with 2.0-mm and 2.4-mm mono- and bicortical screws. Each mandible was tested in bending (class III cantilever model). The bone mineral density of the mandibles was measured by computed tomography scan. The Leibinger 3-D plate showed plate deformation in bending of > 230 N. None of the plates showed failure in the bone/screw interface. The gap and displacement values for the mesh and 3-D plate were comparable to those of the reconstruction plate. These results indicate that a 3-D or mesh plate can be used for fixation of mandibular angle fractures.Entities:
Mesh:
Year: 1997 PMID: 9081259 DOI: 10.1016/s0901-5027(97)80852-1
Source DB: PubMed Journal: Int J Oral Maxillofac Surg ISSN: 0901-5027 Impact factor: 2.789