Literature DB >> 9055375

Accuracy of computer-assisted pedicle screw placement. An in vivo computed tomography analysis.

O Schwarzenbach1, U Berlemann, B Jost, H Visarius, E Arm, F Langlotz, L P Nolte, C Ozdoba.   

Abstract

STUDY
DESIGN: A computer-assisted planning and visualization system (the Orthopaedic Surgery Planning System) was tested for pedicle screw insertion in vivo.
OBJECTIVES: To evaluate the system's applicability for regular intraoperative use and its accuracy for pedicle screw placement in vivo. SUMMARY OF BACKGROUND DATA: Pedicle screw placement poses considerable anatomic and biomechanical risks. The reported rates of screw misplacement with conventional insertion techniques are unacceptably high. It previously has been shown in vitro that computer assistance offers the potential to decrease the number of screws perforating the pedicular cortex.
METHODS: The accuracy of 162 pedicle screws inserted with the Orthopaedic Surgery Planning System was assessed by means of postoperative computed tomography evaluation. Reconstructions of the horizontal, frontal, and sagittal planes were analyzed. Cortex perforations were graded in steps of 2 mm.
RESULTS: The cortex was perforated in 2.7% of pedicles. Complete preoperative computed tomography scanning of the levels to be operated on is essential to allow for a precise image reconstruction. Initial difficulties in applying the system contribute to the malplacements. A learning curve for general handling of the Orthopaedic Surgery Planning System was observed.
CONCLUSIONS: The system provides a safe and reproducible technique for pedicle screw insertion. Other applications in the field of spine surgery are under evaluation.

Mesh:

Year:  1997        PMID: 9055375     DOI: 10.1097/00007632-199702150-00020

Source DB:  PubMed          Journal:  Spine (Phila Pa 1976)        ISSN: 0362-2436            Impact factor:   3.468


  51 in total

1.  Is intraoperative CT of posterior cervical spine instrumentation cost-effective and does it reduce complications?

Authors:  Andrew C Hecht; Steven M Koehler; Janelle C Laudone; Arthur Jenkins; Sheeraz Qureshi
Journal:  Clin Orthop Relat Res       Date:  2011-04       Impact factor: 4.176

2.  [Iso-C(3D0-assisted) navigated implantation of pedicle screws in thoracic lumbar vertebrae].

Authors:  K Wendl; J von Recum; A Wentzensen; P A Grützner
Journal:  Unfallchirurg       Date:  2003-11       Impact factor: 1.000

3.  [Precision of navigation-assisted surgery of the thoracic and lumbar spine].

Authors:  M Arand; M Schempf; D Hebold; S Teller; L Kinzl; F Gebhard
Journal:  Unfallchirurg       Date:  2003-11       Impact factor: 1.000

4.  Computer-assisted posterior instrumentation of the cervical and cervico-thoracic spine.

Authors:  Marcus Richter; Thomas Mattes; Balkan Cakir
Journal:  Eur Spine J       Date:  2003-11-22       Impact factor: 3.134

5.  [Intraoperative three-dimensional navigation for pedicle screw placement].

Authors:  P A Grützner; T Beutler; K Wendl; J von Recum; A Wentzensen; L-P Nolte
Journal:  Chirurg       Date:  2004-10       Impact factor: 0.955

6.  [Accuracy of CT-based navitation of pedicle screws in the thoracic spine compared with conventional technique].

Authors:  K J Schnake; B König; U Berth; R J Schroeder; F Kandziora; U Stöckle; M Raschke; N P Haas
Journal:  Unfallchirurg       Date:  2004-02       Impact factor: 1.000

7.  [Computer-assisted screw placement into the posterior pelvic ring: assessment of different navigated procedures in a cadaver trial].

Authors:  D Briem; J M Rueger; P G C Begemann; Z Halata; T Bock; W Linhart; J Windolf
Journal:  Unfallchirurg       Date:  2006-08       Impact factor: 1.000

8.  Assessment of CAOS as a training model in spinal surgery: a randomised study.

Authors:  P J Richards; I C Kurta; V Jasani; C H Wynn Jones; A Rahmatalla; G Mackenzie; J Dove
Journal:  Eur Spine J       Date:  2006-05-09       Impact factor: 3.134

9.  Observer reliability in evaluating pedicle screw placement using computed tomography.

Authors:  Victor Kosmopoulos; Nicolas Theumann; Stefano Binaghi; Constantin Schizas
Journal:  Int Orthop       Date:  2006-09-12       Impact factor: 3.075

10.  [Navigated drilling for femoral head necrosis. Experimental and clinical results].

Authors:  J Beckmann; M Tingart; L Perlick; C Lüring; J Grifka; S Anders
Journal:  Orthopade       Date:  2007-05       Impact factor: 1.087

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.