The effects of motion on the performance of pulse oximeters in volunteers (revised publication).

S J Barker1, N K Shah.   

Abstract

BACKGROUND: Pulse oximetry is considered a standard of care in both the operating room and the postanesthetic care unit, and it is widely used in all critical care settings. Pulse oximeters may fail to provide valid SpO2 data in various situations that produce low signal-to-noise ratio. Motion artifact is a common cause of oximeter failure and loss of accuracy. This study compares the accuracy and data dropout rates of three current pulse oximeters during standardized motion in healthy volunteers.
METHODS: Ten healthy volunteers were monitored by three different pulse oximeters: Nellcor N-200, Nellcor N-3000, and Masimo SET (prototype). Sensors were placed on digits 2, 3, and 4 of the test hand, which was strapped to a mechanical motion table. The opposite hand was used as a stationary control and was monitored with the same pulse oximeters and an arterial cannula. Arterial oxygen saturation was varied from 100% to 75% by changing the inspired oxygen concentration. While SpO2 was both constant and changing, the oximeter sensors were connected before and during motion. Oximeter errors and dropout rates were digitally recorded continuously during each experiment.
RESULTS: If the oximeter was functioning before motion began, the following are the percentages of time when the instrument displayed an SpO2 value within 7% of control: N-200 = 76%, N-3000 = 87%, and Masimo = 99%. When the oximeter sensor was connected after the beginning of motion, the values were N-200 = 68%, N-3000 = 47%, and Masimo = 97%. If the alarm threshold was chosen SpO2 less than 90%, then the positive predictive values (true alarms/total alarms) are N-200 = 73%, N-3000 = 81%, and Masimo = 100%. In general, N-200 had the greatest SpO2 errors and N-3000 had the highest dropout rates.
CONCLUSIONS: The mechanical motions used in this study significantly affected oximeter function, particularly when the sensors were connected during motion, which requires signal acquisition during motion. The error and dropout rate performance of the Masimo was superior to that of the other two instruments during all test conditions. Masimo uses a new paradigm for oximeter signal processing, which appears to represent a significant advance in low signal-to-noise performance.

Entities:  

Mesh:

Year:  1997        PMID: 9009945     DOI: 10.1097/00000542-199701000-00014

Source DB:  PubMed          Journal:  Anesthesiology        ISSN: 0003-3022            Impact factor:   7.892


  22 in total

1.  A second generation technique for evaluating accuracy and reliability of second generation pulse oximeters.

Authors:  K K Tremper
Journal:  J Clin Monit Comput       Date:  2000       Impact factor: 2.502

2.  Computerised audiovisual event recording for infant apnoea and bradycardia.

Authors:  R T Brouillette; D Tsirigotis; A Leimanis; A Côté; A Morielli
Journal:  Med Biol Eng Comput       Date:  2000-09       Impact factor: 2.602

3.  Major reduction in alarm frequency with a new pulse oximeter.

Authors:  B Bohnhorst; C F Poets
Journal:  Intensive Care Med       Date:  1998-03       Impact factor: 17.440

4.  Comparison between a single-channel nasal airflow device and oximetry for the diagnosis of obstructive sleep apnea.

Authors:  Lydia Makarie Rofail; Keith K H Wong; Gunnar Unger; Guy B Marks; Ronald R Grunstein
Journal:  Sleep       Date:  2010-08       Impact factor: 5.849

5.  Measurement of blood concentration of indocyanine green by pulse dye densitometry--comparison with the conventional spectrophotometric method.

Authors:  T Imai; K Takahashi; F Goto; Y Morishita
Journal:  J Clin Monit Comput       Date:  1998-12       Impact factor: 2.502

6.  A prototype device for standardized calibration of pulse oximeters II.

Authors:  Ch Hornberger; Ph Knoop; H Matz; F Dörries; E Konecny; H Gehring; J Otten; R Bonk; H Frankenberger; P Wouters; J Gil-Rodriguez; L Ponz; J Avgerinos; A Karoutis; A Ikiades; S Weininger
Journal:  J Clin Monit Comput       Date:  2002 Apr-May       Impact factor: 2.502

7.  Masimo signal extraction pulse oximetry.

Authors:  J M Goldman; M T Petterson; R J Kopotic; S J Barker
Journal:  J Clin Monit Comput       Date:  2000       Impact factor: 2.502

8.  Determining the artifact sensitivity of recent pulse oximeters during laboratory benchmarking.

Authors:  S W Kästle; E Konecny
Journal:  J Clin Monit Comput       Date:  2000       Impact factor: 2.502

9.  Forehead reflectance oximetry: a clinical comparison with conventional digit sensors during laparotomic and laparoscopic abdominal surgery.

Authors:  Andrea Casati; Grazia Squicciarini; Marco Baciarello; Marta Putzu; Alessandra Salvadori; Guido Fanelli
Journal:  J Clin Monit Comput       Date:  2007-08-16       Impact factor: 2.502

10.  Motion resistant pulse oximetry in neonates.

Authors:  R Sahni; A Gupta; K Ohira-Kist; T S Rosen
Journal:  Arch Dis Child Fetal Neonatal Ed       Date:  2003-11       Impact factor: 5.747

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.