Measurement of blood concentration of indocyanine green by pulse dye densitometry--comparison with the conventional spectrophotometric method.

T Imai1, K Takahashi, F Goto, Y Morishita.   

Abstract

OBJECTIVE: Pulse dye densitometry (PDD) uses two wavelengths (805 and 890 nm) in association with pulse oximetry to compute the arterial blood concentration ratio of indocyanine green (ICG) to hemoglobin (Hb). When Hb is measured in the usual way, this permits the PDD to compute cardiac output, plasma or blood volume, and liver blood flow following an intravenous injection of ICG. In this study, we evaluate the accuracy of the PDD calculation of dye concentration by comparing it with measurement of the dye concentration in blood (Cb) measured by the spectrophotometric cuvette method during dye clearance in patients.
METHODS: In 25 patients receiving major abdominal surgery, ICG (10, 20, or 40 mg) was injected into a central vein and arterial ICG concentration was continuously and simultaneously monitored at nose and finger by PDD; concurrently, ICG concentrations were measured by a spectrophotometer at 805 nm in 4 radial arterial blood samples. Repeated measures or one-way ANOVA were used for comparison of ICG concentrations and percent errors by PDD at the nose, finger, and Cb.
RESULTS: The percent error (bias) of calculated dye concentration and its standard deviation (precision) was -3.9 +/- 16.8% (p < 0.01) with the probe on a nostril and 3.4 +/- 12.6% using the finger probe. These errors were found to be greatest when the mean transit time of the dye was rapid (-20.7 +/- 6.8% at nose p < 0.01 and -8.5 +/- 2.5% at finger p < 0.05) due to factors other than the time delay of blood sampling.
CONCLUSION: These errors are of similar size to those associated with thermal cardiac output measurement, suggesting that PDD should be valuable clinically as a noninvasive tool especially since it provides values for blood volume and liver blood flow.

Entities:  

Mesh:

Substances:

Year:  1998        PMID: 10385856     DOI: 10.1023/a:1009948128543

Source DB:  PubMed          Journal:  J Clin Monit Comput        ISSN: 1387-1307            Impact factor:   2.502


  5 in total

Review 1.  Recent developments in pulse oximetry.

Authors:  J W Severinghaus; J F Kelleher
Journal:  Anesthesiology       Date:  1992-06       Impact factor: 7.892

2.  The effects of motion on the performance of pulse oximeters in volunteers (revised publication).

Authors:  S J Barker; N K Shah
Journal:  Anesthesiology       Date:  1997-01       Impact factor: 7.892

3.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

4.  Measurement of cardiac output by pulse dye densitometry using indocyanine green: a comparison with the thermodilution method.

Authors:  T Imai; K Takahashi; H Fukura; Y Morishita
Journal:  Anesthesiology       Date:  1997-10       Impact factor: 7.892

5.  Cardiac output and circulating blood volume analysis by pulse dye-densitometry.

Authors:  T Iijima; T Aoyagi; Y Iwao; J Masuda; M Fuse; N Kobayashi; H Sankawa
Journal:  J Clin Monit       Date:  1997-03
  5 in total
  5 in total

Review 1.  [Determinants of insensible fluid loss. Perspiration, protein shift and endothelial glycocalyx].

Authors:  M Jacob; D Chappell; K Hofmann-Kiefer; P Conzen; K Peter; M Rehm
Journal:  Anaesthesist       Date:  2007-08       Impact factor: 1.041

2.  Intraoperative ICG plasma disappearance rate helps to predict absence of early postoperative complications after orthotopic liver transplantation.

Authors:  J J Vos; T W L Scheeren; D J Lukes; M T de Boer; H G D Hendriks; J K G Wietasch
Journal:  J Clin Monit Comput       Date:  2013-05-16       Impact factor: 2.502

3.  Fluid balance and blood volume measurement after aneurysmal subarachnoid hemorrhage.

Authors:  Reinier G Hoff; Gert W van Dijk; Ale Algra; Cor J Kalkman; Gabriel J E Rinkel
Journal:  Neurocrit Care       Date:  2008       Impact factor: 3.210

4.  Comparison between preoperative two-dimensional shear wave elastography and indocyanine green clearance test for prediction of post-hepatectomy liver failure.

Authors:  Tingting Qiu; Rong Fu; Wenwu Ling; Jiawu Li; Jiulin Song; Zhenru Wu; Yujun Shi; Yuqing Zhou; Yan Luo
Journal:  Quant Imaging Med Surg       Date:  2021-05

5.  Quantifying the Photochemical Damage Potential of Contrast-Enhanced Fluorescence Imaging Products: Singlet Oxygen Production.

Authors:  Huang-Chiao Huang; T Joshua Pfefer; Brandon Gaitan; Lucas Frankle; Shruti Vig; Ellen Oskoui; Miriam Adwan; Yu Chen; Rosalie Elespuru
Journal:  Photochem Photobiol       Date:  2022-05-23       Impact factor: 3.521

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.