Accuracy of oscillometric blood pressure measurement in critically ill neonates with reference to the arterial pressure wave shape.

M Gevers1, H R van Genderingen, H N Lafeber, W W Hack.   

Abstract

OBJECTIVE: To perform further evaluation of the oscillometric device for neonatal arterial blood pressure (ABP) measurement, using a catheter-manometer system (CMS) for accurate intraarterial measurement. We aimed to describe the influence of the radial artery wave shape on oscillometric ABP determination, as pressure wave-shape influences the relationships between systolic arterial pressure (SAP), diastolic arterial pressure (DAP) and mean arterial pressure (MAP) in the wave. These relationships are part of the algorithms contributing to the final ABP determination in the oscillometric device.
DESIGN: Intra-patient comparison of two blood pressure measurement systems.
SETTING: Neonatal intensive care unit. PATIENTS: In 51 critically ill newborn infants, ABP was determined oscillometrically in the brachial artery and, simultaneously, invasively in the radial artery using a high-fidelity CMS. Clinical data of the infants were: gestational age: 29 (25-41) weeks; birthweight: 1200 (500-3675) g, postnatal age: 6 (2-46) h.
METHODS: Statistical analysis was performed with the paired Student's t-test. Multiple regression analysis was used to determine the influence of birthweight and height of the blood pressure on the results.
MEASUREMENTS AND MAIN RESULTS: In 51 infants, 255 paired values of SAP, DAP and MAP were recorded. In all recordings, we determined the relationship between SAP, DAP and MAP, using the equation: MAP = alpha%(SAP - DAP) + DAP. For SAP, DAP, MAP and alpha, we computed mean differences (bias) and the limits of agreement (precision). Biases for SAP, DAP, MAP and alpha were significantly different from zero (P < 0.001) and the limits of agreement for SAP, DAP and MAP were wide: 18.8 mmHg, 17.2 mmHg and 15.2 mmHg respectively. The relationship between invasive and noninvasive values is only partly (7-19%) influenced by the height of the blood pressure; low values of SAP, DAP and MAP tend to give overestimated oscillometric values. In the relationship between SAP, DAP and MAP, alpha was found to be 47% invasively (as generally found in the radial artery in newborns) and 34% noninvasively (as generally found in the brachial/radial artery in adults).
CONCLUSIONS: Inaccuracy of the oscillometric device may be partly explained by the incorporation of an inappropriately fixed algorithm for final ABP determination in newborns. Care should be taken when interpreting the oscillometrically derived values in critically ill newborn infants.

Entities:  

Mesh:

Year:  1996        PMID: 8727439     DOI: 10.1007/BF01712244

Source DB:  PubMed          Journal:  Intensive Care Med        ISSN: 0342-4642            Impact factor:   17.440


  31 in total

1.  Blood pressure monitoring: automated oscillometric devices.

Authors:  M Ramsey
Journal:  J Clin Monit       Date:  1991-01

2.  Experimental analysis of catheter-manometer systems in vitro and in vivo.

Authors:  S F Hipkins; A J Rutten; W B Runciman
Journal:  Anesthesiology       Date:  1989-12       Impact factor: 7.892

3.  Oscillometric determination of diastolic, mean and systolic blood pressure--a numerical model.

Authors:  F K Forster; D Turney
Journal:  J Biomech Eng       Date:  1986-11       Impact factor: 2.097

4.  Prediction of arterial blood pressure in the premature neonate using the oscillometric method.

Authors:  J A Wareham; L D Haugh; S B Yeager; J D Horbar
Journal:  Am J Dis Child       Date:  1987-10

5.  A comparative study of the measurement of mean arterial blood pressure using automatic oscillometers, arterial cannulation and auscultation.

Authors:  A J Rutten; A H Ilsley; G A Skowronski; W B Runciman
Journal:  Anaesth Intensive Care       Date:  1986-02       Impact factor: 1.669

6.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

7.  Calculated mean arterial blood pressure in critically ill neonates.

Authors:  M Gevers; W W Hack; E F Ree; H N Lafeber; N Westerhof
Journal:  Basic Res Cardiol       Date:  1993 Jan-Feb       Impact factor: 17.165

8.  Indirect measurement of blood pressure in neonates and infants utilizing an automatic noninvasive oscillometric monitor.

Authors:  R H Friesen; J L Lichtor
Journal:  Anesth Analg       Date:  1981-10       Impact factor: 5.108

9.  Dinamap fails to detect hypotension in very low birthweight infants.

Authors:  G K Diprose; D H Evans; L N Archer; M I Levene
Journal:  Arch Dis Child       Date:  1986-08       Impact factor: 3.791

10.  Accuracy of blood pressure measurement by the Dinamap monitor in infants and children.

Authors:  M K Park; S M Menard
Journal:  Pediatrics       Date:  1987-06       Impact factor: 7.124

View more
  6 in total

1.  The cardiac blood supply-workload balance in children.

Authors:  Tomoaki Murakami; Atsuhito Takeda; Kohta Takei; Shigeru Tateno; Yasutaka Kawasoe; Koichiro Niwa
Journal:  Heart Vessels       Date:  2014-07-03       Impact factor: 2.037

2.  Influence of body height on aortic systolic pressure augmentation and wave reflection in childhood.

Authors:  E V Hidvégi; M Illyés; F T Molnár; A Cziráki
Journal:  J Hum Hypertens       Date:  2014-12-11       Impact factor: 3.012

Review 3.  Perioperative central nervous system injury in neonates.

Authors:  M E McCann; S G Soriano
Journal:  Br J Anaesth       Date:  2012-12       Impact factor: 9.166

4.  Comparison of oscillometric blood pressure measurement by two clinical monitors.

Authors:  Gp Prashanth
Journal:  Indian J Anaesth       Date:  2011-09

5.  Hemodynamic Monitoring in the Acute Management of Pediatric Heart Failure.

Authors:  Rocky Tsang; Paul Checchia; Ronald A Bronicki
Journal:  Curr Cardiol Rev       Date:  2016

Review 6.  Continuous vital sign analysis for predicting and preventing neonatal diseases in the twenty-first century: big data to the forefront.

Authors:  Navin Kumar; Gangaram Akangire; Brynne Sullivan; Karen Fairchild; Venkatesh Sampath
Journal:  Pediatr Res       Date:  2019-08-04       Impact factor: 3.756

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.