Literature DB >> 8567877

Controlled clinical comparison of Isolator and BACTEC 9240 Aerobic/F resin bottle for detection of bloodstream infections.

J K Pohlman1, B A Kirkley, K A Easley, J A Washington.   

Abstract

A controlled clinical comparison was carried out with the BACTEC 9240 Aerobic/F resin bottle and the Isolator system with adult patients suspected of having bloodstream infections. A total of 10,500 paired specimens were collected, of which 1,122 from 520 patients were positive. There were 68 false-positive BACTEC bottles; 259 positive cultures that were excluded from analysis because the bottle, the Isolator, or both failed to meet the minimum volume criterion of 8 ml of blood; and 207 positive cultures that were excluded because the isolates were found to be clinically insignificant or of indeterminate clinical significance on the basis of patient assessment. A total of 656 positive cultures from 258 patients formed the basis of the analysis. Significantly more Staphylococcus aureus isolates (P = 0.03), Staphylococcus epidermidis isolates (P = 0.03), members of the family Enterobacteriaceae (P = 0.03), and Pseudomonas aeruginosa isolates (P = 0.04) were recovered from the resin bottle, and there was no category of organism that was recovered significantly more frequently from the Isolator system. With patients receiving antibiotics at the time of blood culture, S. aureus, S. epidermidis, and gram-negative bacilli were recovered significantly more frequently from the resin bottle. No significant differences between systems were found with cultures from patients not receiving antibiotics at the time of blood culture. Only 12 clinically significant organisms were recovered from the bottle on terminal subcultures, and only 1 of these had not been previously isolated from another blood culture set (10 of the 12) or from the companion Isolator (1 of 12). The Aerobic/F resin bottle continuously monitored in the BACTEC 9240 instrument proved to be superior to the Isolator in overall yield of organisms causing bloodstream infection in adults and required less technician time for specimen processing and examination than the Isolator system.

Entities:  

Mesh:

Substances:

Year:  1995        PMID: 8567877      PMCID: PMC228522          DOI: 10.1128/jcm.33.10.2525-2529.1995

Source DB:  PubMed          Journal:  J Clin Microbiol        ISSN: 0095-1137            Impact factor:   5.948


  11 in total

1.  Staphylococcus aureus bacteremia. Potential impact of a new blood culture technique.

Authors:  J N Sheagren
Journal:  Arch Intern Med       Date:  1986-12

2.  Clinical comparison of lysis-centrifugation and radiometric resin systems for blood culture.

Authors:  P Brannon; T E Kiehn
Journal:  J Clin Microbiol       Date:  1986-11       Impact factor: 5.948

3.  Clinical comparison of isolator and BACTEC 660 resin media for blood culture.

Authors:  M T Kelly; F J Roberts; D Henry; I Geere; J A Smith
Journal:  J Clin Microbiol       Date:  1990-09       Impact factor: 5.948

4.  Multicenter clinical evaluation of a continuous monitoring blood culture system using fluorescent-sensor technology (BACTEC 9240).

Authors:  F S Nolte; J M Williams; R C Jerris; J A Morello; C D Leitch; S Matushek; L D Schwabe; F Dorigan; F E Kocka
Journal:  J Clin Microbiol       Date:  1993-03       Impact factor: 5.948

5.  CDC definitions for nosocomial infections, 1988.

Authors:  J S Garner; W R Jarvis; T G Emori; T C Horan; J M Hughes
Journal:  Am J Infect Control       Date:  1988-06       Impact factor: 2.918

6.  Lysis-centrifugation blood culture technique. Clinical impact in Staphylococcus aureus bacteremia.

Authors:  R C Walker; N K Henry; J A Washington; R L Thompson
Journal:  Arch Intern Med       Date:  1986-12

7.  Clinical comparison of the resin-containing BACTEC 26 Plus and the Isolator 10 blood culturing systems.

Authors:  J J Tarrand; C Guillot; M Wenglar; J Jackson; J D Lajeunesse; K V Rolston
Journal:  J Clin Microbiol       Date:  1991-10       Impact factor: 5.948

8.  Controlled comparison of the BACTEC high-blood-volume fungal medium, BACTEC Plus 26 aerobic blood culture bottle, and 10-milliliter isolator blood culture system for detection of fungemia and bacteremia.

Authors:  M L Wilson; T E Davis; S Mirrett; J Reynolds; D Fuller; S D Allen; K K Flint; F Koontz; L B Reller
Journal:  J Clin Microbiol       Date:  1993-04       Impact factor: 5.948

9.  Microplate-based DNA hybridization assays for detection of human retroviral gene sequences.

Authors:  L M Dyster; L Abbott; V Bryz-Gornia; B J Poiesz; L D Papsidero
Journal:  J Clin Microbiol       Date:  1994-02       Impact factor: 5.948

10.  Microbiological and clinical evaluation of the isolator lysis-centrifugation blood culture tube.

Authors:  N K Henry; C A McLimans; A J Wright; R L Thompson; W R Wilson; J A Washington
Journal:  J Clin Microbiol       Date:  1983-05       Impact factor: 5.948

View more
  15 in total

1.  Performance of five agar media for recovery of fungi from isolator blood cultures.

Authors:  G W Procop; F R Cockerill; E A Vetter; W S Harmsen; J G Hughes; G D Roberts
Journal:  J Clin Microbiol       Date:  2000-10       Impact factor: 5.948

2.  Evaluation of a rapid direct assay for identification of bacteria and the mec A and van genes from positive-testing blood cultures.

Authors:  U Eigner; M Weizenegger; A-M Fahr; W Witte
Journal:  J Clin Microbiol       Date:  2005-10       Impact factor: 5.948

3.  Performance of a new combination of blood culture vials in sepsis detection: a 2-year retrospective comparison.

Authors:  Paolo Bottino; Fabio Rapallo; Elisa Gamalero; Andrea Rocchetti
Journal:  Eur J Clin Microbiol Infect Dis       Date:  2019-05-08       Impact factor: 3.267

4.  Advantage of combining resin with lytic BACTEC blood culture media.

Authors:  P Rohner; B Pepey; R Auckenthaler
Journal:  J Clin Microbiol       Date:  1997-10       Impact factor: 5.948

5.  Controlled comparative evaluation of BacT/Alert FAN and ESP 80A aerobic media as means for detecting bacteremia and fungemia.

Authors:  G V Doern; A Barton; S Rao
Journal:  J Clin Microbiol       Date:  1998-09       Impact factor: 5.948

6.  Comparison of five days versus seven days of incubation for detection of positive blood cultures by the Bactec 9240 system.

Authors:  A H Huang; J J Yan; J J Wu
Journal:  Eur J Clin Microbiol Infect Dis       Date:  1998-09       Impact factor: 3.267

7.  Use of simulated blood cultures to compare a specific fungal medium with a standard microorganism medium for yeast detection.

Authors:  H Fricker-Hidalgo; F Chazot; B Lebeau; H Pelloux; P Ambroise-Thomas; R Grillot
Journal:  Eur J Clin Microbiol Infect Dis       Date:  1998-02       Impact factor: 3.267

8.  Multicenter comparison of BACTEC 9050 and BACTEC 9240 blood culture systems.

Authors:  P R Murray; G E Hollick; R C Jerris; M L Wilson
Journal:  J Clin Microbiol       Date:  1998-06       Impact factor: 5.948

9.  Effectiveness of resins in neutralizing antibiotic activities in bactec plus Aerobic/F culture medium.

Authors:  J Spaargaren; C P van Boven; G P Voorn
Journal:  J Clin Microbiol       Date:  1998-12       Impact factor: 5.948

10.  Effects of rapid detection of bloodstream infections on length of hospitalization and hospital charges.

Authors:  S E Beekmann; D J Diekema; K C Chapin; G V Doern
Journal:  J Clin Microbiol       Date:  2003-07       Impact factor: 5.948

View more