Comparison of two reflectance photometers in the assessment of neonatal hypoglycaemia.

P Kirkham1, A Watkins.   

Abstract

To assess the accuracy and reliability of reflectance photometers in estimating blood glucose concentrations, two were assessed: the Ames Glucometer 3 (Bayer Diagnostics) with Glucofilm Test strips; and the Reflolux S (Boehringer Mannheim) with BM Test Glycemia 20-800 strips. These were compared with laboratory blood glucose estimations in 100 assays (50 comparisons for each machine, measuring the difference (d) between the glucose reading and the mean of the reflectance photometer and the laboratory value). The Ames Glucometer 3 (mean d = + 0.7 mmol/l, (SD 1.1) mmol/l) was less accurate than the Boehringer Reflolux S (mean d = 0.2 mmol/l, (SD (0.7) mmol/l). The range of error of both machines is wide (Ames 2 SD range + 2.9 mmol/l to -1.5 mmol/l true readings; Boehringer + 1.8 mmol/l to -1.2 mmol/l of true readings). Because of this, any reflectance photometer readings that are even slightly low should be checked with laboratory estimations. The clinical value of such machines is limited in infants with low blood glucose concentrations.

Entities:  

Mesh:

Substances:

Year:  1995        PMID: 8535875      PMCID: PMC2528469          DOI: 10.1136/fn.73.3.f170

Source DB:  PubMed          Journal:  Arch Dis Child Fetal Neonatal Ed        ISSN: 1359-2998            Impact factor:   5.747


  9 in total

1.  The effect of haematocrit on reagent strip tests for glucose.

Authors:  K Wiener
Journal:  Diabet Med       Date:  1991 Feb-Mar       Impact factor: 4.359

2.  Clinical application of a new glucose analyzer in the neonatal intensive care unit: comparison with other methods.

Authors:  P D Conrad; J W Sparks; I Osberg; L Abrams; W W Hay
Journal:  J Pediatr       Date:  1989-02       Impact factor: 4.406

3.  Adverse neurodevelopmental outcome of moderate neonatal hypoglycaemia.

Authors:  A Lucas; R Morley; T J Cole
Journal:  BMJ       Date:  1988-11-19

4.  Capillary blood glucose monitor. Evaluation in a newborn nursery.

Authors:  B Presti; T Kircher; C Reed
Journal:  Clin Pediatr (Phila)       Date:  1989-09       Impact factor: 1.168

5.  Plasma or whole blood glucose?

Authors:  B Morrison; A Fleck
Journal:  Clin Chim Acta       Date:  1973-05-18       Impact factor: 3.786

6.  Incidence of hypoglycemia in newborn infants classified by birth weight and gestational age.

Authors:  L O Lubchenco; H Bard
Journal:  Pediatrics       Date:  1971-05       Impact factor: 7.124

7.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

Review 8.  Prevention and management of neonatal hypoglycaemia.

Authors:  J M Hawdon; M P Ward Platt; A Aynsley-Green
Journal:  Arch Dis Child Fetal Neonatal Ed       Date:  1994-01       Impact factor: 5.747

9.  Screening for hypoglycemia with plasma in neonatal blood of high hematocrit value.

Authors:  M Kaplan; O Blondheim; I Alon; U Eylath; S Trestian; A I Eidelman
Journal:  Crit Care Med       Date:  1989-03       Impact factor: 7.598

  9 in total
  3 in total

1.  Hypoglycemia in small for gestational age babies.

Authors:  M A Bhat; P Kumar; A Bhansali; S Majumdar; A Narang
Journal:  Indian J Pediatr       Date:  2000-06       Impact factor: 1.967

2.  Comparison of two cotside methods for the detection of hypoglycaemia among neonates in Nepal.

Authors:  M Ellis; D S Manandhar; N Manandhar; J M Land; N Patel; A M de L Costello
Journal:  Arch Dis Child Fetal Neonatal Ed       Date:  1996-09       Impact factor: 5.747

3.  Evaluation of "point of care" devices in the measurement of low blood glucose in neonatal practice.

Authors:  H T Ho; W K Y Yeung; B W Y Young
Journal:  Arch Dis Child Fetal Neonatal Ed       Date:  2004-07       Impact factor: 5.747

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.