Abstract
The performance of a rapid capillary blood glucose monitor, the Glucoscan 2000, was evaluated in a neonatal patient population by comparison with a reference method on 283 paired patient samples. The median glucose was 56 mg/dL (3.1 mmol/L). The slope of the regression line was .93 with a correlation coefficient r of .82. The average error in Glucoscan values at low, intermediate and high glucose ranges was 43%, 20% and 12% respectively. Precision analysis by replicate determinations also showed an increased coefficient of variation, 7.4%, at low glucose values.Entities: