Literature DB >> 8300680

Causes of dislocation of total hip arthroplasty. CT study of component alignment.

F Pierchon1, G Pasquier, A Cotten, C Fontaine, J Clarisse, A Duquennoy.   

Abstract

We reviewed radiographs and CT scans of 38 total hip arthroplasties which had dislocated (36 posteriorly; 2 anteriorly) and compared the alignment of the prosthetic components with those of 14 uncomplicated arthroplasties. No difference was found between the alignment of the prosthetic components in the two groups. In the seven patients who had reoperations, the cause of dislocation diagnosed by CT was confirmed in only two cases (one retroversion of the cup and one protruding osteophyte). Muscular imbalance rather than malposition of the components was the major factor determining dislocation. CT allows accurate measurement of cup and neck anteversion but contributes little to preoperative planning.

Entities:  

Mesh:

Year:  1994        PMID: 8300680

Source DB:  PubMed          Journal:  J Bone Joint Surg Br        ISSN: 0301-620X


  30 in total

1.  Combined anteversion technique reduced the dislocation in cementless total hip arthroplasty.

Authors:  Yasuharu Nakashima; Masanobu Hirata; Mio Akiyama; Takashi Itokawa; Takuaki Yamamoto; Goro Motomura; Masanobu Ohishi; Satoshi Hamai; Yukihide Iwamoto
Journal:  Int Orthop       Date:  2013-09-13       Impact factor: 3.075

2.  Range of motion measurement of an artificial hip joint using CT images.

Authors:  Haosheng Sun; Hidenori Inaoka; Yutaka Fukuoka; Tadashi Masuda; Akimasa Ishida; Sadao Morita
Journal:  Med Biol Eng Comput       Date:  2007-09-26       Impact factor: 2.602

3.  Intra- and intersurgeon variability in image-free navigation system for THA.

Authors:  Hirotsugu Ohashi; Masanori Matsuura; Yusaku Okamoto; Yoshiaki Okajima
Journal:  Clin Orthop Relat Res       Date:  2009-04-21       Impact factor: 4.176

4.  Computer-assisted versus manual alignment in THA: a probabilistic approach to range of motion.

Authors:  Anthony J Petrella; Joshua Q Stowe; Darryl D D'Lima; Paul J Rullkoetter; Peter J Laz
Journal:  Clin Orthop Relat Res       Date:  2008-10-22       Impact factor: 4.176

5.  Combined anteversion technique for total hip arthroplasty.

Authors:  Lawrence D Dorr; Aamer Malik; Manish Dastane; Zhinian Wan
Journal:  Clin Orthop Relat Res       Date:  2008-11-01       Impact factor: 4.176

Review 6.  Imaging and navigation measurement of acetabular component position in THA.

Authors:  Zhinian Wan; Aamer Malik; Branislav Jaramaz; Lisa Chao; Lawrence D Dorr
Journal:  Clin Orthop Relat Res       Date:  2008-11-01       Impact factor: 4.176

7.  A cadaveric study of posterior dislocation after total hip replacement-effects of head diameter and acetabular anteversion.

Authors:  Fu Yuen Ng; Jiang Tao Zhang; Kwong Yuen Chiu; Chun Hoi Yan
Journal:  Int Orthop       Date:  2010-02-24       Impact factor: 3.075

8.  The effect of acetabular and femoral component version on dislocation in primary total hip arthroplasty.

Authors:  Takaaki Fujishiro; Takafumi Hiranaka; Shingo Hashimoto; Shinya Hayashi; Masahiro Kurosaka; Taiki Kanno; Takeshi Masuda
Journal:  Int Orthop       Date:  2015-07-23       Impact factor: 3.075

9.  Accuracy of plain antero-posterior radiographic-based methods for measurement of acetabular cup version.

Authors:  Mohammad Ali Alzohiry; Mohammad Kamal Abdelnasser; Mohamed Moustafa; Mohamed Mahran; Hatem Bakr; Yaser Khalifa; Ahmed Abelaal; Haisam Atta; Galal Zaki Said
Journal:  Int Orthop       Date:  2018-06-04       Impact factor: 3.075

10.  Early complications of primary total hip arthroplasty in the supine position with a modified Watson-Jones anterolateral approach.

Authors:  Tsuyoshi Nakai; Naxin Liu; Kazumasa Fudo; Toshikazu Mohri; Masaaki Kakiuchi
Journal:  J Orthop       Date:  2014-09-11
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.