| Literature DB >> 28018564 |
Ki-Joon Yoon1, Young-Bum Park2, Hyunmin Choi1, Youngsung Cho1, Jae-Hoon Lee1, Keun-Woo Lee1.
Abstract
PURPOSE: The purpose of this study is to evaluate the stability of interface between Co-Cr-Mo (CCM) UCLA abutment and external hex implant.Entities:
Keywords: CCM UCLA abutment; Chewing simulator; External hex implant; Removal torque value
Year: 2016 PMID: 28018564 PMCID: PMC5179485 DOI: 10.4047/jap.2016.8.6.465
Source DB: PubMed Journal: J Adv Prosthodont ISSN: 2005-7806 Impact factor: 1.904
Fig. 1Chewing simulator CS4 (SD Mechatronic, Feldkirchen, Westerham, Germany).
Fig. 2A schematic diagram of connecting implant and prosthesis & loading position that is 3 mm away from the crown center.
Materials and manufacturer of implant components used in this study
| Component | Material | Manufacturer |
|---|---|---|
| US II fixture | CP Titanium grade 4 | Osstem Co., Seoul, Korea |
| Gold UCLA abutment | Au, Pt, Pd alloy | Osstem Co., Seoul, Korea |
| Gold UCLA abutment screw | Ebonygold | Osstem Co., Seoul, Korea |
| CCM UCLA abutment | Co, Cr, Mo alloy | Geodent Co., Seoul, Korea |
| CCM ULCA abutment screw | Ti Alloy | Geodent Co., Seoul, Korea |
Test parameters of chewing simulation
| Test parameter | Setting | Test parameter | Setting |
|---|---|---|---|
| Chewing cycles | 1,000,000 | Applied weight per sample | 5 kg |
| Cycle frequency | 1 Hz | Hot dwell time | 60 sec |
| Vertical movement | 3 mm | Hot bath temperature | 55℃ |
| Descending speed | 30 mm/s | Cold dwell time | 60 sec |
| Rising speed | 30 mm/s | Cold bath temperature | 5℃ |
| Forward speed | 20 mm/s | Intermediate pause | 0 sec |
| Backward speed | 20 mm/s |
Fig. 3Typical graph of changes in the removal torque values with respect to elapsed time, which were measured every 0.1 sec by a driving torque tester.
Fig. 4Means and standard deviations of removal torque value of the CCM and Gold UCLA abutment groups.
Fig. 5Representative SEM photograph at interface of abutment in the CCM group (original magnification: ×50).
Fig. 6Representative SEM photograph at the interface of abutment in the Gold group (arrows : remarkable wear patterns, original magnification: ×50)
Fig. 7Representative SEM photograph at the interface of fixture in the CCM group (original magnification : ×50).
Fig. 8Representative SEM photograph at the interface of fixture in the Gold group (original magnification : ×50).