X-ray tube potential, filtration, and detector considerations in dual-energy chest radiography.

D M Gauntt1, G T Barnes.   

Abstract

The effect of x-ray tube potential and prepatient and interdetector filtration in single exposure dual energy chest imaging has been studied employing a carefully benchmarked model. The analysis utilized published methodology. Noise in simulated lung and mediastinum fields of the aluminum (bone) and Lucite (soft tissue) images were studied at fixed entrance skin exposure (ESE) for commonly employed sandwich detector and sandwich imaging plate configurations. Our results indicate noise in the lung increases slowly with tube potential above 120 kVp, while noise in the mediastinum decreases rapidly. Also, at high tube potential (> or = 120 kVp) adding moderate amounts of prepatient K-edge filtration (approximately equal to 100 mg/cm2) while optimizing imaging conditions for the lung tends to decrease noise in the lungs by approximately equal to 30% while increasing noise in the mediastinum by a similar amount. Without K-edge prepatient filtration, noise in the lung is minimized with Cu interdetector filter weights near 400 mg/cm2. In the mediastinum noise is minimized with heavier interdetector filter and prepatient K-edge filter weights. Prepatient K-edge filter weights that minimize image noise in either field can increase the tube loading by factors ranging from 10 to 10(10). Systems designed with sandwich detectors using commercially available phosphors and coating weights can produce contrast-to-noise ratios (CNRs) as high as 50% of the theoretical limit (defined as an optimized system with a totally absorbing rear detector).

Entities:  

Mesh:

Year:  1994        PMID: 8177153     DOI: 10.1118/1.597373

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  9 in total

1.  Improved dual-energy material discrimination for dual-source CT by means of additional spectral filtration.

Authors:  A N Primak; J C Ramirez Giraldo; X Liu; L Yu; C H McCollough
Journal:  Med Phys       Date:  2009-04       Impact factor: 4.071

2.  Quantification of breast density with dual energy mammography: a simulation study.

Authors:  Justin L Ducote; Sabee Molloi
Journal:  Med Phys       Date:  2008-12       Impact factor: 4.071

3.  Comparison of three tissue composition measurement techniques using digital mammograms--a signal-to-noise study.

Authors:  D S Breitenstein; C C Shaw
Journal:  J Digit Imaging       Date:  1998-08       Impact factor: 4.056

4.  Back-irradiated and dual-screen sandwich detector configurations for radiography.

Authors:  Anthony R Lubinsky; Adrian Howansky; Hao Zheng; Wei Zhao
Journal:  J Med Imaging (Bellingham)       Date:  2019-07-09

5.  Theoretical and Monte Carlo optimization of a stacked three-layer flat-panel x-ray imager for applications in multi-spectral medical imaging.

Authors:  Sebastian Lopez Maurino; Aldo Badano; Ian A Cunningham; Karim S Karim
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2016-03-29

6.  Dual-source dual-energy CT with additional tin filtration: Dose and image quality evaluation in phantoms and in vivo.

Authors:  Andrew N Primak; Juan Carlos Ramirez Giraldo; Christian D Eusemann; Bernhard Schmidt; Birgit Kantor; Joel G Fletcher; Cynthia H McCollough
Journal:  AJR Am J Roentgenol       Date:  2010-11       Impact factor: 3.959

7.  CsI-detector-based dual-exposure dual energy in chest radiography for lung nodule detection: results of an international multicenter trial.

Authors:  Ricarda Rühl; Magdalena M Wozniak; Michael Werk; François Laurent; Georg Mager; Michel Montaudon; Andreas Pattermann; Antoine Scherrer; Jean-Pierre Tasu; Maciej Pech; Jens Ricke
Journal:  Eur Radiol       Date:  2008-04-19       Impact factor: 5.315

8.  Parameters and computer software for the evaluation of mass attenuation and mass energy-absorption coefficients for body tissues and substitutes.

Authors:  Akintunde A Okunade
Journal:  J Med Phys       Date:  2007-07

9.  The complete evaluation of tube potential on clinical image quality when using direct digital detectors for pelvis and lumbar spine radiography: Re: 'An evaluation of the effect of tube potential on clinical image quality using direct digital detectors for pelvis and lumbar spine radiographs' by Peacock, Steward and Riley.

Authors:  Kholoud Alzyoud; Beverly Snaith; Peter Hogg; Andrew England
Journal:  J Med Radiat Sci       Date:  2020-12-04
  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.