Severe polyethylene wear in total hip arthroplasty. Observations from retrieved AML PLUS hip implants with an ACS polyethylene liner.

J V Bono1, L Sanford, J T Toussaint.   

Abstract

A retrospective review was performed of 94 consecutive Anatomic Medullary Locking Plus (AML+, DePuy, Warsaw, IN) cementless acetabular components implanted between January 1988 and January 1990. All acetabular cups utilized the Acetabular Cup System (ACS) polyethylene liner (Depuy). Of these hips, 72 had been followed for more than 2 years (average, 43 months). There have been 15 (21%) clinical failures, all due to catastrophic wear of the acetabular polyethylene component surfaces. Of the 15 failures, clinical symptoms were absent in 7, an audible squeak was present in 3, and 5 patients reported having pain. All clinical failures had accelerated acetabular wear necessitating revision. Patients in the failure population were younger (56 years vs 62 years) than the remainder of the patients, and had greater cup abduction angles (55 degrees vs 49 degrees). All of the failed ACS polyethylene components had a 32 mm inner diameter articulating surface. The AML + acetabular component outer diameter averaged 56 mm (range, 50-64 mm). Six of 15 failures occurred in cups 58 mm or larger. Acetabular wear in the failure group was 0.77 mm/y (average). The incidence of acetabular (78%) and femoral (71%) osteolysis in the failure population is alarming. The rate of failure (21%) at 46 months of the AML + acetabular component with the ACS polyethylene liner appears to be related to design. The ACS design is flawed by a lack of hemispherical geometry, leading to failure at the superior rim with penetration of the femoral head through the polyethylene and against the metal shell.(ABSTRACT TRUNCATED AT 250 WORDS)

Entities:  

Mesh:

Substances:

Year:  1994        PMID: 8014641     DOI: 10.1016/0883-5403(94)90059-0

Source DB:  PubMed          Journal:  J Arthroplasty        ISSN: 0883-5403            Impact factor:   4.757


  10 in total

1.  Squeaking: Current knowledge and how to avoid it.

Authors:  Arjuna M Imbuldeniya; Simon J Pearce; William L Walter; Bernard A Zicat; William K Walter
Journal:  Curr Rev Musculoskelet Med       Date:  2013-12

2.  Early follow-up for a hybrid total hip arthroplasty using a metal-backed acetabular component designed to reduce "backside" polyethylene wear.

Authors:  Colin E Poole; Shantanu S Patil; Darryl D D'Lima; Clifford W Colwell
Journal:  HSS J       Date:  2005-09

3.  Long-term results of uncemented acetabular cups with an ACS polyethylene liner. a 14-16-year follow-up study.

Authors:  E Garcia-Rey; E Garcia-Cimbrelo
Journal:  Int Orthop       Date:  2006-05-31       Impact factor: 3.075

4.  The 2014 Frank Stinchfield Award: The 'landing zone' for wear and stability in total hip arthroplasty is smaller than we thought: a computational analysis.

Authors:  Jacob M Elkins; John J Callaghan; Thomas D Brown
Journal:  Clin Orthop Relat Res       Date:  2015-02       Impact factor: 4.176

5.  The Safe Zone Range for Cup Anteversion Is Narrower Than for Inclination in THA.

Authors:  William S Murphy; Ho Hyun Yun; Brett Hayden; Jens H Kowal; Stephen B Murphy
Journal:  Clin Orthop Relat Res       Date:  2018-02       Impact factor: 4.176

6.  Clinical and radiographic results and wear performance in different generations of a cementless porous-coated acetabular cup.

Authors:  E García-Rey; E García-Cimbrelo
Journal:  Int Orthop       Date:  2007-01-31       Impact factor: 3.075

7.  Clinical biomechanics of wear in total hip arthroplasty.

Authors:  John J Callaghan; Douglas R Pedersen; Richard C Johnston; Thomas D Brown
Journal:  Iowa Orthop J       Date:  2003

8.  Survivorship of monoblock trabecular metal cups in primary THA : midterm results.

Authors:  Konstantinos N Malizos; Konstantinos Bargiotas; Loukia Papatheodorou; Michael Hantes; Theofilos Karachalios
Journal:  Clin Orthop Relat Res       Date:  2008-01-03       Impact factor: 4.176

9.  The effect of geometry and abduction angle on the stresses in cemented UHMWPE acetabular cups--finite element simulations and experimental tests.

Authors:  Rami K Korhonen; Arto Koistinen; Yrjö T Konttinen; Seppo S Santavirta; Reijo Lappalainen
Journal:  Biomed Eng Online       Date:  2005-05-17       Impact factor: 2.819

10.  Improving the postmarket surveillance of total joint arthroplasty devices.

Authors:  Nizar N Mahomed; Khalid Syed; Clement B Sledge; Troyen A Brennan; Matthew H Liang
Journal:  Open Rheumatol J       Date:  2008-02-25
  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.