Literature DB >> 17265157

Clinical and radiographic results and wear performance in different generations of a cementless porous-coated acetabular cup.

E García-Rey1, E García-Cimbrelo.   

Abstract

We compared clinical results and wear performance in two different generations of a cementless porous-coated cup, analysing the long-term results of 83 uncemented Harris-Galante I cups (32-mm femoral head) and 93 uncemented Harris-Galante II cups (28-mm femoral head). All polyethylene liners were gamma irradiated in air. Polyethylene linear wear was estimated using a software package. The minimum follow-up was 10 years. Nine Harris-Galante I cups and two Harris-Galante II cups were revised due to aseptic loosening or polyethylene problems. The mean femoral head penetration at 6 weeks after surgery was 0.15 +/- 0.05 mm for the Harris-Galante I cups and 0.12 +/- 0.03 for the Harris-Galante II cups (p < 0.001);but mean wear was 0.13 +/- 0.23 mm per year for the Harris-Galante I cups and 0.11 +/- 0.10 for the Harris-Galante II cups (p = 0.740). Most of the metallic shells in both groups showed stable fixation. The so-called second-generation cups had lower initial polyethylene wear that resulted in less polyethylene wear at the latest the follow-up, but the overall wear rate was similar in both groups despite the different femoral head sizes and the improved locking mechanism.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 17265157      PMCID: PMC2269018          DOI: 10.1007/s00264-006-0305-6

Source DB:  PubMed          Journal:  Int Orthop        ISSN: 0341-2695            Impact factor:   3.075


  24 in total

1.  Analysis of temporal wear patterns of porous-coated acetabular components: distinguishing between true wear and so-called bedding-in.

Authors:  C J Sychterz; C A Engh; A Yang; C A Engh
Journal:  J Bone Joint Surg Am       Date:  1999-06       Impact factor: 5.284

2.  Functional results of hip arthroplasty with acrylic prosthesis.

Authors:  R M D'AUBIGNE; M POSTEL
Journal:  J Bone Joint Surg Am       Date:  1954-06       Impact factor: 5.284

3.  Radiographic signs of osseointegration in porous-coated acetabular components.

Authors:  Milan S Moore; James P McAuley; Anthony M Young; Charles A Engh
Journal:  Clin Orthop Relat Res       Date:  2006-03       Impact factor: 4.176

4.  The porous-coated anatomic total hip prosthesis: failure of the metal-backed acetabular component.

Authors:  D J Astion; P Saluan; B N Stulberg; C M Rimnac; S Li
Journal:  J Bone Joint Surg Am       Date:  1996-05       Impact factor: 5.284

5.  Determination of polyethylene wear in total hip replacements with use of digital radiographs.

Authors:  J M Martell; S Berdia
Journal:  J Bone Joint Surg Am       Date:  1997-11       Impact factor: 5.284

6.  Measurement of polyethylene wear in metal-backed acetabular cups. I. Three-dimensional technique.

Authors:  P A Devane; R B Bourne; C H Rorabeck; R M Hardie; J G Horne
Journal:  Clin Orthop Relat Res       Date:  1995-10       Impact factor: 4.176

7.  Radiographic evaluation of penetration by the femoral head into the polyethylene liner over time.

Authors:  C J Sychterz; C A Engh; N Shah; C A Engh
Journal:  J Bone Joint Surg Am       Date:  1997-07       Impact factor: 5.284

8.  Inadequate polyethylene thickness and osteolysis in cementless hip arthroplasty.

Authors:  I D Learmonth; J G Hussell; E J Smith
Journal:  J Arthroplasty       Date:  1997-04       Impact factor: 4.757

9.  Three-dimensional polyethylene wear of a press-fit titanium prosthesis. Factors influencing generation of polyethylene debris.

Authors:  P A Devane; J G Horne; K Martin; G Coldham; B Krause
Journal:  J Arthroplasty       Date:  1997-04       Impact factor: 4.757

10.  Severe polyethylene wear in total hip arthroplasty. Observations from retrieved AML PLUS hip implants with an ACS polyethylene liner.

Authors:  J V Bono; L Sanford; J T Toussaint
Journal:  J Arthroplasty       Date:  1994-04       Impact factor: 4.757

View more
  3 in total

1.  Ten-year results with the Morscher press-fit cup: an uncemented, non-modular, porous-coated cup inserted without screws.

Authors:  Guido Garavaglia; Anne Lübbeke; Christophe Barea; Constantinos Roussos; Robin Peter; Pierre Hoffmeyer
Journal:  Int Orthop       Date:  2010-06-04       Impact factor: 3.075

2.  Polyethylene in total hip arthroplasty: half a century in the limelight.

Authors:  Eduardo García-Rey; Eduardo García-Cimbrelo
Journal:  J Orthop Traumatol       Date:  2010-05-27

Review 3.  Early proximal migration of cups is associated with late revision in THA: a systematic review and meta-analysis of 26 RSA studies and 49 survivalstudies.

Authors:  Bart G Pijls; Marc J Nieuwenhuijse; Marta Fiocco; Josepha Wm Plevier; Saskia Middeldorp; Rob Ghh Nelissen; Edward R Valstar
Journal:  Acta Orthop       Date:  2012-11-05       Impact factor: 3.717

  3 in total