A S Beddar1, D J Mason, P F O'Brien.
Abstract
Uncertainties in dose measurements which may be considered acceptable for standard radiotherapy techniques may not be acceptable for radiosurgery or other precision radiotherapy techniques. One example of this type of uncertainty are the small perturbations to measured dose distributions which are caused by the use of diode detectors. In this article the effect of using diodes to measure small field penumbra is assessed and comments are made about diode orientation and diode construction.Mesh:
Year: 1994 PMID: 7968839 DOI: 10.1118/1.597350