Total scatter factors of small beams: a multidetector and Monte Carlo study.

Paolo Francescon1, Stefania Cora, Carlo Cavedon.   

Abstract

The scope of this study was to estimate total scatter factors (S(c,p)) of the three smallest collimators of the Cyberknife radiosurgery system (5-10 mm in diameter), combining experimental measurements and Monte Carlo simulation. Two microchambers, a diode, and a diamond detector were used to collect experimental data. The treatment head and the detectors were simulated by means of a Monte Carlo code in order to calculate correction factors for the detectors and to estimate total scatter factors by means of a consistency check between measurement and simulation. Results for the three collimators were: S(c,p) (5 mm) = 0.677 +/- 0.004, S(c,p) (7.5 mm) = 0.820 +/- 0.008, S(c,p) (10 mm) = 0.871 +/- 0.008, all relative to the 60 mm collimator at 80 cm source-to-detector distance. The method also allows the full width at half maximum of the electron beam to be estimated; estimations made with different collimators and different detectors were in excellent agreement and gave a value of 2.1 mm. Correction factors to be applied to the detectors for the measurement of S(c,p) were consistent with a prevalence of volume effect for the microchambers and the diamond and a prevalence of scattering from high-Z material for the diode detector. The proposed method is more sensitive to small variations of the electron beam diameter with respect to the conventional method used to commission Monte Carlo codes, i.e., by comparison with measured percentage depth doses (PDD) and beam profiles. This is especially important for small fields (less than 10 mm diameter), for which measurements of PDD and profiles are strongly affected by the type of detector used. Moreover, this method should allow S(c,p) of Cyberknife systems different from the unit under investigation to be estimated without the need for further Monte Carlo calculation, provided that one of the microchambers or the diode detector of the type used in this study are employed. The results for the diamond are applicable only to the specific detector that was investigated due to excessive variability in manufacturing.

Entities:  

Mesh:

Year:  2008        PMID: 18383671      PMCID: PMC3899064          DOI: 10.1118/1.2828195

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  58 in total

1.  Comparative dosimetry of diode and diamond detectors in electron beams for intraoperative radiation therapy.

Authors:  P Björk; T Knöös; P Nilsson
Journal:  Med Phys       Date:  2000-11       Impact factor: 4.071

2.  Comparison of measured and Monte Carlo calculated dose distributions from the NRC linac.

Authors:  D Sheikh-Bagheri; D W Rogers; C K Ross; J P Seuntjens
Journal:  Med Phys       Date:  2000-10       Impact factor: 4.071

3.  Sensitivity of megavoltage photon beam Monte Carlo simulations to electron beam and other parameters.

Authors:  Daryoush Sheikh-Bagheri; D W O Rogers
Journal:  Med Phys       Date:  2002-03       Impact factor: 4.071

4.  An investigation of the operating characteristics of two PTW diamond detectors in photon and electron beams.

Authors:  C De Angelis; S Onori; M Pacilio; G A P Cirrone; G Cuttone; L Raffaele; M Bucciolini; S Mazzocchi
Journal:  Med Phys       Date:  2002-02       Impact factor: 4.071

5.  Temperature dependence of commercially available diode detectors.

Authors:  Amarjit S Saini; Timothy C Zhu
Journal:  Med Phys       Date:  2002-04       Impact factor: 4.071

6.  Comparison of ionization chambers of various volumes for IMRT absolute dose verification.

Authors:  Leonid B Leybovich; Anil Sethi; Nesrin Dogan
Journal:  Med Phys       Date:  2003-02       Impact factor: 4.071

7.  Evaluation and characterization of parallel plate microchamber's functionalities in small beam dosimetry.

Authors:  Heung-Rae Lee; Mike Pankuch; James C Chu; John J Spokas
Journal:  Med Phys       Date:  2002-11       Impact factor: 4.071

8.  Deconvolution of detector size effect for output factor measurement for narrow Gamma Knife radiosurgery beams.

Authors:  G Bednarz; M Saiful Huq; U F Rosenow
Journal:  Phys Med Biol       Date:  2002-10-21       Impact factor: 3.609

9.  Monte Carlo calculation of depth doses for small field of CyberKnife.

Authors:  Tokihiro Yamamoto; Teruki Teshima; Satoshi Miyajima; Masao Matsumoto; Hiroya Shiomi; Takehiro Inoue; Toshihiko Inoue; Hideo Hirayama
Journal:  Radiat Med       Date:  2002 Nov-Dec

10.  Investigation of photon beam output factors for conformal radiation therapy--Monte Carlo simulations and measurements.

Authors:  F Haryanto; M Fippel; W Laub; O Dohm; F Nüsslin
Journal:  Phys Med Biol       Date:  2002-06-07       Impact factor: 3.609

View more
  16 in total

1.  Direct tumor in vivo dosimetry in highly-conformal radiotherapy: a feasibility study of implantable MOSFETs for hypofractionated extracranial treatments using the Cyberknife system.

Authors:  Paolo Scalchi; Roberto Righetto; Carlo Cavedon; Paolo Francescon; Federico Colombo
Journal:  Med Phys       Date:  2010-04       Impact factor: 4.071

2.  Dosimetric characterization and behaviour in small X-ray fields of a microchamber and a plastic scintillator detector.

Authors:  Massimo Pasquino; Claudia Cutaia; Lorenzo Radici; Serena Valzano; Eva Gino; Carlo Cavedon; Michele Stasi
Journal:  Br J Radiol       Date:  2016-11-09       Impact factor: 3.039

3.  Output factor determination based on Monte Carlo simulation for small cone field in 10-MV photon beam.

Authors:  Kyohei Fukata; Satoru Sugimoto; Chie Kurokawa; Akito Saito; Tatsuya Inoue; Keisuke Sasai
Journal:  Radiol Phys Technol       Date:  2018-04-04

4.  Monte carlo model and output factors of elekta infinity™ 6 and 10 MV photon beam.

Authors:  Sitti Yani; Indra Budiansah; Mohamad Fahdillah Rhani; Freddy Haryanto
Journal:  Rep Pract Oncol Radiother       Date:  2020-04-28

5.  Anthropomorphic Phantoms for Confirmation of Linear Accelerator-Based Small Animal Irradiation.

Authors:  Julian R Perks; Steven Lucero; Arta M Monjazeb; Jian Jian Li
Journal:  Cureus       Date:  2015-03-05

6.  Deriving detector-specific correction factors for rectangular small fields using a scintillator detector.

Authors:  Yujiao Qin; Hualiang Zhong; Ning Wen; Karen Snyder; Yimei Huang; Indrin J Chetty
Journal:  J Appl Clin Med Phys       Date:  2016-11-08       Impact factor: 2.102

7.  Characterization of a new commercial single crystal diamond detector for photon- and proton-beam dosimetry.

Authors:  Yuichi Akino; Archana Gautam; Len Coutinho; Jan Würfel; Indra J Das
Journal:  J Radiat Res       Date:  2015-08-12       Impact factor: 2.724

8.  Application of a Monte Carlo-based method for total scatter factors of small beams to new solid state micro-detectors.

Authors:  Paolo Francescon; Stefania Cora; Carlo Cavedon; Paolo Scalchi
Journal:  J Appl Clin Med Phys       Date:  2009-01-27       Impact factor: 2.102

9.  Diamond detector in absorbed dose measurements in high-energy linear accelerator photon and electron beams.

Authors:  Ramamoorthy Ravichandran; John Pichy Binukumar; Iqbal Al Amri; Cheriyathmanjiyil Antony Davis
Journal:  J Appl Clin Med Phys       Date:  2016-03-08       Impact factor: 2.102

10.  Practical considerations of linear accelerator-based frameless extracranial radiosurgery for treatment of occipital neuralgia for nonsurgical candidates.

Authors:  Travis R Denton; Lisa B E Shields; Jonathan N Howe; Todd S Shanks; Aaron C Spalding
Journal:  J Appl Clin Med Phys       Date:  2017-05-18       Impact factor: 2.102

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.