Characterization of a microSilicon diode detector for small-field photon beam dosimetry.

Yuichi Akino1,2, Masateru Fujiwara2, Keita Okamura3, Hiroya Shiomi4, Hirokazu Mizuno5, Fumiaki Isohashi4, Osamu Suzuki6, Yuji Seo4, Keisuke Tamari4, Kazuhiko Ogawa4.   

Abstract

This study characterized a new unshielded diode detector, the microSilicon (model 60023), for small-field photon beam dosimetry by evaluating the photon beams generated by a TrueBeam STx and a CyberKnife. Temperature dependence was evaluated by irradiating photons and increasing the water temperature from 11.5 to 31.3°C. For Diode E, microSilicon, microDiamond and EDGE detectors, dose linearity, dose rate dependence, energy dependence, percent-depth-dose (PDD), beam profiles and detector output factor (OFdet) were evaluated. The OFdet of the microSilicon detector was compared to the field output factors of the other detectors. The microSilicon exhibited small temperature dependence within 0.4%, although the Diode E showed a linear variation with a ratio of 0.26%/°C. The Diode E and EDGE detectors showed positive correlations between the detector reading and dose rate, whereas the microSilicon showed a stable response within 0.11%. The Diode E and microSilicon demonstrated negative correlations with the beam energy. The OFdet of microSilicon was the smallest among all the detectors. The maximum differences between the OFdet of microSilicon and the field output factors of microDiamond were 2.3 and 1.6% for 5 × 5 mm2 TrueBeam and 5 mm φ CyberKnife beams, respectively. The PDD data exhibited small variations in the dose fall-off region. The microSilicon and microDiamond detectors yielded similar penumbra widths, whereas the other detectors showed steeper penumbra profiles. The microSilicon demonstrated favorable characteristics including small temperature and dose rate dependence as well as the small spatial resolution and output factors suitable for small field dosimetry.
© The Author(s) 2020. Published by Oxford University Press on behalf of The Japanese Radiation Research Society and Japanese Society for Radiation Oncology.

Entities:  

Keywords:  CyberKnife; beam data commissioning; diode detector; small field dosimetry

Mesh:

Substances:

Year:  2020        PMID: 32211851      PMCID: PMC7299273          DOI: 10.1093/jrr/rraa010

Source DB:  PubMed          Journal:  J Radiat Res        ISSN: 0449-3060            Impact factor:   2.724


  30 in total

1.  Comparative dosimetry of diode and diamond detectors in electron beams for intraoperative radiation therapy.

Authors:  P Björk; T Knöös; P Nilsson
Journal:  Med Phys       Date:  2000-11       Impact factor: 4.071

2.  Dose rate and SDD dependence of commercially available diode detectors.

Authors:  Amarjit S Saini; Timothy C Zhu
Journal:  Med Phys       Date:  2004-04       Impact factor: 4.071

3.  Calculation of k(Q(clin),Q(msr) ) (f(clin),f(msr) ) for several small detectors and for two linear accelerators using Monte Carlo simulations.

Authors:  P Francescon; S Cora; N Satariano
Journal:  Med Phys       Date:  2011-12       Impact factor: 4.071

4.  Correction factor measurements for multiple detectors used in small field dosimetry on the Varian Edge radiosurgery system.

Authors:  Sean Tanny; Nicholas Sperling; E Ishmael Parsai
Journal:  Med Phys       Date:  2015-09       Impact factor: 4.071

5.  Dosimetric characteristics of a new unshielded silicon diode and its application in clinical photon and electron beams.

Authors:  Irmgard Griessbach; Markus Lapp; Jörg Bohsung; Günther Gademann; Dietrich Harder
Journal:  Med Phys       Date:  2005-12       Impact factor: 4.071

6.  A new single crystal diamond dosimeter for small beam: comparison with different commercial active detectors.

Authors:  F Marsolat; D Tromson; N Tranchant; M Pomorski; M Le Roy; M Donois; F Moignau; A Ostrowsky; L De Carlan; C Bassinet; C Huet; S Derreumaux; M Chea; K Cristina; G Boisserie; P Bergonzo
Journal:  Phys Med Biol       Date:  2013-10-11       Impact factor: 3.609

7.  Characterization of the plastic scintillation detector Exradin W2 for small field dosimetry.

Authors:  Paulina E Galavis; Lei Hu; Shannon Holmes; Indra J Das
Journal:  Med Phys       Date:  2019-04-08       Impact factor: 4.071

8.  On the correction, perturbation and modification of small field detectors in relative dosimetry.

Authors:  P Papaconstadopoulos; F Tessier; J Seuntjens
Journal:  Phys Med Biol       Date:  2014-09-11       Impact factor: 3.609

9.  Output correction factors for nine small field detectors in 6 MV radiation therapy photon beams: a PENELOPE Monte Carlo study.

Authors:  Hamza Benmakhlouf; Josep Sempau; Pedro Andreo
Journal:  Med Phys       Date:  2014-04       Impact factor: 4.071

10.  Technical Note: Characterization of the new microSilicon diode detector.

Authors:  Ann-Britt Schönfeld; Daniela Poppinga; Rafael Kranzer; Rudy Leon De Wilde; Kay Willborn; Björn Poppe; Hui Khee Looe
Journal:  Med Phys       Date:  2019-07-31       Impact factor: 4.071

View more
  1 in total

1.  Determination of correction factors in small MLC-defined fields for the Razor and microSilicon diode detectors and evaluation of the suitability of the IAEA TRS-483 protocol for multiple detectors.

Authors:  Andrew N McGrath; Samane Golmakani; Timothy J Williams
Journal:  J Appl Clin Med Phys       Date:  2022-06-02       Impact factor: 2.243

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.