Literature DB >> 7886649

Effect of a spacer on pulmonary aerosol deposition from a jet nebuliser during mechanical ventilation.

C J Harvey1, M J O'Doherty, C J Page, S H Thomas, T O Nunan, D F Treacher.   

Abstract

BACKGROUND: Several factors have been identified which improve nebulised aerosol delivery in vitro. One of these is the addition of a spacer to the ventilator circuit which improves aerosol delivery from a jet nebuliser to a model lung by approximately 30%. The current study was designed to demonstrate whether similar improvements could be demonstrated in vivo.
METHODS: Ten patients (seven men) were studied during mechanical ventilation (Siemens Servo 900C) after open heart surgery. Aerosol was delivered using a Siemens Servo 945 nebuliser system (high setting) driving a System 22 Acorn jet nebuliser (Medic-Aid) containing 3 ml technetium-99m labelled human serum albumin (99mTc-HSA (50 micrograms); activity in the first nebulisation, 90 MBq; in the second nebulisation, 185 MBq). Central and peripheral lung aerosol deposition and the time to complete deposition were measured using a gamma camera and compared when the nebuliser was connected to the inspiratory limb using a simple T-piece or a 600 ml spacer.
RESULTS: The addition of the spacer increased total lung deposition (mean (SD) percentage initial nebuliser activity) from 2.2 (0.7)% to 3 (0.8)%. There was no difference in the time required to complete nebulisation (18.2 min v 18.3 min respectively for T-piece and spacer) or in the retention of activity in the nebuliser (46.2% v 47.1% respectively).
CONCLUSIONS: The combination of a spacer with a jet nebuliser increased lung deposition by 36% in mechanically ventilated patients and is a simple way of increasing drug deposition or reducing the amount of an expensive drug required for nebulisation.

Entities:  

Mesh:

Substances:

Year:  1995        PMID: 7886649      PMCID: PMC473707          DOI: 10.1136/thx.50.1.50

Source DB:  PubMed          Journal:  Thorax        ISSN: 0040-6376            Impact factor:   9.139


  10 in total

1.  Nebulizer function during mechanical ventilation.

Authors:  T G O'Riordan; M J Greco; R J Perry; G C Smaldone
Journal:  Am Rev Respir Dis       Date:  1992-05

2.  Aerosol delivery in intubated, mechanically ventilated patients.

Authors:  N R MacIntyre; R M Silver; C W Miller; F Schuler; R E Coleman
Journal:  Crit Care Med       Date:  1985-02       Impact factor: 7.598

3.  Improving the efficiency of drug administration with jet nebulisers.

Authors:  S H Thomas; J A Langford; R D George; D M Geddes
Journal:  Lancet       Date:  1988-01-16       Impact factor: 79.321

4.  Improvement of pressurised aerosol deposition with Nebuhaler spacer device.

Authors:  S P Newman; A B Millar; T R Lennard-Jones; F Morén; S W Clarke
Journal:  Thorax       Date:  1984-12       Impact factor: 9.139

5.  Deposition of pressurized suspension aerosols inhaled through extension devices.

Authors:  S P Newman; F Morén; D Pavia; F Little; S W Clarke
Journal:  Am Rev Respir Dis       Date:  1981-09

6.  Pressurized aerosol versus jet aerosol delivery to mechanically ventilated patients. Comparison of dose to the lungs.

Authors:  H D Fuller; M B Dolovich; G Posmituck; W W Pack; M T Newhouse
Journal:  Am Rev Respir Dis       Date:  1990-02

7.  Pulmonary deposition of a nebulised aerosol during mechanical ventilation.

Authors:  S H Thomas; M J O'Doherty; H M Fidler; C J Page; D F Treacher; T O Nunan
Journal:  Thorax       Date:  1993-02       Impact factor: 9.139

8.  Delivery of ultrasonic nebulized aerosols to a lung model during mechanical ventilation.

Authors:  S H Thomas; M J O'Doherty; C J Page; D F Treacher; T O Nunan
Journal:  Am Rev Respir Dis       Date:  1993-10

9.  Delivery of a nebulized aerosol to a lung model during mechanical ventilation. Effect of ventilator settings and nebulizer type, position, and volume of fill.

Authors:  M J O'Doherty; S H Thomas; C J Page; D F Treacher; T O Nunan
Journal:  Am Rev Respir Dis       Date:  1992-08

10.  Aerosol deposition in mechanically ventilated patients. Optimizing nebulizer delivery.

Authors:  T G O'Riordan; L B Palmer; G C Smaldone
Journal:  Am J Respir Crit Care Med       Date:  1994-01       Impact factor: 21.405

  10 in total
  13 in total

1.  Aerosolization of perfluorocarbons during mechanical ventilation: an in vitro study.

Authors:  Tobias Gregor; Gerd Schmalisch; Wolfram Burkhardt; Hans Proquitté; Roland R Wauer; Mario Rüdiger
Journal:  Intensive Care Med       Date:  2003-04-16       Impact factor: 17.440

2.  In vitro assessment of drug delivery through an endotracheal tube using a dry powder inhaler delivery system.

Authors:  M L Everard; S G Devadason; P N Le Souëf
Journal:  Thorax       Date:  1996-01       Impact factor: 9.139

3.  SPECT-CT Comparison of Lung Deposition using a System combining a Vibrating-mesh Nebulizer with a Valved Holding Chamber and a Conventional Jet Nebulizer: a Randomized Cross-over Study.

Authors:  Jonathan Dugernier; Michel Hesse; Rita Vanbever; Virginie Depoortere; Jean Roeseler; Jean-Bernard Michotte; Pierre-François Laterre; François Jamar; Gregory Reychler
Journal:  Pharm Res       Date:  2016-11-07       Impact factor: 4.200

Review 4.  Inhaled Antibiotics for Gram-Negative Respiratory Infections.

Authors:  Eric Wenzler; Dustin R Fraidenburg; Tonya Scardina; Larry H Danziger
Journal:  Clin Microbiol Rev       Date:  2016-07       Impact factor: 26.132

5.  In vitro study and semiempirical model for aerosol delivery control during mechanical ventilation.

Authors:  Laurent Vecellio; Claude Guérin; Daniel Grimbert; Michele De Monte; Patrice Diot
Journal:  Intensive Care Med       Date:  2005-04-01       Impact factor: 17.440

Review 6.  Cromolyn sodium for the prevention of chronic lung disease in preterm infants.

Authors:  Geraldine Ng; Arne Ohlsson
Journal:  Cochrane Database Syst Rev       Date:  2017-01-23

Review 7.  Bronchodilators for the prevention and treatment of chronic lung disease in preterm infants.

Authors:  Geraldine Ng; Orlando da Silva; Arne Ohlsson
Journal:  Cochrane Database Syst Rev       Date:  2016-12-14

8.  In vitro evaluation of aerosol bronchodilator delivery during mechanical ventilation: pressure-control vs. volume control ventilation.

Authors:  Dean R Hess; Christine Dillman; Robert M Kacmarek
Journal:  Intensive Care Med       Date:  2003-05-15       Impact factor: 17.440

Review 9.  The pulmonary physician in critical care . 12: Acute severe asthma in the intensive care unit.

Authors:  P Phipps; C S Garrard
Journal:  Thorax       Date:  2003-01       Impact factor: 9.139

10.  Pharmacokinetics of high-dose nebulized amikacin in mechanically ventilated healthy subjects.

Authors:  Stephan Ehrmann; Emmanuelle Mercier; Laurent Vecellio; David Ternant; Gilles Paintaud; Pierre-François Dequin
Journal:  Intensive Care Med       Date:  2007-11-29       Impact factor: 17.440

View more