Literature DB >> 27536622

Fracture of Fully-coated Femoral Stem after Primary Total Hip Arthroplasty for Nonunion of Intertrochanteric Fracture: A Case Report.

Young Soo Chun1, Hyung Suk Juh1, Yoon Je Cho2, Kee Hyung Rhyu1.   

Abstract

Femoral stem fracture is an uncommon reason for the failure of total hip arthroplasty, with only 16 cases of fully coated stem fractures reported to date. Here we report a case in which a fully coated primary femoral stem fracture occurred after conversion to total hip arthroplasty for the non-union of an intertrochanteric fracture of the femur. Metallurgic evaluation of the etiology and mechanism revealed that the fracture was initiated by fatigue-related failure and completed by ductile failure on the posterior side of the fracture. Considering the recent trend of treating an intertrochanteric fracture with hip arthroplasty, possible stem failure should be considered, since most patients will have at least one of the known risk factors for stem fracture.

Entities:  

Keywords:  Cementless; Fracture; Stem; Total hip arthroplasty

Year:  2015        PMID: 27536622      PMCID: PMC4972723          DOI: 10.5371/hp.2015.27.3.179

Source DB:  PubMed          Journal:  Hip Pelvis        ISSN: 2287-3260


  11 in total

1.  Fracture of fully coated echelon femoral stems in revision total hip arthroplasty.

Authors:  Joshua Landa; Michael Benke; Alan Dayan; Gavin Pereira; Paul E Di Cesare
Journal:  J Arthroplasty       Date:  2009-02       Impact factor: 4.757

Review 2.  The incidence of implant fractures after total hip arthroplasty.

Authors:  Patrick Sadoghi; Wolfram Pawelka; Michael C Liebensteiner; Alexandra Williams; Andreas Leithner; Gerold Labek
Journal:  Int Orthop       Date:  2014-01       Impact factor: 3.075

3.  Fatigue fracture of a forged cobalt-chromium-molybdenum femoral component inserted with cement. A report of ten cases.

Authors:  S T Woolson; J P Milbauer; J D Bobyn; S Yue; W J Maloney
Journal:  J Bone Joint Surg Am       Date:  1997-12       Impact factor: 5.284

Review 4.  Early failure of modern cemented stems.

Authors:  R L Barrack
Journal:  J Arthroplasty       Date:  2000-12       Impact factor: 4.757

5.  Fracture of VerSys fully bead-coated long femoral stems: report on four fractures in 41 hips.

Authors:  Hsueh-Che Lu; Cheng-Li Lin; Chih-Wei Chang; Kuo-An Lai
Journal:  Kaohsiung J Med Sci       Date:  2012-04-03       Impact factor: 2.744

6.  Cementless femoral components should be made from cobalt chrome.

Authors:  N G Sotereanos; C A Engh; A H Glassman; G E Macalino; C A Engh
Journal:  Clin Orthop Relat Res       Date:  1995-04       Impact factor: 4.176

7.  Fracture of the femoral component after total hip replacement. An analysis of fifty-eight cases.

Authors:  E Y Chao; M B Coventry
Journal:  J Bone Joint Surg Am       Date:  1981-09       Impact factor: 5.284

8.  Defects in failed stems of hip prostheses.

Authors:  W Rostoker; E Y Chao; J O Galante
Journal:  J Biomed Mater Res       Date:  1978-09

9.  Fractured stem in total hip replacement. A clinical review of 120 cases.

Authors:  B M Wroblewski
Journal:  Acta Orthop Scand       Date:  1982-04

10.  Rupture of Extensively Porous Coated Stems. - A Case Series of 2 Patients.

Authors:  X Gallart; G Claret; S Garcia; J A Fernández-Valencia; J Riba
Journal:  J Orthop Case Rep       Date:  2014 Jul-Sep
View more
  1 in total

1.  Bilateral Fractures of Anatomic Medullary Locking Hip Arthroplasty Stems in a Single Patient: A Case Report.

Authors:  Sean E Slaven; Colin J Harrington; John P Cody
Journal:  J Orthop Case Rep       Date:  2020
  1 in total