Evaluation of two bioluminescence-measuring instruments, the Turner Design and Lumac systems, for the rapid screening of urine specimens.

W D Welch, L Thompson, M Layman, P M Southern.   

Abstract

Two bioluminescence-measuring instruments, the Turner Design and Lumac systems, were compared with a standard plate culture method for their ability to rapidly screen 400 urine specimens. For cultures with less than 1,000 CFU/ml the Turner Design, with old and new evaluation formulas, gave 6.5 and 50.6% false-positive results, respectively, versus 17.6% at greater than or equal to 500 relative light units with the Lumac. For cultures which had greater than 10(5) CFU/ml the Turner Design gave 39% (old formula) and 14% (new formula) false-negative results compared with 4% at less than 200 relative units with the Lumac. The microorganisms most frequently isolated in the false-negative cultures from either system were gram-positive cocci. Predictive values for a positive test at greater than 10(5) CFU/ml were 77.4% (old formula) and 35.7% (new formula) for the Turner Design versus only 50% for the Lumac at greater than or equal to 500 relative light units. Predictive values for a negative test for both instruments were greater than 88% at greater than 10(5) CFU/ml. The Turner Design and Lumac systems were 4.0 and 3.7 times as expensive, respectively, as the plate culture method. Although both systems greatly reduce the time required to process urine specimens, their high costs as compared with that of plate culture, their failure to detect many specimens having greater than 10(5) CFU of gram-positive cocci per ml, and the numerous false-positives reported by both instruments suggest that additional improvements in the systems are warranted.

Mesh:

Year:  1984        PMID: 6394619      PMCID: PMC271538          DOI: 10.1128/jcm.20.6.1165-1170.1984

Source DB:  PubMed          Journal:  J Clin Microbiol        ISSN: 0095-1137            Impact factor:   5.948


  12 in total

Review 1.  Analytical luminescence: its potential in the clinical laboratory.

Authors:  T P Whitehead; L J Kricka; T J Carter; G H Thorpe
Journal:  Clin Chem       Date:  1979-09       Impact factor: 8.327

2.  Detection of bacteriuria by luciferase assay of adenosine triphosphate.

Authors:  A Thore; S Anséhn; A Lundin; S Bergman
Journal:  J Clin Microbiol       Date:  1975-01       Impact factor: 5.948

3.  Polymicrobic bacteriuria: significant association with bacteremia.

Authors:  P A Gross; M Flower; G Barden
Journal:  J Clin Microbiol       Date:  1976-03       Impact factor: 5.948

4.  Urine screening with the MS-2.

Authors:  D J Hoban; J C Koss; C A Gratton; A R Ronald
Journal:  J Clin Microbiol       Date:  1983-06       Impact factor: 5.948

5.  Evaluation of a two-minute test for urine screening.

Authors:  M T Pezzlo; M A Wetkowski; E M Peterson; L M de la Maza
Journal:  J Clin Microbiol       Date:  1983-09       Impact factor: 5.948

6.  Use of leukocyte esterase-nitrate activity as predictive assays of significant bacteriuria.

Authors:  D L Smalley; A N Dittmann
Journal:  J Clin Microbiol       Date:  1983-11       Impact factor: 5.948

7.  Screening of urine cultures by three automated systems.

Authors:  M T Pezzlo; G L Tan; E M Peterson; L M de la Maza
Journal:  J Clin Microbiol       Date:  1982-03       Impact factor: 5.948

8.  An automated test for the detection of significant bacteriuria.

Authors:  H H Johnston; C J Mitchell; G D Curtis
Journal:  Lancet       Date:  1976-08-21       Impact factor: 79.321

9.  Adenosinetriphosphate (ATP) levels in foods contaminated by bacteria.

Authors:  A N Sharpe; M N Woodrow; A K Jackson
Journal:  J Appl Bacteriol       Date:  1970-12

10.  Rapid diagnosis of bacteraemia by measuring bacterial adenosine triphosphate in blood culture bottles using bioluminescence.

Authors:  B Beckers; H R Lang
Journal:  Med Microbiol Immunol       Date:  1983       Impact factor: 3.402

View more
  9 in total

1.  Rapid bioluminescence method for bacteriuria screening.

Authors:  M T Pezzlo; V Ige; A P Woolard; E M Peterson; L M de la Maza
Journal:  J Clin Microbiol       Date:  1989-04       Impact factor: 5.948

2.  Rapid screening for bacteriuria in pregnancy.

Authors:  W Graninger; D Fleischmann; B Schneeweiss; L Aram; F Stockenhuber
Journal:  Infection       Date:  1992 Jan-Feb       Impact factor: 3.553

3.  Rapid antimicrobial susceptibility determination of uropathogens in clinical urine specimens by use of ATP bioluminescence.

Authors:  Vesna Ivancic; Mitra Mastali; Neil Percy; Jeffrey Gornbein; Jane T Babbitt; Yang Li; Elliot M Landaw; David A Bruckner; Bernard M Churchill; David A Haake
Journal:  J Clin Microbiol       Date:  2008-02-13       Impact factor: 5.948

4.  Automated bacteriuria screening using the Berthold LB 950 luminescence analyser.

Authors:  G D Curtis; H H Johnston; A R Hack
Journal:  Eur J Clin Microbiol       Date:  1987-06       Impact factor: 3.267

5.  Bioluminescence screening for bacteriuria.

Authors:  J C Kolbeck; R A Padgett; E G Estevez; L J Harrell
Journal:  J Clin Microbiol       Date:  1985-04       Impact factor: 5.948

6.  Clinical evaluation of the Lumac bioluminescence method for screening urine specimens.

Authors:  E T Martin; J A Cote; L K Perry; W J Martin
Journal:  J Clin Microbiol       Date:  1985-07       Impact factor: 5.948

Review 7.  Detection of urinary tract infections by rapid methods.

Authors:  M Pezzlo
Journal:  Clin Microbiol Rev       Date:  1988-07       Impact factor: 26.132

8.  Leukocyte esterase-nitrite and bioluminescence assays as urine screens.

Authors:  B M Males; W R Bartholomew; D Amsterdam
Journal:  J Clin Microbiol       Date:  1985-10       Impact factor: 5.948

9.  An IMS/ATP Assay for the Detection of Mycobacterium tuberculosis in Urine.

Authors:  Dawn M Hunter; Daniel V Lim
Journal:  Tuberc Res Treat       Date:  2012-05-09
  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.