Reproducibility of linear cardiac output measurement by Doppler ultrasound alone.

F M McLennan, N E Haites, J D Mackenzie, M K Daniel, J M Rawles.   

Abstract

Inclusion of a pig aorta in an artificial circulation with pulsed blood flow allowed correlation of minute distance, measured in the aorta by Doppler ultrasound, and absolute blood flow, measured by timed blood-volume collection. The correlation coefficient was 0.99 with a standard error of prediction that was 5.4% of the minute distance predicted at a standard flow rate of 5 litres per minute. The horizontal distance between 95% confidence limits for a single prediction expressed as a percentage of 5 litres per minute was 33%, and this corresponded to the range of flow rates of 1.65 litres per minute that could give rise to the same measurement. In 142 patients duplicate measurements of minute distance were made with repositioning of the ultrasound transducer between recordings. The mean difference between paired readings, expressed as a percentage of the average (SD) of each pair was 5.4 (4.7)%. Thus, the non-invasive measurement of linear cardiac output by Doppler ultrasound is similarly reproducible in vitro and in vivo and compares favourably with the measurement of volumetric cardiac output by thermodilution.

Entities:  

Mesh:

Year:  1986        PMID: 3511929      PMCID: PMC1232064          DOI: 10.1136/hrt.55.1.25

Source DB:  PubMed          Journal:  Br Heart J        ISSN: 0007-0769


  20 in total

1.  Transcutaneous aortovelography. A quantitative evaluation.

Authors:  R F Sequeira; L H Light; G Cross; E B Raftery
Journal:  Br Heart J       Date:  1976-05

2.  Transcutaneous aortovelography: reproducibility in adults and children.

Authors:  C B Fraser; L H Light; E A Shinebourne; A Buchthal; M J Healy; J A Beardshaw
Journal:  Eur J Cardiol       Date:  1976-06

Review 3.  Transcutaneous aortovelography. A new window on the circulation?

Authors:  H Light
Journal:  Br Heart J       Date:  1976-05

4.  Central haemodynamics in essential hypertension.

Authors:  P Lund-Johansen
Journal:  Acta Med Scand Suppl       Date:  1977

5.  Suprasternal ultrasonography.

Authors:  B B Goldberg
Journal:  JAMA       Date:  1971-01-11       Impact factor: 56.272

6.  Method of assessing the reproducibility of blood flow measurement: factors influencing the performance of thermodilution cardiac output computers.

Authors:  J D Mackenzie; N E Haites; J M Rawles
Journal:  Br Heart J       Date:  1986-01

7.  Doppler echocardiographic measurement of cardiac output: comparison with a non-golden standard.

Authors:  A H Schuster; N C Nanda
Journal:  Am J Cardiol       Date:  1984-01-01       Impact factor: 2.778

8.  Combined doppler and phased-array echocardiographic estimation of cardiac output.

Authors:  P A Magnin; J A Stewart; S Myers; O von Ramm; J A Kisslo
Journal:  Circulation       Date:  1981-02       Impact factor: 29.690

9.  How far is the cardiac output?

Authors:  N E Haites; F M McLennan; D H Mowat; J M Rawles
Journal:  Lancet       Date:  1984-11-03       Impact factor: 79.321

10.  Assessment of cardiac output by the Doppler ultrasound technique alone.

Authors:  N E Haites; F M McLennan; D H Mowat; J M Rawles
Journal:  Br Heart J       Date:  1985-02
View more
  14 in total

1.  Reproducibility of Doppler blood flow velocity waveform measurements: study on variability within and between day and during haemodynamic intervention in normal subjects.

Authors:  G H Rimoy; N K Bhaskar; P C Rubin
Journal:  Eur J Clin Pharmacol       Date:  1991       Impact factor: 2.953

2.  Improved method for cardiac output determination in man using ultrasound Doppler technique.

Authors:  M Eriksen; L Walløe
Journal:  Med Biol Eng Comput       Date:  1990-11       Impact factor: 2.602

3.  Haemoglobin concentration and linear cardiac output, peripheral resistance, and oxygen transport.

Authors:  M K Daniel; B Bennett; A A Dawson; J M Rawles
Journal:  Br Med J (Clin Res Ed)       Date:  1986-04-05

4.  Disturbed synthesis of insulinlike growth factor I and its binding proteins may influence renal function changes in liver cirrhosis.

Authors:  C M Fernández-Rodriguez; I Prada; A Andrade; M Moreiras; R Guitián; R Aller; J L Lledó; G Cacho; J Quiroga; J Prieto
Journal:  Dig Dis Sci       Date:  2001-06       Impact factor: 3.199

5.  Effect of age and blood pressure on aortic size and stroke distance.

Authors:  B A Towfiq; J Weir; J M Rawles
Journal:  Br Heart J       Date:  1986-06

6.  Reproducibility of cardiac output measurement by cross sectional and Doppler echocardiography.

Authors:  S C Robson; A Murray; I Peart; A Heads; S Hunter
Journal:  Br Heart J       Date:  1988-06

7.  Method of assessing the reproducibility of blood flow measurement: factors influencing the performance of thermodilution cardiac output computers.

Authors:  J D Mackenzie; N E Haites; J M Rawles
Journal:  Br Heart J       Date:  1986-01

Review 8.  Reliability and validity of measures of cardiac output during incremental to maximal aerobic exercise. Part II: Novel techniques and new advances.

Authors:  D E Warburton; M J Haykowsky; H A Quinney; D P Humen; K K Teo
Journal:  Sports Med       Date:  1999-04       Impact factor: 11.136

9.  Feasibility and variability of six methods for the echocardiographic and Doppler determination of cardiac output.

Authors:  G L Nicolosi; E Pungercic; E Cervesato; D Pavan; L Modena; E Moro; V Dall'Aglio; D Zanuttini
Journal:  Br Heart J       Date:  1988-03

10.  Cardiac function in healthy infants and children: Doppler echocardiographic evaluation.

Authors:  K Hanseus; G Björkhem; N R Lundström
Journal:  Pediatr Cardiol       Date:  1994 Sep-Oct       Impact factor: 1.655

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.