Validation of the beat to beat measurement of blood velocity in the human ascending aorta by a new high temporal resolution Doppler ultrasound spectral analyser.

A J Coats1, C Murphy, J Conway, P Sleight.   

Abstract

OBJECTIVE: To develop and validate a high temporal resolution spectral analysis system for Doppler measurements of blood velocity in the ascending aorta.
DESIGN: An observational laboratory and clinical study comparing Doppler velocity-based measurements with fluid collection, electromagnetic flow catheters and probes, and thermodilution.
SETTING: Tertiary referral cardiology unit and cardiac catheter laboratory. PATIENTS: Patients undergoing routine cardiac catheterisation for ischaemic heart disease, cardiac failure, and primary pulmonary hypertension.
RESULTS: There was good agreement between Doppler-derived and electromagnetic cuff or catheter measurements of velocity in an experimental flow rig (SD of differences 4.75% for velocity integral) and in the patients (SD of differences 4% for velocity integral). There was also reasonably good agreement between simultaneous Doppler-derived and thermodilution-estimated cardiac output measurements in patients undergoing cardiac catheterisation (SD of differences 12.6%).
CONCLUSIONS: This new method of high temporal resolution spectral analysis improves the resolution of rapidly changing blood velocities and may improve the ability to describe blood velocity patterns in the ascending aorta.

Entities:  

Mesh:

Year:  1992        PMID: 1389745      PMCID: PMC1025022          DOI: 10.1136/hrt.68.8.223

Source DB:  PubMed          Journal:  Br Heart J        ISSN: 0007-0769


  22 in total

Review 1.  Arterial stiffness, systolic blood pressure, and logical treatment of arterial hypertension.

Authors:  M O'Rourke
Journal:  Hypertension       Date:  1990-04       Impact factor: 10.190

2.  Determination of stroke volume and cardiac output during exercise: comparison of two-dimensional and Doppler echocardiography, Fick oximetry, and thermodilution.

Authors:  J Christie; L M Sheldahl; F E Tristani; K B Sagar; M J Ptacin; S Wann
Journal:  Circulation       Date:  1987-09       Impact factor: 29.690

3.  Simultaneous Doppler blood velocity measurements from aorta and radial artery in normal human subjects.

Authors:  R C Targett; B Levy; A Bardou; M B McIlroy
Journal:  Cardiovasc Res       Date:  1985-07       Impact factor: 10.787

4.  The effects of sampling site on the two-dimensional echo-Doppler determination of cardiac output.

Authors:  A J Labovitz; T A Buckingham; K Habermehl; J Nelson; H L Kennedy; G A Williams
Journal:  Am Heart J       Date:  1985-02       Impact factor: 4.749

5.  Validation of beat by beat pulsed Doppler measurements of ascending aortic blood velocity in man.

Authors:  J A Innes; C J Mills; M I Noble; K Murphy; S Pugh; A C Shore; A Guz
Journal:  Cardiovasc Res       Date:  1987-01       Impact factor: 10.787

6.  Accuracy of cardiac output measured by continuous wave Doppler echocardiography during dynamic exercise testing in the supine position in patients with coronary artery disease.

Authors:  M Maeda; M Yokota; M Iwase; T Miyahara; H Hayashi; I Sotobata
Journal:  J Am Coll Cardiol       Date:  1989-01       Impact factor: 24.094

7.  Variable effects of changes in flow rate through the aortic, pulmonary and mitral valves on valve area and flow velocity: impact on quantitative Doppler flow calculations.

Authors:  W J Stewart; L Jiang; R Mich; N Pandian; J L Guerrero; A E Weyman
Journal:  J Am Coll Cardiol       Date:  1985-09       Impact factor: 24.094

8.  Trancutaneous determination of aortic blood-flow velocities in man.

Authors:  L L Huntsman; E Gams; C C Johnson; E Fairbanks
Journal:  Am Heart J       Date:  1975-05       Impact factor: 4.749

9.  Determination of cardiac output by transcutaneous continuous-wave ultrasonic Doppler computer.

Authors:  P A Chandraratna; M Nanna; C McKay; A Nimalasuriya; R Swinney; U Elkayam; S H Rahimtoola
Journal:  Am J Cardiol       Date:  1984-01-01       Impact factor: 2.778

10.  Effects of antihypertensive agents on blood velocity: implications for atherogenesis.

Authors:  J D Spence
Journal:  Can Med Assoc J       Date:  1982-10-15       Impact factor: 8.262

View more
  2 in total

1.  Load dependence of changes in forearm and peripheral vascular resistance after acute leg exercise in man.

Authors:  M Piepoli; J E Isea; G Pannarale; S Adamopoulos; P Sleight; A J Coats
Journal:  J Physiol       Date:  1994-07-15       Impact factor: 5.182

2.  Cardiac output measurements via echocardiography versus thermodilution: A systematic review and meta-analysis.

Authors:  Yun Zhang; Yan Wang; Jing Shi; Zhiqiang Hua; Jinyu Xu
Journal:  PLoS One       Date:  2019-10-03       Impact factor: 3.240

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.