Determination of cardiac output by an angle and diameter independent dual beam Doppler technique in critically ill infants.

C F Wippermann1, D Schranz, R Huth, F Zepp, H Oelert, B K Jüngst.   

Abstract

OBJECTIVE: To compare cardiac output measurements in critically ill infants by the dual beam Doppler and thermodilution techniques.
DESIGN: Prospective direct comparison of the two techniques. For statistical evaluation one randomly assigned paired measurement of every patient was used.
SETTING: Paediatric intensive care unit in a university hospital. PATIENTS: 18 infants after open heart surgery aged 4-25 months (weight 4-10 kg).
INTERVENTIONS: Cardiac output measurements by dual beam Doppler and thermodilution techniques were performed within 10 minutes of each other and without knowledge of the results of the other methods. Multiple measurements were performed on some patients with a pharmacological or electrophysiological intervention or with a minimum of six hours between each pair of measurements.
MEASUREMENTS AND MAIN RESULTS: Three patients were excluded because of an inadequate Doppler signal or a significant residual shunt. Cardiac output measurements ranged from 0.4 to 2.2 l/min for the thermodilution technique and from 0.5 to 2.1 l/min for the dual beam Doppler technique. Agreement between both methods was acceptable. The mean difference between the two methods was 0.026 l/min with two standard deviations ranging from -0.20 to 0.26 l/min.
CONCLUSION: The dual beam Doppler technique was shown to have promise for the non-invasive determination of cardiac output in critically ill infants.

Entities:  

Mesh:

Year:  1992        PMID: 1540440      PMCID: PMC1024751          DOI: 10.1136/hrt.67.2.180

Source DB:  PubMed          Journal:  Br Heart J        ISSN: 0007-0769


  19 in total

1.  Automated non-invasive measurement of cardiac output by the carbon dioxide rebreathing method: comparisons with dye dilution and thermodilution.

Authors:  A E Russell; S A Smith; M J West; P E Aylward; R J McRitchie; R M Hassam; R B Minson; L M Wing; J P Chalmers
Journal:  Br Heart J       Date:  1990-03

2.  A new approach to the noninvasive measurement of cardiac output using an annular array Doppler technique--II. Practical implementation and results.

Authors:  J M Evans; R Skidmore; J D Baker; P N Wells
Journal:  Ultrasound Med Biol       Date:  1989       Impact factor: 2.998

3.  A new approach to the noninvasive measurement of cardiac output using an annular array Doppler technique--I. Theoretical considerations and ultrasonic fields.

Authors:  J M Evans; R Skidmore; N P Luckman; P N Wells
Journal:  Ultrasound Med Biol       Date:  1989       Impact factor: 2.998

4.  Blood flow measurement using the attenuation-compensated volume flowmeter.

Authors:  C F Hottinger; J D Meindl
Journal:  Ultrason Imaging       Date:  1979-01       Impact factor: 1.578

5.  Clinical evaluation of a diameter sensing Doppler cardiac output meter.

Authors:  R Niclou; S M Teague; R Lee
Journal:  Crit Care Med       Date:  1990-04       Impact factor: 7.598

6.  Determination of cardiac output in critically ill patients by dual beam Doppler echocardiography.

Authors:  D S Looyenga; P R Liebson; R C Bone; R A Balk; J V Messer
Journal:  J Am Coll Cardiol       Date:  1989-02       Impact factor: 24.094

7.  A comparison of thermodilution and pulsed Doppler cardiac output measurement in critically ill children.

Authors:  D A Notterman; F V Castello; C Steinberg; B M Greenwald; J E O'Loughlin; J P Gold
Journal:  J Pediatr       Date:  1989-10       Impact factor: 4.406

8.  Comparison of pulsed Doppler and thermodilution methods for measuring cardiac output in critically ill patients.

Authors:  K D Donovan; G J Dobb; M A Newman; B E Hockings; M Ireland
Journal:  Crit Care Med       Date:  1987-09       Impact factor: 7.598

9.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

10.  Bedside pulmonary artery catheterization in pediatrics.

Authors:  M M Pollack; T P Reed; P R Holbrook; A I Fields
Journal:  J Pediatr       Date:  1980-02       Impact factor: 4.406

View more
  2 in total

Review 1.  Accuracy and repeatability of pediatric cardiac output measurement using Doppler: 20-year review of the literature.

Authors:  Michelle S Chew; Jan Poelaert
Journal:  Intensive Care Med       Date:  2003-09-04       Impact factor: 17.440

Review 2.  Accuracy and precision of minimally-invasive cardiac output monitoring in children: a systematic review and meta-analysis.

Authors:  Koichi Suehiro; Alexandre Joosten; Linda Suk-Ling Murphy; Olivier Desebbe; Brenton Alexander; Sang-Hyun Kim; Maxime Cannesson
Journal:  J Clin Monit Comput       Date:  2015-08-29       Impact factor: 2.502

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.