B F Woolfrey, R T Lally, M N Ederer.
Abstract
A total of 398 consecutive clinical staphylococcus isolates, of which 205 were coagulase negative and 193 were coagulase positive, were tested in parallel by using AutoMicrobic system Gram-Positive Susceptibility-MIC cards and modified Mueller-Hinton agar containing 4% NaCl and oxacillin (6 micrograms/ml). The AutoMicrobic system cards correctly detected 103 of 104 (99%) oxacillin-resistant coagulase-negative isolates with no reports of false resistance.Entities:
Mesh:
Substances:
Year: 1986 PMID: 3633924 PMCID: PMC268710 DOI: 10.1128/jcm.23.3.629-630.1986
Source DB: PubMed Journal: J Clin Microbiol ISSN: 0095-1137 Impact factor: 5.948