Rapid recognition of methicillin-resistant Staphylococcus aureus by use of automated test systems.

J H Jorgensen, J Redding, J E Johnson, V Holloway, R J Almeida.   

Abstract

The ability to rapidly recognize methicillin-resistant Staphylococcus aureus by use of two automated instrument systems, the MS-2 system (Abbott Laboratories, Diagnostics Division, Irving, Tex.) and the AutoMicrobic system (Vitek Systems, Hazelwood, Mo.), was evaluated on a collection of 95 methicillin-resistant S. aureus isolates recovered from at least six geographical areas of the United States. Isolates were simultaneously tested with both systems, and the results were compared with MIC tests performed by the National Committee for Clinical Laboratory Standards agar dilution method. Methicillin-resistant S. aureus isolates were defined as those with a methicillin MIC greater than or equal to 8 micrograms/ml by the reference procedure. Overall, with the AutoMicrobic system, 94.7% of 95 methicillin-resistant S. aureus isolates were detected, and with the MS-2 system, 91.6% of the isolates were detected. Isolates with methicillin MICs greater than or equal to 32 micrograms/ml were readily detected with both systems (41 of 42 isolates). Of 53 isolates from three locales with methicillin MICs of 8 or 16 micrograms/ml, 90.6% (48) were detected by the AutoMicrobic system, whereas 86.8% (46) were detected by the MS-2 system. A program update which has been added to the MS-2 system prints a warning message indicating possible methicillin-resistant S. aureus with isolates which demonstrate multiple antibiotic resistance (greater than or equal to four drugs other than methicillin). This warning message would have provided presumptive recognition of six of eight isolates with discrepant results for methicillin by the MS-2 system.

Entities:  

Mesh:

Substances:

Year:  1984        PMID: 6386859      PMCID: PMC271344          DOI: 10.1128/jcm.20.3.430-433.1984

Source DB:  PubMed          Journal:  J Clin Microbiol        ISSN: 0095-1137            Impact factor:   5.948


  21 in total

1.  Clinical, epidemiologic and bacteriologic observations of an outbreak of methicillin-resistant Staphylococcus aureus at a large community hospital.

Authors:  J J Klimek; F J Marsik; R C Bartlett; B Weir; P Shea; R Quintiliani
Journal:  Am J Med       Date:  1976-09       Impact factor: 4.965

2.  Methicillin-resistant Staphylococcus aureus susceptibility testing by an automated system, Autobac I.

Authors:  T J Cleary; D Maurer
Journal:  Antimicrob Agents Chemother       Date:  1978-05       Impact factor: 5.191

3.  The detection of methicillin resistance in Staphylococcus aureus.

Authors:  J H Hewitt; A W Coe; M T Parker
Journal:  J Med Microbiol       Date:  1969-11-04       Impact factor: 2.472

4.  Methicillin-resistant Staphylococcus aureus: introduction and spread within a hospital.

Authors:  J E Peacock; F J Marsik; R P Wenzel
Journal:  Ann Intern Med       Date:  1980-10       Impact factor: 25.391

5.  Reliability of the microdilution technic for detection of methicillin-resistant strains of staphylococcus aureus.

Authors:  A L Barry; R E Badal
Journal:  Am J Clin Pathol       Date:  1977-05       Impact factor: 2.493

6.  Genetic behavior of the methicillin resistance determinant in Staphylococcus aureus.

Authors:  G C Stewart; E D Rosenblum
Journal:  J Bacteriol       Date:  1980-12       Impact factor: 3.490

7.  Reliability of the MS-2 system in detecting methicillin-resistant Staphylococcus aureus.

Authors:  J M Boyce; R L White; M C Bonner; W R Lockwood
Journal:  J Clin Microbiol       Date:  1982-02       Impact factor: 5.948

8.  Methicillin-resistant Staphylococcus aureus susceptibility testing with Abbott MS-2 system.

Authors:  J R Carlson; F E Conley; D L Cahall
Journal:  Antimicrob Agents Chemother       Date:  1982-04       Impact factor: 5.191

9.  Effect of temperature on the in vitro susceptibility of Staphylococcus aureus to penicillinase-resistant penicillins.

Authors:  C Thornsberry; J Q Caruthers; C N Baker
Journal:  Antimicrob Agents Chemother       Date:  1973-09       Impact factor: 5.191

10.  Reliability of the Kirby-Bauer disc diffusion method for detecting methicillin-resistant strains of Staphylococcus aureus.

Authors:  W L Drew; A L Barry; R O'Toole; J C Sherris
Journal:  Appl Microbiol       Date:  1972-08
View more
  7 in total

1.  Evaluation of the AutoMicrobic system Gram-Positive Susceptibility-MIC card for detection of oxacillin-resistant coagulase-negative staphylococci.

Authors:  B F Woolfrey; R T Lally; M N Ederer
Journal:  J Clin Microbiol       Date:  1986-03       Impact factor: 5.948

2.  Evaluation of laboratory tests for detection of methicillin-resistant Staphylococcus aureus and Staphylococcus epidermidis.

Authors:  P E Coudron; D L Jones; H P Dalton; G L Archer
Journal:  J Clin Microbiol       Date:  1986-11       Impact factor: 5.948

3.  Rapid detection (4 h) of methicillin-resistant Staphylococcus aureus by a bioluminescence method.

Authors:  C H Park; D L Hixon; C M McLaughlin; J F Cook
Journal:  J Clin Microbiol       Date:  1988-06       Impact factor: 5.948

4.  Use of the Autobac system for detection of methicillin-resistant Staphylococcus aureus.

Authors:  Y de Rautlin de la Roy; E Chevalier-Burbaud; S Pannetier; F Souchaud
Journal:  J Clin Microbiol       Date:  1985-09       Impact factor: 5.948

5.  Misclassification of susceptible strains of Staphylococcus aureus as methicillin-resistant S. aureus By a rapid automated susceptibility testing system.

Authors:  J Ribeiro; F D Vieira; T King; J B D'Arezzo; J M Boyce
Journal:  J Clin Microbiol       Date:  1999-05       Impact factor: 5.948

Review 6.  Laboratory and epidemiologic experience with methicillin-resistant Staphylococcus aureus in the USA.

Authors:  J H Jorgensen
Journal:  Eur J Clin Microbiol       Date:  1986-12       Impact factor: 3.267

7.  Comparative evaluation of four systems for determining susceptibility of gram-positive organisms.

Authors:  D Henry; L Kunzer; J Ngui-Yen; J Smith
Journal:  J Clin Microbiol       Date:  1986-04       Impact factor: 5.948

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.