Three-Dimensional Finite Element Analysis of the Stress Distribution at the Internal Implant-Abutment Connection.

Sung-Yong Cho, Yoon-Hyuk Huh, Chan-Jin Park, Lee-Ra Cho.   

Abstract

This study investigated stress distribution in four different implant-abutment interface conditions in the internal tapered connection implant system. Four different implant diameters (3.5 mm, 4.0 mm, 4.5 mm, and 5.0 mm) and two abutment types (hexagonal and conical) were simulated. Four unique implant-abutment interface conditions were assumed based on wall thickness, mating surface length, distance to the vertical stop, and abutment shape. Axial and oblique loading was applied during abutment screw preload, and the Von Mises stresses were measured at the implant-abutment and abutment-screw interfaces. The implant-abutment interface stress decreased as the wall thickness increased. As the mating surface increased, the stress distribution trended downward, and when the distance to the implant vertical stop was 0 μm, the Von Mises stress was extremely high at the vertical stop. Despite their different shapes, the abutments showed similar stress distributions. However, the maximum Von Mises stress was higher in the conical connection than in the hexagonal connection, particularly at the contralateral side to loading. To decrease the stress distribution at the implant-abutment interface, the implant wall thickness, mating surface contact length, distance to the vertical stop, and abutment shape should be carefully considered.

Mesh:

Substances:

Year:  2016        PMID: 27100815     DOI: 10.11607/prd.2351

Source DB:  PubMed          Journal:  Int J Periodontics Restorative Dent        ISSN: 0198-7569            Impact factor:   1.840


  3 in total

1.  Cad-cam procedure and implant-prosthetic rehaBilitation. Case report.

Authors:  F Guzzo; G DE Luca; P Barnaba; D Severino
Journal:  Oral Implantol (Rome)       Date:  2016-11-13

2.  A Finite Element Stress Analysis of a Concical Triangular Connection in Implants: A New Proposal.

Authors:  Romy Angeles Maslucan; John Alexis Dominguez
Journal:  Materials (Basel)       Date:  2022-05-20       Impact factor: 3.748

3.  Stress analysis and factor of safety in three dental implant systems by finite element analysis.

Authors:  E Menacho-Mendoza; R Cedamanos-Cuenca; A Díaz-Suyo
Journal:  Saudi Dent J       Date:  2022-08-27
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.