Thierry Gaillard1, Sonia Ramos-Pascual2, Mo Saffarini3, Jean-Pierre Piton4. 1. Polyclinique du Beaujolais, 120 Anc. Rte de Beaujeu, 69400, Arnas, France. 2. ReSurg SA, Rue Saint-Jean 22, 1260, Nyon, Switzerland. journals@resurg.com. 3. ReSurg SA, Rue Saint-Jean 22, 1260, Nyon, Switzerland. 4. UNEOS - Hôpitaux Privés de Metz, 15 Rue de Sarre, 57070, Metz, France.
Abstract
PURPOSE: To report long-term survival and clinical outcomes of primary total hip arthroplasty (THA) using a Saturne cementless dual-mobility (DM) cup, and investigate whether patient demographics or surgical parameters affect clinical scores. METHODS: A consecutive series of primary THAs implanted with Saturne cementless DM cups between 01/09/2009-31/12/ 2011 was retrospectively assessed. Patients were postoperatively evaluated using modified Harris hip score (mHHS) and forgotten joint score (FJS). Complications, reoperations, and revisions were noted. Regression analyses were performed to determine associations of postoperative mHHS with preoperative and intraoperative variables. Ten-year Kaplan-Meier survival was calculated. RESULTS: Of 308 patients (308 hips), 111 (36%) had died with their original cups in place, 29 (9%) were lost-to-follow-up, and 5 (2%) required cup revision, leaving a final cohort of 163 (53%) with their original cup in place at a follow-up of ≥ 10 years. Ten-year survival was 98% considering cup revision for any reason as endpoint; 99% considering cup revision for aseptic loosening as endpoint; 96% considering stem revision for any reason as endpoint; and 96% considering any revision as endpoint. The final cohort of 163 patients was assessed at 11 ± 1 years (range, 10-13), mHHS was 85 ± 16 (range, 31-100) and FJS was 84 ± 24 (range, 0-100). Multivariable regression analysis revealed that postoperative mHHS significantly worsened with age (β = -0.48, p = 0.007) and BMI (β = -0.70, p = 0.008), as well as for 22 mm head sizes (β = -6.98, p = 0.046). CONCLUSIONS: The Saturne DM cup granted satisfactory survival and clinical outcomes at a minimum follow-up of 10 years, and resulted in no cases of intra- or extra-prosthetic dislocations.
PURPOSE: To report long-term survival and clinical outcomes of primary total hip arthroplasty (THA) using a Saturne cementless dual-mobility (DM) cup, and investigate whether patient demographics or surgical parameters affect clinical scores. METHODS: A consecutive series of primary THAs implanted with Saturne cementless DM cups between 01/09/2009-31/12/ 2011 was retrospectively assessed. Patients were postoperatively evaluated using modified Harris hip score (mHHS) and forgotten joint score (FJS). Complications, reoperations, and revisions were noted. Regression analyses were performed to determine associations of postoperative mHHS with preoperative and intraoperative variables. Ten-year Kaplan-Meier survival was calculated. RESULTS: Of 308 patients (308 hips), 111 (36%) had died with their original cups in place, 29 (9%) were lost-to-follow-up, and 5 (2%) required cup revision, leaving a final cohort of 163 (53%) with their original cup in place at a follow-up of ≥ 10 years. Ten-year survival was 98% considering cup revision for any reason as endpoint; 99% considering cup revision for aseptic loosening as endpoint; 96% considering stem revision for any reason as endpoint; and 96% considering any revision as endpoint. The final cohort of 163 patients was assessed at 11 ± 1 years (range, 10-13), mHHS was 85 ± 16 (range, 31-100) and FJS was 84 ± 24 (range, 0-100). Multivariable regression analysis revealed that postoperative mHHS significantly worsened with age (β = -0.48, p = 0.007) and BMI (β = -0.70, p = 0.008), as well as for 22 mm head sizes (β = -6.98, p = 0.046). CONCLUSIONS: The Saturne DM cup granted satisfactory survival and clinical outcomes at a minimum follow-up of 10 years, and resulted in no cases of intra- or extra-prosthetic dislocations.
The use of dual-mobility (DM) cups in total hip arthroplasty (THA) has gained popularity over the last few decades, as their advantageous head-to-neck ratio and double articulation result in lower dislocation risk and greater range of motion [5, 16]. First- and second- generation DM cups were associated with cases of intra-prosthetic dislocation and aseptic loosening; however, modern DM cups have shown low revision rates and good clinical outcomes, even in unselected populations [1, 7, 22].In 2000, the Saturne cementless DM cup (Amplitude, France) was introduced with a quasi-anatomic rim to prevent dislocations without increasing risks of psoas impingement. The rim is therefore augmented at the postero-superior half to increase the ‘jump distance’ but follows the hemispherical equator at the antero-inferior half to avoid prosthetic overhang at the psoas valley [17, 18]. In a recent study, Gaillard et al. [8] reported the 10-year survival and clinical outcomes of 310 THAs using the Saturne DM cup implanted by one surgeon at a university hospital using a posterolateral approach. These encouraging findings have not yet been corroborated by other surgeons at private hospitals, who may use different surgical techniques and implant combinations.The purpose of this study was to confirm previous findings by reporting long-term survival and clinical outcomes of primary THA using a Saturne cementless DM cup, and to investigate whether patient demographics or surgical parameters affect clinical scores.
Materials and methods
Study design
The authors retrospectively assessed a consecutive series of patients that underwent primary THA between September 2009 and December 2011 by 2 surgeons at 2 centers. Patients were included in the present study if they received the Saturne cementless DM cup. Patients were excluded from the study if they did not have ≥ 10 years follow-up. The initial cohort comprised 308 patients (308 hips), of which 150 were males and 158 were females, aged 74 ± 9 years (range, 42–91) at index surgery with body mass index (BMI) of 28 ± 5 kg/m2 (range, 18–52) (Table 1). Indications for surgery were primary osteoarthritis (n = 270, 87.7%), secondary osteoarthritis (n = 12, 3.9%), femoral neck fracture (n = 12, 3.9%), avascular necrosis (n = 11, 3.6%), and rheumatoid arthritis (n = 3, 1.0%). All patients gave informed consent to participate in the study, and this study was approved by the institutional review board of GCS Ramsay Santé pour l’Enseignement et la Recherche (IRB number: COS-RGDS-2022-08-001-GAILLARD-T).
Table 1
Pre- and intra-operative data
Original Cohort
(n = 308 hips)
Mean ± SD or n (%)
Range
Preoperative data
Age
73.6 ± 8.8
(42- 91)
BMI
27.8 ± 4.6
(18- 52)
HHS
34.1 ± 10.4
(6- 66)
Male sex
150 (49%)
Indication
Primary OA
270 (88%)
Secondary OA
12 (4%)
Femoral neck fracture
12 (4%)
Avascular necrosis
11 (4%)
RA
3 (1%)
Intraoperative data
Surgical approach
Anterolateral
300 (97%)
Anterior
6 (2%)
Posterolateral
2 (1%)
Navigation
No
196 (64%)
Yes
112 (36%)
Stem type
Modular Acor
108 (35%)
Fixed neck Integrale
166 (54%)
Modular Integrale
34 (11%)
Stem fixation
Cemented
49 (16%)
Cementless
259 (84%)
Neck length
Short
46 (15%)
Medium
207 (67%)
Long
55 (18%)
Head size
22
136 (44%)
28
172 (56%)
Cup size
44
1 (0%)
46
7 (2%)
48
53 (17%)
50
53 (17%)
52
55 (18%)
54
61 (20%)
56
56 (18%)
58
16 (5%)
60
6 (2%)
Abbreviations: BMI body mass index, HHS Harris hip score, OA osteoarthritis, RA rheumatoid arthritis, SD standard deviation
Pre- and intra-operative dataAbbreviations: BMI body mass index, HHS Harris hip score, OA osteoarthritis, RA rheumatoid arthritis, SD standard deviation
Surgical information
The surgical approach was anterolateral in most cases (n = 300, 97%), while it was anterior (n = 6, 2%) or posterolateral (n = 2, 1%) in few cases (Table 1). Navigation with a computer assisted surgery system (Amplivision, Amplitude, France) was used in over one-third of cases (n = 112, 36%). Three types of stems were used: the Integrale cementless fixed-neck stem (n = 116, 54%) (Amplitude, France), the Integrale cementless modular stem (n = 34, 11%) (Amplitude, France), and the Acor modular stem (n = 108, 35%) (Amplitude, France) available cementless and cemented.
Clinical assessment
Patients were evaluated preoperatively at the clinic using the Harris hip score (HHS). The latest clinical evaluation was performed via telephone by an independent observer, who recorded the modified HHS (mHHS), forgotten joint score (FJS), and satisfaction level (very satisfied, satisfied, disappointed, dissatisfied). Complications, reoperations, and revisions were noted from patient files and from phone interviews with patients. If a patient was deceased, the family doctor or next of kin was contacted for the date of death and to confirm that the patient died with the cup in place.
Statistical analyses
Descriptive statistics were used to summarise patient demographics, surgical data, and clinical outcomes. Preoperative HHS was converted into mHHS to calculate the net change in score. Univariable linear regression analyses were performed to determine associations of postoperative mHHS with 9 variables (age, BMI, sex, use of navigation, stem type, stem fixation, neck length, head size, and cup size). The variables: indication, surgical approach, and cup size were not included in regression analyses because at least one subgroup had less than 10 patients. The variables: stem type, stem fixation, and navigation were not included in multivariable regression analyses because of significant collinearities between them. Furthermore, the Kaplan–Meier (KM) method was used to estimate survival and 95% confidence intervals (CI) at 10 years for 4 different end points: (i) cup revision for any reason; (ii) cup revision for aseptic reasons; (iii) stem revision for any reason; (iv) any revision for any reason. Statistical analyses were performed using R version 4.1.3 (R Foundation for Statistical Computing, Vienna, Austria). P-values < 0.05 were considered statistically significant.
Results
Complications, reoperations, revisions, deaths, and losses to follow-up
From the initial cohort of 308 patients (308 hips), two (0.6%) patients had complications that did not require reoperation: one (0.32%) had an infected hematoma treated with antibiotics and one (0.32%) had a femoral fracture for which revision was contraindicated because the patient was too frail for surgery. It is important to note that there were no intra- or extra-prosthetic dislocations. Six (1.9%) patients required reoperations without implant removal: two (0.65%) had lavage/debridement for infection and four (1.3%) had osteosynthesis using plate and screws for femoral fracture. Four (1.3%) patients required cup and stem revision: two (0.65%) for periprothstic joint infection at 1 and 3 years, one (0.32%) for aseptic loosening at 6 years, and 1 (0.32%) for femoral fracture at 7 years. One (0.3%) patient required cup revision only, for psoas impingement at 8 years. Six (1.9%) patients required stem revision only: five (1.6%) for femoral fracture at 0, 1, 3, 8 and 10 years, and one (0.32%) for fracture of the modular stem neck at 4 years.At a minimum follow-up of 10 years, 111 (36%) patients had died with their original cups in place and 29 (9%) could not be reached, but their most recent follow-up records indicated that none had cup revision (Fig. 1). Patients were excluded from the final cohort if they had died, were lost to follow-up or had cup revision; thus, the final cohort comprised 163 patients (53%) with their original cup in place with a follow-up of ≥ 10 years.
Fig. 1
Flowchart indicating that from the initial cohort of 308 patients (308 hips), 111 (36%) had died with their original cups in place, 29 (9%) were lost to follow-up, and five (1.6%) required cup revision, thus leaving a final cohort of 163 patients (53%) with their original cup in place with a follow-up of ≥ 10 years
Flowchart indicating that from the initial cohort of 308 patients (308 hips), 111 (36%) had died with their original cups in place, 29 (9%) were lost to follow-up, and five (1.6%) required cup revision, thus leaving a final cohort of 163 patients (53%) with their original cup in place with a follow-up of ≥ 10 years
Survival
Survival at 10 years estimated using the KM method was (i) 98.0% (CI, 95.1–99.2%) considering cup revision for any reason as endpoint (Fig. 2); (ii) 98.6% (CI, 95.8–99.6%) considering cup revision for aseptic loosening as endpoint (Fig. 3); (iii) 96.1% (CI, 92.8–97.9%) considering stem revision for any reason as endpoint (Fig. 4); and (iv) 95.6% (CI, 92.1–97.6%) considering any revision for any reason as endpoint (Fig. 5).
Fig. 2
Survival at 10 years estimated using the KM method was 98.0% (CI, 95.1–99.2%) considering cup revision for any reason as endpoint
Fig. 3
Survival at 10 years estimated using the KM method was 98.6% (CI, 95.8–99.6%) considering cup revision for aseptic loosening as endpoint
Fig. 4
Survival at 10 years estimated using the KM method was 96.1% (CI, 92.8–97.9%) considering stem revision for any reason as endpoint
Fig. 5
Survival at 10 years estimated using the KM method was 95.6% (CI, 92.1–97.6%) considering any revision for any reason as endpoint
Survival at 10 years estimated using the KM method was 98.0% (CI, 95.1–99.2%) considering cup revision for any reason as endpointSurvival at 10 years estimated using the KM method was 98.6% (CI, 95.8–99.6%) considering cup revision for aseptic loosening as endpointSurvival at 10 years estimated using the KM method was 96.1% (CI, 92.8–97.9%) considering stem revision for any reason as endpointSurvival at 10 years estimated using the KM method was 95.6% (CI, 92.1–97.6%) considering any revision for any reason as endpoint
Clinical outcomes
The final cohort of 163 patients (163 hips) comprised 80 males and 83 females, aged 82 ± 7 years (range, 64–88) at a follow-up of 11 ± 1 years (range, 10–13). The postoperative mHHS was 85 ± 16 (range, 31–100) and the net change in mHHS was 53 ± 20 (range, -19–82; p < 0.001) (Table 2). The postoperative FJS was 84 ± 24 (range, 0–100). Most patients were very satisfied (n = 104, 64%) or satisfied (n = 53, 33%) with surgery.
Table 2
Clinical scores and satisfaction with surgery
Cohort with ≥ 10 year FU
(n = 163 hips)
Mean ± SD or n (%)
Range
Follow-up (years)
11.3 ± 0.7
(10.0 – 12.6)
Modified HHS
Preoperative
32.2 ± 10.8
(3 – 55)
Postoperative
85.4 ± 16.3
(31 – 100)
Net change
53.2 ± 20.4
(-19 – 82)
Postoperative FJS
84.0 ± 23.6
(0 – 100)
Satisfaction with surgery
Very satisfied
104 (64%)
Satisfied
53 (33%)
Dissapointed
4 (2%)
Dissatisfied
2 (1%)
Abbrevations: HHS Harris hip score, FJS forgotten joint score, FU follow-up, SD standard deviation
Clinical scores and satisfaction with surgeryAbbrevations: HHS Harris hip score, FJS forgotten joint score, FU follow-up, SD standard deviationUnivariable linear regression analyses revealed that postoperative mHHS significantly worsened with BMI (β = -0.68, p = 0.011), as well as for 22 mm head sizes (β = -6.38, p = 0.012), while it was significantly better for navigated surgery (β = 10.55, p < 0.001), modular Acor stems (β = 11.24, p < 0.001), short neck stems (β = 10.68, p = 0.003), and long neck stems (β = 8.56, p = 0.012) (Table 3). Multivariable linear regression analysis confirmed that postoperative mHHS significantly worsened with age (β = -0.48, p = 0.007) and BMI (β = -0.70, p = 0.008), as well as for 22 mm head sizes (β = -6.98, p = 0.046).
Table 3
Uni- and multi-variable regression analyses of postoperative modified Harris hip score
Univariable
Multivariable
n
β
95%CI
P-value
β
95% C.I
P-value
Preoperative data
Age
163
-0.29
(-0.64 – 0.06)
0.099
-0.48
(-0.83 – -0.13)
0.007
BMI
163
-0.68
(-1.21 – -0.16)
0.011
-0.70
(-1.21 – -0.19)
0.008
Male sex
80
2.78
(-2.26 – 7.82)
0.278
0.12
(-6.41 – 6.64)
0.972
Intraoperative data
Navigation
No
113
REF
Yes
50
10.55
(5.29 – 15.81)
< 0.001
Stem type
Modular Acor
47
11.24
(5.75 – 16.73)
< 0.001
Fixed neck Integrale
99
REF
Modular Integrale
17
1.62
(-6.45 – 9.69)
0.693
Stem fixation
Cemented
17
1.16
(-7.09 – 9.42)
0.781
Cementless
146
REF
Neck length
Short
25
10.68
(3.81 – 17.56)
0.003
6.46
(-1.55 – 14.47)
0.113
Medium
111
REF
Long
27
8.56
(1.89 – 15.22)
0.012
6.62
(-0.15 – 13.40)
0.055
Head size
28
86
REF
22
77
-6.38
(-11.35 – -1.41)
0.012
-6.98
(-13.82 – -0.13)
0.046
Cup size
Small (44–48)
36
-6.06
(-12.39 – 0.28)
0.061
-3.60
(-10.50 – 3.31)
0.305
Medium (50–54)
88
REF
Large (56–60)
39
-1.66
(-7.82 – 4.50)
0.595
-5.52
(-13.08 – 2.04)
0.151
Abbreviations: BMI body mass index, REF reference, CI confidence interval
Uni- and multi-variable regression analyses of postoperative modified Harris hip scoreAbbreviations: BMI body mass index, REF reference, CI confidence interval
Discussion
The principal findings of this study are that, even when implanted in various settings and with different implant combinations, the Saturne DM cup confirmed satisfactory survival and clinical outcomes at a minimum follow-up of 10 years. The Saturne DM cup was introduced in 2000, with a quasi-anatomic rim to prevent dislocations without increasing risks of psoas impingement, augmented at the postero-superior half to increase the ‘jump distance’ but following the hemispherical equator at the antero-inferior half to avoid prosthetic overhang at the psoas valley [17, 18]. Only one previous study has reported on 10-year survival and clinical outcomes of this DM cup, in a series of 300 THAs performed by one surgeon using a posterolateral approach at a university hospital, but these findings had not been corroborated in other settings. In the present series, there were no cases of intra- or extra-prosthetic dislocations, indicating that the quasi-anatomic rim is effective at increasing the jump distance without causing neck-cup impingement. The secondary findings of this study are that, in multi-variable linear regression analysis, postoperative mHHS was significantly worse for 22 mm head sizes, as well as for patients of greater age and BMI.Modern cementless DM cups are designed with a thin metal shell, and have porous coating on their outer surface to favour bone ingrowth, and are mirror-polished on their inner surface to optimise articulation against a large-diameter polyethylene (PE) insert, which assembles with a small retentive femoral head. The present study included young patients (< 55 years), obese patients (BMI > 30 kg/m2), as well as patients with femoral neck fracture, who may have higher risks of dislocations; nonetheless, according to the systematic review by Batailler et al. [1], DM cups provide satisfactory outcomes, which are better than those provided by standard cups, in these populations. The findings of the present study are in line with those of other modern DM cups. Ten-year survival of the acetabular cup with revision for any reason as endpoint in the literature ranged between 95–100% [7, 8, 13, 15], which is comparable to the present study (98.0%). Intra- and extra-prosthetic dislocation rates in the literature ranged between 0–2.4% and 0–1.9% respectively [2, 3, 7, 8, 13, 21], with a recent systematic review stating that there are no intraprosthetic dislocations with modern DM cups [1], these are comparable to the intra- and extra-prosthetic dislocation rates of the present study (0%). Finally, the average HHS in the literature ranged between 84–95 [2, 3, 8, 13, 21], which is comparable to the mHHS of the present study (85 ± 16).The present series had a revision rate of 1.6%, with deep infection being the main cause of revision (0.6%), followed by aseptic loosening (0.3%), femoral fracture (0.3%), and psoas impingement (0.3%). The present revision rate is comparable to that reported for other modern DM cups; while studies on smaller series (40–104 hips) had no revisions at 5–10 years of follow-up [12, 14, 19, 20], larger cohorts (167–3474 hips) had revision rates of 0.5–3.6%, at 5–13 years of follow-up [3, 4, 6–9, 11].The study by Gaillard et al. [8] evaluated 310 THAs using the Saturne cementless DM cup and the Integrale cementless stem with head sizes of 22 mm (36% of hips) or 28 mm (64% of hips), implanted using a posterolateral approach by one surgeon at a university hospital. The study reported 10-year survival considering any revision as endpoint of 98% (CI, 97–99%), postoperative HHS of 95 (range, 76–100), and no intra- or extra-prosthetic dislocations. In contrast, the present study used three different types of stems (Integrale cementless fixed-neck stems in 54% of hips, Integrale cementless modular stems in 11% of hips, and Acor cementless/cemented modular stems in 35% of hips) with head sizes of 22 mm (44% of hips) or 28 mm (56% of hips) implanted using an anterolateral approach in most cases (97% of hips) by two surgeons at private hospitals. The present study reported a comparable 10-year survival considering any revision as endpoint of 96% (CI, 92–98%), a lower postoperative mHHS of 85 (range, 31–100), and also no intra- or extra-prosthetic dislocations. It is worth noting that Gaillard et al. [8] had a smaller proportion of 22 mm heads (36% vs 44%), and included patients of younger age (68 [range, 40–84] vs 74 [range, 42–91]) and lower BMI (25 [range, 18–43] vs 28 [range, 18–52]).The effect of patient demographics and intraoperative parameters on clinical outcomes were evaluated using multivariable linear regression analysis, which revealed significantly worse postoperative mHHS for 22 mm head sizes (β = -6.98, p = 0.046), as well as for patients of greater age (β = -0.48, p = 0.007) and BMI (β = -0.70, p = 0.008). It is important to note that use of navigation, stem type, and stem fixation were not included in multivariable analysis because of significant collinearities between them. In a previous large series, Fessy et al. [7] advised against the use of 22 mm heads, as they are associated with increased risks of intraprosthetic dislocations due to reduced neck-to-head ratio, favoring earlier impingement between the neck of the stem and the retaining ring of the cup. While there were no intraprosthetic dislocations in the present series, it is possible that 22 mm heads could have limited range of motion, and thereby compromise mHHS in some patients [7, 10, 23].This retrospective study has a number of limitations. First, radiographs were not available, and therefore the effect of navigated versus manual surgery on cup positioning could not be investigated. Second, the net change in HHS could not be calculated, because postoperatively an independent observer assessed mHHS via telephone; patients were not asked to go to the clinic routinely to reduce the risk of exposure to COVID-19. Instead, the net change in mHHS is presented. Third, not all intraoperative data could be included in regression analyses as some variables had subgroups with less than 10 hips and other variables demonstrated significant collinearities.
Conclusions
This study has demonstrated that, even when implanted in various settings and with different implant combinations, the Saturne DM cup granted satisfactory survival and clinical outcomes at a minimum follow-up of 10 years, and resulted in no cases of intra- or extra-prosthetic dislocations.