| Literature DB >> 36217120 |
Zhiqing Yang1,2, Na Miao1, Lixiang Wang1, Ke Ma3.
Abstract
BACKGROUND: To compare the accuracy of non-contact tonometry, Pentacam, and corneal visualization Scheimpflug technology (Corvis ST) for the measurement of intraocular pressure (IOP) after small incision lenticule extraction (SMILE) or femtosecond laser-assisted in situ keratomileusis (FS-LASIK) surgery.Entities:
Keywords: Corvis ST; Intraocular pressure; Non-contact tonometry; Pentacam; Refractive corneal surgery; Topography map
Mesh:
Year: 2022 PMID: 36217120 PMCID: PMC9549448 DOI: 10.1186/s12886-022-02620-7
Source DB: PubMed Journal: BMC Ophthalmol ISSN: 1471-2415 Impact factor: 2.086
Comparison of IOP measurements with different methods and correction formulations before and after SMILE surgery
| Timepoint | No | NCT- measured IOP | Corvis ST-corrected IOP (bIOP) | Pentacam-corrected IOP (mmHg) | ||||
|---|---|---|---|---|---|---|---|---|
| Ehlers | Shah | Dresden | Orssengo-Pye | Kohlhaas | ||||
| (IOPe) | (IOPs) | (IOPd) | (IOPo) | (IOPk) | ||||
| Preoperative | 58 | 15.78 ± 2.20 | 15.82 ± 1.41 | 15.35 ± 2.02 | 15.73 ± 1.89 | 15.74 ± 1.88 | 15.50 ± 1.75 | 16.28 ± 1.95 |
| 1 month | 11.19 ± 2.06 | 14.73 ± 1.48 | 17.19 ± 2.20 | 15.68 ± 1.96 | 14.77 ± 1.91 | 14.36 ± 2.45 | 14.45 ± 1.92 | |
| Change | -4.59 ± 2.63 | -1.09 ± 1.95 | 1.83 ± 2.43 | -0.05 ± 2.39 | -0.97 ± 2.42 | -1.14 ± 2.78 | -1.83 ± 2.44 | |
| 13.27 | 4.23 | -5.75 | 0.17 | 3.04 | 3.12 | 5.7 | ||
| < 0.001 | < 0.001 | < 0.001 | 0.869 | 0.004 | 0.030 | < 0.001 | ||
1 mmHg = 0.133 kPa
Comparison of IOP measurements with different methods and correction formulations before and after FS-LASIK surgery
| Timepoint | No | NCT- measured IOP | Corvis ST-corrected IOP (bIOP) | Pentacam-corrected IOP (mmHg) | ||||
|---|---|---|---|---|---|---|---|---|
| Ehlers | Shah | Dresden | Orssengo-Pye | Kohlhaas | ||||
| (IOPs) | (IOPd) | (IOPo) | (IOPk) | |||||
| Preoperative | 40 | 15.58 ± 2.42 | 16.85 ± 2.10 | 16.40 ± 2.10 | 16.50 ± 2.08 | 16.35 ± 2.10 | 16.10 ± 2.22 | 16.40 ± 2.42 |
| 1 month | 40 | 12.14 ± 3.18 | 15.33 ± 2.43 | 19.30 ± 3.10 | 17.46 ± 2.43 | 16.38 ± 3.00 | 16.39 ± 3.89 | 15.90 ± 3.22 |
| Change | -3.67 ± 2.70 | -1.52 ± 2.01 | 2.90 ± 2.92 | 0.96 ± 2.79 | 0.03 ± 2.75 | 0.29 ± 3.49 | -0.50 ± 2.60 | |
| 8.58 | 4.77 | -6.28 | -2.18 | -0.08 | -0.52 | 1.22 | ||
| < 0.001 | < 0.001 | < 0.001 | 0.035 | 0.941 | 0.604 | 0.231 | ||
Comparison of IOP measurements with different methods and correction formulations before and after either type of corneal refractive surgeries
| Time | No. eyes | NCT- measured IOP | Corvis ST-corrected (bIOP) | Pentacam-corrected IOP (mmHg) | ||||
|---|---|---|---|---|---|---|---|---|
| Ehlers IOP (IOPe) | Shah | Dresden | Orssengo-Pye | Kohlhaas | ||||
| (IOPs) | (IOPd) | (IOPo) | (IOPk) | |||||
| Preoperative | 98 | 15.79 ± 2.28 | 16.24 ± 1.79 | 15.78 ± 2.11 | 16.04 ± 1.99 | 15.99 ± 1.99 | 15.75 ± 1.97 | 16.33 ± 2.14 |
| 1 month | 98 | 11.58 ± 2.60 | 14.98 ± 1.94 | 18.05 ± 2.78 | 16.41 ± 2.58 | 15.43 ± 2.53 | 15.19 ± 3.25 | 15.04 ± 2.62 |
| Changes | -4.21 ± 2.69 | -1.26 ± 1.98 | 2.27 ± 2.68 | 0.36 ± 2.60 | -0.56 ± 2.60 | -0.56 ± 3.16 | -1.29 ± 2.58 | |
| t value | 15.52 | 6.31 | -8.38 | -1.38 | 2.13 | 1.75 | 4.94 | |
| < 0.001 | < 0.001 | < 0.001 | 0.17 | 0.035 | 0.084 | < 0.001 | ||