Literature DB >> 15093634

Assessment of the Pascal dynamic contour tonometer in monitoring intraocular pressure in unoperated eyes and eyes after LASIK.

Dimitrios S Siganos1, Georgios I Papastergiou, Carlos Moedas.   

Abstract

PURPOSE: To assess the performance of the Pascal dynamic contour tonometer (PDCT) (Swiss Microtechnology AG) by comparing it to Goldmann applanation tonometry (GAT) and noncontact air tonometry (NCT) before and after laser in situ keratomileusis (LASIK).
SETTING: Vlemma Eye Institute, Athens, Greece.
METHODS: Intraocular pressure was measured in 118 eyes before and 1 and 4 weeks after LASIK using GAT, NCT, and the PDCT, which allows direct, digital, transcorneal measurement of intraocular pressure.
RESULTS: Preoperatively, central corneal thickness (CCT) correlated with GAT and NCT measurements but not with PDCT measurements. After LASIK, the mean reduction in CCT ranged from -3.0 to -171.0 microm (median 78 microm). The mean GAT measurement dropped by -4.9 mm Hg +/- 2.7 (SD) at 1 week and was -5.4 +/- 3.0 mm Hg at 4 weeks. Similar drops were observed in NCT measurements. The observed post-LASIK changes in GAT and NCT IOP measurements were not directly proportional to the change in CCT, refractive error, or mean keratometric readings. The preoperative and postoperative PDCT measurements did not differ significantly.
CONCLUSIONS: The structural corneal changes induced by LASIK appeared to influence GAT and NCT measurements but not PDCT measurements. Therefore, PDCT may be better suited for monitoring IOP in unoperated eyes and in eyes that have had LASIK.

Entities:  

Mesh:

Year:  2004        PMID: 15093634     DOI: 10.1016/j.jcrs.2003.12.033

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  27 in total

1.  Intraocular pressure and ocular pulse amplitude variations during the Valsalva maneuver.

Authors:  Umit Aykan; Mesut Erdurmus; Burhan Yilmaz; Ahmet Hamdi Bilge
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2010-03-24       Impact factor: 3.117

2.  Comparison between Pascal dynamic contour tonometer and Goldmann applanation tonometer after different types of refractive surgery.

Authors:  Antonios P Aristeidou; Georgios Labiris; Andreas Katsanos; Michalis Fanariotis; Nikitas C Foudoulakis; Vassilios P Kozobolis
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2010-06-25       Impact factor: 3.117

3.  The influence of central corneal thickness and age on intraocular pressure measured by pneumotonometry, non-contact tonometry, the Tono-Pen XL, and Goldmann applanation tonometry.

Authors:  P-A Tonnu; T Ho; T Newson; A El Sheikh; K Sharma; E White; C Bunce; D Garway-Heath
Journal:  Br J Ophthalmol       Date:  2005-07       Impact factor: 4.638

Review 4.  [Dynamic contour tonometry].

Authors:  C Kniestedt; H E Kanngiesser
Journal:  Ophthalmologe       Date:  2006-08       Impact factor: 1.059

5.  Underestimation of intraocular pressure after photorefractive keratectomy: a biomechanical analysis.

Authors:  Per Hallberg; Anders Eklund; Kenneth Santala; Timo Koskela; Olof Lindahl; Christina Lindén
Journal:  Med Biol Eng Comput       Date:  2006-07-26       Impact factor: 2.602

6.  [Influence of residual corneal bed thickness after myopic LASIK on intraocular pressure measurements. Goldmann applanation tonometry and dynamic contour tonometry].

Authors:  L Muller; T Kohnen
Journal:  Ophthalmologe       Date:  2009-01       Impact factor: 1.059

7.  The relative effects of corneal thickness and age on Goldmann applanation tonometry and dynamic contour tonometry.

Authors:  A Kotecha; E T White; J M Shewry; D F Garway-Heath
Journal:  Br J Ophthalmol       Date:  2005-12       Impact factor: 4.638

8.  Dynamic contour tonometry over silicone hydrogel contact lens.

Authors:  Andrew K C Lam; Jimmy S H Tse
Journal:  J Optom       Date:  2013-08-28

9.  Physiological diurnal variability and characteristics of the ocular pulse amplitude (OPA) with the dynamic contour tonometer (DCT-Pascal).

Authors:  Sayeh Pourjavan; Pierre-Yves Boëlle; Michéle Detry-Morel; Patrick De Potter
Journal:  Int Ophthalmol       Date:  2007-10-23       Impact factor: 2.031

10.  Comparison of Goldmann applanation and dynamic contour tonometry in a population of Mexican open-angle glaucoma patients.

Authors:  J Jimenez-Roman; F Gil-Carrasco; A Martinez; O Albis-Donado; J D De la Torre-Tovar
Journal:  Int Ophthalmol       Date:  2012-11-16       Impact factor: 2.031

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.