Suyon Chang1, Jinho Park2, Young-Joong Yang2, Kyongmin Sarah Beck1, Pan Ki Kim2, Byoung Wook Choi2,3, Jung Im Jung1.
Abstract
BACKGROUND: Myocardial T2* mapping at 1.5T remains the gold standard, but the use of 3T scanners is increasing. We aimed to determine the conversion equations in different scanners with clinically available, vendor-provided T2* mapping sequences using a phantom and evaluated the feasibility of the phantom-based conversion method.Entities:
Keywords: 3T; T2*; heart; iron overload; magnetic resonance imaging
Year: 2022 PMID: 36005435 PMCID: PMC9410052 DOI: 10.3390/jcdd9080271
Source DB: PubMed Journal: J Cardiovasc Dev Dis ISSN: 2308-3425
Figure 1T2* maps in a phantom (A) and normal myocardium (B).
T2* values in a phantom study measured by two radiologist readers.
| Concentration (mM) | Institution | Scanner | Vendor | Field Strength | T2* (Mean ± Standard Deviation) | ||
|---|---|---|---|---|---|---|---|
| Reader 1 + Reader 2 | Reader 1 | Reader 2 | |||||
| Overall | A | MR-A1 | Philips | 1.5T | 29.6 ± 22.0 | 29.5 ± 22.2 | 29.7 ± 22.5 |
| MR-A2 | Philips | 3.0T | 21.3 ± 15.7 | 21.4 ± 16.0 | 21.3 ± 15.9 | ||
| MR-A3 | Siemens | 3.0T | 12.8 ± 7.7 | 12.8 ± 7.8 | 12.9 ± 7.9 | ||
| B | MR-B | Siemens | 3.0T | 20.8 ± 15.7 | 20.8 ± 16.1 | 20.8 ± 15.9 | |
| 3.53 | A | MR-A1 | Philips | 1.5T | 62.5 ± 0.7 | 62.0 ± 0.2 | 62.9 ± 0.7 |
| MR-A2 | Philips | 3.0T | 44.2 ± 0.3 | 44.4 ± 0.3 | 44.0 ± 0.3 | ||
| MR-A3 | Siemens | 3.0T | 24.4 ± 0.2 | 24.3 ± 0.3 | 24.5 ± 0.2 | ||
| B | MR-B | Siemens | 3.0T | 45.5 ± 0.4 | 45.8 ± 0.1 | 45.2 ± 0.2 | |
| 4.32 | A | MR-A1 | Philips | 1.5T | 44.5 ± 0.1 | 44.5 ± 0.1 | 44.5 ± 0.0 |
| MR-A2 | Philips | 3.0T | 32.6 ± 0.2 | 32.5 ± 0.2 | 32.8 ± 0.1 | ||
| MR-A3 | Siemens | 3.0T | 17.4 ± 0.2 | 17.3 ± 0.1 | 17.5 ± 0.2 | ||
| B | MR-B | Siemens | 3.0T | 29.9 ± 0.1 | 29.9 ± 0.1 | 29.9 ± 0.1 | |
| 5.90 | A | MR-A1 | Philips | 1.5T | 26.8 ± 0.1 | 26.8 ± 0.2 | 26.8 ± 0.2 |
| MR-A2 | Philips | 3.0T | 19.5 ± 0.1 | 19.5 ± 0.1 | 19.5 ± 0.1 | ||
| MR-A3 | Siemens | 3.0T | 12.7 ± 0.1 | 12.7 ± 0.1 | 12.7 ± 0.1 | ||
| B | MR-B | Siemens | 3.0T | 18.2 ± 0.1 | 18.2 ± 0.1 | 18.3 ± 0.1 | |
| 10.63 | A | MR-A1 | Philips | 1.5T | 10.0 ± 0.0 | 10.0 ± 0.0 | 10.0 ± 0.0 |
| MR-A2 | Philips | 3.0T | 7.2 ± 0.1 | 7.2 ± 0.1 | 7.1 ± 0.1 | ||
| MR-A3 | Siemens | 3.0T | 6.0 ± 0.1 | 6.0 ± 0.1 | 6.0 ± 0.0 | ||
| B | MR-B | Siemens | 3.0T | 7.2 ± 0.0 | 7.2 ± 0.1 | 7.2 ± 0.0 | |
| 20.09 | A | MR-A1 | Philips | 1.5T | 4.2 ± 0.0 | 4.2 ± 0.0 | 4.2 ± 0.1 |
| MR-A2 | Philips | 3.0T | 3.2 ± 0.0 | 3.2 ± 0.0 | 3.2 ± 0.0 | ||
| MR-A3 | Siemens | 3.0T | 3.7 ± 0.1 | 3.7 ± 0.1 | 3.7 ± 0.0 | ||
| B | MR-B | Siemens | 3.0T | 3.2 ± 0.1 | 3.2 ± 0.0 | 3.2 ± 0.1 | |
Linear regression model of 1.5T on 3.0T.
| Institution | Scanner | Vendor | Field Strength | Reader 1 + Reader 2 | Reader 1 | Reader 2 | |||
|---|---|---|---|---|---|---|---|---|---|
| Model | R-Square | Model | R-Square | Model | R-Square | ||||
| 1st + 2nd + 3rd | |||||||||
| A | MR-A2 | Philips | 3.0T | −2.23 + 0.76 × 3.0T | 0.999 | −2.23 + 0.76 × 3.0T | 0.999 | −2.23 + 0.76 × 3.0T | 0.999 |
| MR-A3 | Siemens | 3.0T | −34.28 + 0.95 × 3.0T | 0.973 | −34.23 + 0.95 × 3.0T | 0.975 | −34.33 + 0.95 × 3.0T | 0.972 | |
| B | MR-B | Siemens | 3.0T | −3.02 + 0.76 × 3.0T | 0.999 | −2.75 + 0.75 × 3.0T | 1 | −3.30 + 0.76 × 3.0T | 0.999 |
| 1st | |||||||||
| A | MR-A2 | Philips | 3.0T | −2.45 + 0.77 × 3.0T | 0.999 | −2.18 + 0.76 × 3.0T | 0.999 | −2.73 + 0.77 × 3.0T | 0.999 |
| MR-A3 | Siemens | 3.0T | −35.28 + 0.96 × 3.0T | 0.972 | −35.59 + 0.97 × 3.0T | 0.973 | −34.99 + 0.96 × 3.0T | 0.972 | |
| B | MR-B | Siemens | 3.0T | −2.37 + 0.75 × 3.0T | 1 | −2.52 + 0.75 × 3.0T | 1 | −2.23 + 0.74 × 3.0T | 1 |
| 2nd | |||||||||
| A | MR-A2 | Philips | 3.0T | −2.42 + 0.77 × 3.0T | 0.999 | −2.60 + 0.77 × 3.0T | 0.999 | −2.24 + 0.76 × 3.0T | 0.999 |
| MR-A3 | Siemens | 3.0T | −33.89 + 0.95 × 3.0T | 0.975 | −33.66 + 0.94 × 3.0T | 0.98 | −34.17 + 0.95 × 3.0T | 0.971 | |
| B | MR-B | Siemens | 3.0T | −3.28 + 0.76 × 3.0T | 0.999 | −3.16 + 0.76 × 3.0T | 0.999 | −3.42 + 0.77 × 3.0T | 0.999 |
| 3rd | |||||||||
| A | MR-A2 | Philips | 3.0T | −1.84 + 0.75 × 3.0T | 0.999 | −1.94 + 0.75 × 3.0T | 0.999 | −1.73 + 0.75 × 3.0T | 1 |
| MR-A3 | Siemens | 3.0T | −33.72 + 0.94 × 3.0T | 0.973 | −33.65 + 0.93 × 3.0T | 0.973 | −33.84 + 0.95 × 3.0T | 0.974 | |
| B | MR-B | Siemens | 3.0T | −3.47 + 0.77 × 3.0T | 0.999 | −2.58 + 0.75 × 3.0T | 1 | −4.54 + 0.79 × 3.0T | 0.999 |
Figure 2Scatter plots of a phantom with linear regression with repeated measures by two radiologist readers.
Results of intra-observer and inter-observer reproducibility analyses for the phantom study.
| Scanner | Intra-Observer Reliability | Inter-Observer Reliability | ||||
|---|---|---|---|---|---|---|
| Reader 1 | Reader 2 | 1st + 2nd + 3rd | 1st | 2nd | 3rd | |
| MR-A1 | 1 | 0.9998 | 0.9997 | 0.9996 | 1 | 0.9996 |
| (0.9998–1.0000) | (0.9991–1.0000) | (0.9991–0.9999) | (0.9971–1.0000) | (0.9998–1.0000) | (0.9971–1.0000) | |
| MR-A2 | 0.9999 | 0.9999 | 0.9998 | 0.9997 | 0.9999 | 0.9999 |
| (0.9994–1.0000) | (0.9997–1.0000) | (0.9995–0.9999) | (0.9978–1.0000) | (0.9995–1.0000) | (0.9992–1.0000) | |
| MR-A3 | 0.9997 | 0.9999 | 0.9995 | 0.9999 | 0.9996 | 0.9993 |
| (0.9986–1.0000) | (0.9994–1.0000) | (0.9986–0.9998) | (0.9993–1.0000) | (0.9967–1.0000) | (0.9946–0.9999) | |
| MR-B | 1 | 1 | 0.9998 | 0.9999 | 0.9998 | 0.9999 |
| (1.0000–1.0000) | (0.9998–1.0000) | (0.9995–0.9999) | (0.9995–1.0000) | (0.9983–1.0000) | (0.9991–1.0000) | |
Data are intra-class correlation coefficient (95% confidence interval).
Comparison of T2* measured in 1.5T (T2*1.5T) and equivalent T2* values from 3.0T (T2*eq) in normal myocardium.
| Values | Reader 1 + Reader 2 | Reader 1 | Reader 2 | ||||||
|---|---|---|---|---|---|---|---|---|---|
| T2* | Difference | T2* | Difference | T2* | Difference | ||||
| (Mean ± SD) | (95% CI) | (Mean ± SD) | (95% CI) | (Mean ± SD) | (95% CI) | ||||
| 1st + 2nd + 3rd | |||||||||
| T2*1.5T | 34.5 ± 1.2 | −0.94 | 0.34 | 34.6 ± 1.4 | −0.82 | 0.625 | 34.5 ± 1.0 | −1.05 | 0.373 |
| T2*eq | 35.5 ± 3.5 | (−2.95–1.08) | 35.4 ± 3.8 | (−4.56–2.91) | 35.5 ± 3.4 | (−3.61–1.51) | |||
| 1st | |||||||||
| T2 *1.5T | 34.7 ± 1.2 | −1.19 | 0.484 | 34.8 ± 1.8 | −1.22 | 0.721 | 34.6 ± 0.6 | −1.16 | 0.605 |
| T2*eq | 35.9 ± 3.3 | (−5.23–2.85) | 36.0 ± 4.1 | (−13.9–11.50) | 35.8 ± 3.2 | (−9.34–7.03) | |||
| 2nd | |||||||||
| T2*1.5T | 34.5 ± 0.9 | −0.93 | 0.626 | 34.6 ± 1.3 | −0.29 | 0.937 | 34.4 ± 0.4 | −1.58 | 0.566 |
| T2*eq | 35.4 ± 3.9 | (−5.55–3.69) | 34.9 ± 4.3 | (−14.2–13.58) | 36.0 ± 4.4 | (−11.5–8.38) | |||
| 3rd | |||||||||
| T2*1.5T | 34.4 ± 1.6 | −0.69 | 0.73 | 34.4 ± 1.6 | −0.96 | 0.809 | 34.4 ± 1.9 | −0.42 | 0.876 |
| T2*eq | 35.1 ± 3.8 | (−5.53–4.15) | 35.4 ± 4.5 | (−16.0–14.04) | 34.8 ± 4.0 | (−10.522129.67) | |||
CI: confidence interval; SD: standard deviation.
Figure 3T2* values in three healthy volunteers with repeated measures by two radiologist readers. T2*3.0T = T2* measured at 3T, T2*eq = equivalent T2* calculated from the site-specific equation, T2*1.5T= T2* measured at 1.5T.
Figure 4Bland–Altman plots of equivalent T2* values and measured T2* values at 1.5T.