Barbara Zollner1, Christian Heinz1, Sabrina Pitzler1, Farkhad Manapov1, Steffi Kantz1, Maya Christine Rottler1, Maximilian Niyazi1, Ute Ganswindt1, Claus Belka1, Hendrik Ballhausen2.
Abstract
BACKGROUND: To assess the accuracy and precision of a fully integrated pilot installation of stereoscopic X-ray imaging and kV-CBCT for automatic couch positioning in stereotactic radiotherapy of intracranial tumors. Positioning errors as detected by stereoscopic X-ray imaging are compared to those by kV-CBCT (i.e. the accuracy of the new method is verified by the established method), and repeated X-ray images are compared (i.e. the precision of new method is determined intra-modally).Entities:
Keywords: Cone beam computed tomography; Image guided radiotherapy; Intracranial tumors; Patient positioning; Radiosurgery; Stereoscopic X-ray imaging; Stereotactic radiotherapy
Mesh:
Year: 2016 PMID: 27927235 PMCID: PMC5142336 DOI: 10.1186/s13014-016-0735-2
Source DB: PubMed Journal: Radiat Oncol ISSN: 1748-717X Impact factor: 3.481
Fig. 1Installation of Elekta Versa HD with integrated kV-CBCT (highlighted in blue), Brainlab ExacTrac stereoscopic X-ray imaging (detectors highlighted in red, sources are hidden in floor) and robotic patient couch with 6 DOF Hexapod (highlighted in green)
Discrepancies in position detection between ExacTrac and CBCT (N = 42)
| mean ± SD | median | range | 95% (abs.) | |
|---|---|---|---|---|
| lateral | +0.1 ± 0.3 | +0.1 | –0.9 to +0.8 | <0.8 |
| longitudinal | –0.5 ± 0.4 | –0.4 | –1.6 to +0.2 | <1.2 |
| vertical | –0.3 ± 0.5 | –0.4 | –1.4 to +0.8 | <1.0 |
| radial | 0.8 ± 0.3 | 0.8 | 0.3 to 1.8 | <1.4 |
CBCT Cone Beam Computed Tomography
All units are in mm
Improvement after correction and re-verification by ExacTrac (N = 70)
| mean ± SD | median | range | 95% (abs.) | |
|---|---|---|---|---|
| Before correction by ExacTrac | ||||
| lateral | –0.7 ± 2.3 | –0.7 | –7.7 to +5.9 | <5.2 |
| longitudinal | +0.3 ± 1.5 | +0.2 | –3.6 to +4.6 | <3.5 |
| vertical | –1.2 ± 1.8 | –0.8 | –10.7 to +1.3 | <3.1 |
| radial | 2.9 ± 2.0 | 2.3 | 0.2 to 11.8 | <6.1 |
| After correction by ExacTrac | ||||
| lateral | +0.03 ± 0.25 | +0.02 | –0.40 to +1.13 | <0.38 |
| longitudinal | –0.02 ± 0.21 | –0.04 | –0.54 to +0.95 | <0.40 |
| vertical | +0.01 ± 0.22 | +0.01 | –0.67 to +0.56 | <0.51 |
| radial | 0.32 ± 0.22 | 0.26 | 0.07 to 1.33 | <0.69 |
All units are in mm
Repeatability of position detection by ExacTrac (N = 14)
| mean ± SD | range | 95% (abs.) | 50% (abs.) | |
|---|---|---|---|---|
| Difference before – after | ||||
| lateral | –0.08 ± 0.52 | –0.94 to +0.53 | <0.8 | <0.5 |
| longitudinal | –0.05 ± 0.70 | –0.74 to +1.01 | <0.8 | <0.1 |
| vertical | –0.03 ± 0.42 | –0.67 to +0.83 | <0.7 | <0.3 |
| radial | 0.73 ± 0.32 | 0.26 to 1.28 | <1.2 | <0.7 |
All units are in mm
Shifts of a head phantom as preset by the robotic couch and as detected by both ExacTrac and CBCT (N = 24, resp. 8 per axis)
| Discrepancy between … | … ExacTrac and robotic couch | … CBCT and robotic couch | … ExacTrac and CBCT |
|---|---|---|---|
| lateral | 0.0 ± 0.5 ( | 0.0 ± 0.6 ( | 0.0 ± 0.1 ( |
| longitudinal | 0.4 ± 0.3 ( | 0.3 ± 0.3 ( | 0.1 ± 0.1 ( |
| vertical | 0.0 ± 0.4 ( | 0.0 ± 0.4 ( | 0.0 ± 0.1 ( |
| radial | 0.7 ± 0.3 ( | 0.7 ± 0.3 ( | 0.1 ± 0.0 ( |
All units are in mm, average ± SD (p values from two-sided t-test)
Time requirements during the workflow of the head phantom measurement (N = 24)
| Gantry in position | Scan | Registration | |
|---|---|---|---|
| ExacTrac | 15.7 ± 2.9 | 6,8 ± 0.3 | 4.1 ± 0.2 |
| CBCT | 25.1 ± 1.1 | 41.2 ± 0.2 | 8.2 ± 2.0 |
All units are in seconds, average ± SD